FilmArray Pneumonia Panel (Pneumo) recallZ-2040-2021 · BioFire Diagnostics, LLC · initiated Jun 7, 2021
Terminated
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from BioFire Diagnostics, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Due to real-time stability study failures resulting in increase rate of false negative results between the 6 to 12 month expiry time-frame.
FDA root cause category: Under Investigation by firm
What the company is doing
On 06/03/2021, the firm sent an "URGENT: MEDICAL DEVICE CORRECTION" Notification Letter via overnight FedEx informing its customer that it has identified a potential for obtaining elevated rates of negative adenovirus results when using the pneumonia panel. The potential false negative results are due to a 10-100x reduction in sensitivity specifically for adenovirus species C when using the pneumonia panel kits within 6 months of expiration. In addition: -The test performance is NOT impacted if kits are more than 6 months from expiration date. -Performance for other adenovirus species (e.g. A, B, D, E, F, and G) are NOT affected -The adenovirus assays on all other of the Recalling Firm's Respiratory Panels are NOT affected. Customer are instructed to: 1) Immediately examine their inventory for product identified in this recall. -If the customer identifies any affected Pneumonia Panels in their inventory (i.e. Pneumonia Panels within 6 months of shelf-life), they may continue using the affected product; however, when adenovirus C infection is suspected, negative results for Adenovirus should be confirmed by another method prior to reporting test results to clinicians, or alternatively, the Recalling Firm's Pneumonia Panel Adenovirus negative results should not be reported. 2) If a customer has further distributed this product, they are to identify their customers and notify them at once of this product recall. 3) Complete the enclosed Acknowledgement of Receipt Form as soon as possible and return it to the Recalling Firm via one of the methods listed on the Acknowledgement of Receipt Form - 1) Fill out electronically using the following URL: (https://www.biofiredx.com/pn-5183/) 2) Scan to PDF and return completed form to: [email protected] Any questions or assistance - contact customer support department at [email protected] or via telephone 1-800-735-6544 option 5.
Recall details
- Recall number
- Z-2040-2021
- Status
- Terminated
- Initiated by the company
- Jun 7, 2021
- Recalling firm
- BioFire Diagnostics, LLC
- Quantity
- 3,559 kits (Globally)
- Product code
- QDS
- FDA submissions
- K181324
- Distribution
- U.S.: AK, AL, AZ, CA, CA, CO, CT, DC, DE, FL, GA, Guam, HI, IA, ID, IL, IN, IN, KS, KY, KY, LA, MA, MD, ME, MI, MI, MN, MO, MO, MS, MT, NC, ND, NE, NJ, NM, NV, NY, OH, OK, OR, PA, Puerto Rico, SC, TN, TX, UT, VA, VT, WA, WI, WV and WY. O.U.S.: Algeria, Angola, Anguilla, Argentina, Australia, Austria, Bahrain, Belarus, Belgium, Bermuda, Bolivia, Bosnia-Herz., Brazil, Bulgaria, Chile, China, Colombia, Costa Rica, Cote d'Ivoire, Croatia, Czech Republic, Dominican Republic, Ecuador, Egypt, El Salvador, Estonia, France, Georgia, Germany, Greece , Guatemala, Honduras, Hong Kong, Hungary, India, Indonesia, Iraq, Israel, Italy, Japan, Jordan, Kenya, Kuwait, Latvia, Lebanon, Lithuania, Macedonia, Malaysia, Mexico, Morocco, Myanmar, Nepal, Netherlands, Nicaragua, Oman, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Qatar, Romania, Saudi Arabia, Serbia, Singapore, Slovenia, South Africa, Spain, Sweden, Switzerland, Thailand, Tunisia, Turkey, United Kingdom, Uruguay, United Arab Emirates, and Vietnam.
Product description
FilmArray Pneumonia Panel plus (Pneumoplus) Part Numbers: Part No: RFIT-ASY-0143 / UDI: 00815381020314 (30 test kit) Part No: RFIT-ASY-0142 / UDI: 00815381020321 (6 test kit)
Code information (models, lots, serial numbers)
Part No: RFIT-ASY-0144 / UDI: 00815381020178 (30 test kit) Part No: RFIT-ASY-0145 / UDI: 00815381020185 (6 test kit)
Devices named in this recall
2 devices in the FDA device registry whose identifier appears in the recall's code information.
- FilmArray® Pneumonia Panel (Pneumo) RFIT-ASY-0144Biofire Diagnostics, LLC · Respiratory Panel5 recalls7 reports
- FilmArray® Pneumonia Panel (Pneumo) RFIT-ASY-0145Biofire Diagnostics, LLC · Respiratory Panel2 recalls0 reports
Product lines
Other devices under the same FDA submission
1 device cleared under K181324 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- FilmArray® Pneumonia Panel plus (Pneumoplus) RFIT-ASY-0143Biofire Diagnostics, LLC · Mers-Cov And Common Respiratory Pathogens Semi-Quantitative And Quantitative Multiplex Nucleic Acid Detection System3 recalls3 deaths4 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 88117).
- Z-2039-2021 · TerminatedFilmArray Pneumonia Panel (Pneumo) Part Numbers: Part No: RFIT-ASY-0144 / UDI: 00815381020178 (30 test kit) Part No: RFIT-ASY-0145 / UDI: 00815381020185 (6 test kit)
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page