MaxPlus, MaxGuard recallZ-1995-2021 · Carefusion · initiated Jun 9, 2021

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Carefusion or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

A portion of a validation lot was inadvertently released to the US Market. The product did not meet the validation testing criteria regarding leakage.

FDA root cause category: Under Investigation by firm

What the company is doing

On 06/09/2021, the firm sent an "URGENT MEDICAL DEVICE RECALL" Notification letter via email and FedEx to customers informing them that a validation lot was inadvertently released to the US market. This product did not meet the validation testing criteria regarding leakage. Potential health consequences associated with the use of the non-validated product is under-infusion. Customers are instructed to: 1. Immediately review their inventory for the specific Catalog (Ref) and lot numbers listed. Destroy all product subject to the recall following their institutions process for destruction. The Recalling Firm will replace discarded product. 2. Share this recall notification with all users within or outside their facility network to which you they have further distributed the affected products to ensure they are also aware of this recall. 3. Complete the attached Customer Response Form and return to the Recalling Firm's contact noted on the form (BD Post Market Quality Email: [email protected], Or Fax No.: 312-949-0236) whether or not they have any of the impacted material so that the Recalling Firm may acknowledge their receipt of this notification per FDA requirements and subsequently process their product replacement. For any questions or assistance contact: North American Regional Complaint Center 1-84BD-LIFE (1-844-823-5433) Say Recall when prompted MonFri 8:00am and 5:00pm CT

Recall details

Recall number
Z-1995-2021
Status
Terminated
Initiated by the company
Jun 9, 2021
Recalling firm
Carefusion
Quantity
100 units
Product code
FPA
Distribution
U.S.: TN O.U.S.: None

Product description

MaxPlusTM bi fuse extension set with 2 clear needleless connectors - Product Usage: The needleless connector can be used for direct injection, intermittent infusion, continuous infusion, or aspiration.

Code information (models, lots, serial numbers)

Catalog Number:MP9246-C GTIN: 10885403237157 Lot Number: 20087194