Philips Respironics V60 Ventilator with Software Versions… recallZ-2073-2021 · Respironics California, LLC · initiated Jun 18, 2021

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Respironics California, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

It has been identified that ventilators equipped with High Flow Therapy (Software Versions 3.00 and 3.10), that includes a designed safety mechanism that limits the amount of pressure that can be delivered to a patient, in situations where the system pressure reaches the maximum limit and sounds the "Cannot Reach Target Flow" alarm , the system will reduce pressure and simultaneously decreasing the flow rate to a level below what was set by the clinicians. If the reason for the increase in pressure is not resolved, the ventilator will continue to provide the lower flow rate, which result in patients experiencing oxygen desaturation which can be characterized as moderate or severe hypoxemia.

FDA root cause category: Device Design

What the company is doing

On 06/18/2021, the Recalling Firm sent an "URGENT - Medical Device Recall" Notification Letter via USPS certified mail to customers informing them that a problem has been detected in their ventilators that could pose a risk to patients or users. All units with High Flow Therapy option (Software Versions 3.00 and 3.10) are impacted. These ventilators with the High Flow Therapy software were designed with a safety mechanism to limit the amount of pressure that can be delivered to the patient when the ventilator is operating in High Flow Therapy. In situations where the high flow therapy pressure reaches the maximum limit, the ventilator will sound a low priority "Cannot Reach Target Flow (CRTF) alarm and reduces the pressure which also simultaneously decreases the flow rate to a level below what was set by the clinician for as long as the condition persists. Customers are instructed/informed to: When administering HFT via High Flow Nasal Cannula the following action must be taken: (1) Monitor the patients SpO2 continuously; (2) All patients that are dependent on supplemental oxygen, and at risk of clinical deterioration should be under constant and close monitoring by the clinician to prevent dangerous drops in blood oxygen levels, work of breathing and respiratory distress and resulting escalation in medical treatment (3) Respond to all alarms urgently, regardless of alarm priority; (4) If the patient cannot be constantly and closely monitored by the clinician, do not use High Flow Therapy. In addition; (5) Refer to and follow the attached user manual addendum titled V60 Plus User Manual Addendum: High-Flow Therapy Safety and Alarm Features. This user manual addendum provides additional technical details about the CRTF alarm functionality. The addendum is also available at www.philips.com/hrcmanuals under the V60/V60 Plus link titled "V60 Plus User Manual Addendum - High-Flow Therapy Safety and Alarm Features. For additional information and educ

Recall details

Recall number
Z-2073-2021
Status
Open, Classified
Initiated by the company
Jun 18, 2021
Recalling firm
Respironics California, LLC
Quantity
Total of V60/V60 Plus = 23,115 systems
Product code
MNT
Distribution
Worldwide Distribution: U.S (nationwide) to states of.: AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SC, TN, TX, UT, VA, VT, WA, WI, WV, and WY; and O.U.S. countries of: Andorra, United Arab Emirates, Albania, Argentina, Austria, Australia, Belgium, Canada, Switzerland, Chile, China, Colombia, Czech Republic, Germany, Denmark, Spain, Finland, France, United Kingdom, French Guiana, Guadeloupe, Greece, Hong Kong, Indonesia, Ireland, India, Iceland, Italy, Jordan, Kenya, Kuwait, Lebanon, Latvia, Libya, Morocco, Martinique, Malta, Mexico, Malaysia, New Caledonia, Nigeria, Netherlands, Norway, New Zealand, Philippines, Poland, Portugal, Qatar, R¿union, Russian Federation, Saudi Arabia, Sweden, Singapore, Slovenia, Thailand, and South Africa.

Product description

Philips Respironics V60 Ventilator with Software Versions 3.00 and 3.10. The Respironics V60/V60 Plus Ventilator is an assist ventilator and is intended to augment patient breathing. It is intended for spontaneously breathing individuals who require mechanical ventilation: patients with respiratory failure, chronic respiratory insufficiency, or obstructive sleep apnea in a hospital or other institutional settings under the direction of a physician.

Code information (models, lots, serial numbers)

Show the full code information (36k characters)

Model V60 Ventilator: P/N 1053613, 1053614, 1053615, 1053616, 1053617, DU1053617, U1053617, 1076716, 1137292  US System Serial Number/UDI Code: 100351272 (01)00884838020054(21)100351272 100006691  100193211  100220566 (01)00884838020054(21)100220566 100313757 (01)00884838020054(21)100313757 100313412 (01)00884838020054(21)100313412 100080135  100081569  100110187  100121703  100013472  100230650 (01)00884838020054(21)100230650 100374411 (01)00884838020054(21)100374411 100054305  100068863  100188390  100181283  100030036  100030039  100030046  100030040  100043349  100030041  100037246  201029401 (01)00884838020054(21)201029401 100278366 (01)00884838020054(21)100278366 100088340  100131320  201030489 (01)00884838020054(21)201030489 100057026  100074563  201005638  201001514  100027860  201005635  100154662  100056752  100130449  100154846  100078004  100055123  100191452  100077709  100015080  100191788  100129710  100028046  100077079  201017909  100224361 (01)00884838020054(21)100224361 100043830  100024302  100133355  100077493  100077503  100039424  100177846  201011206  100076877  100191271  100152171  100056385  100120289  100262177 (01)00884838020054(21)100262177 100120262  100047561  100156074  100104165  100115189  201026331 (01)00884838020054(21)201026331 100057549  100133002  100258925 (01)00884838020054(21)100258925 201030702 (01)00884838020054(21)201030702 100040748  100290605 (01)00884838020054(21)100290605 100044617  100147649  100038426  100049697  100318142 (01)00884838020054(21)100318142 100062636  100047178  100130954  100017106  100172675  100265434 (01)00884838020054(21)100265434 100097540  100174538  100028947  201033184 (01)00884838020054(21)201033184 100158623  100184938  100184879  100094731  100325804 (01)00884838020054(21)100325804 100117611  100048641  100267973 (01)00884838020054(21)100267973 100148068  100416709 (01)00884838020054(21)100416709 100118800  100177855  100023735  100025030  100040300  100279146 (01)00884838020054(21)100279146 100008228  100054315  100008224  100008438  100258476 (01)00884838020054(21)100258476 100008146  100006219  100063484  100074097  100349391 (01)00884838020054(21)100349391 100064083  100187263  100168073  100280881 (01)00884838020054(21)100280881 100256575 (01)00884838020054(21)100256575 100245312 (01)00884838020054(21)100245312 100038024  100088492  100017237  100039542  100245557 (01)00884838020054(21)100245557 100275362 (01)00884838020054(21)100275362 100053364  100190401  100237792 (01)00884838020054(21)100237792 100059977  100360168 (01)00884838020054(21)100360168 100268194 (01)00884838020054(21)100268194 100147646  100063411  201024993 (01)00884838020054(21)201024993 100112558  100113474  100266415 (01)00884838020054(21)100266415 100031093  100037777  100051583  100105808  100411005 (01)00884838020054(21)100411005 100148200  100376334 (01)00884838020054(21)100376334 100057733  100091066  100122067  100073597  100057124  100146619  100302964 (01)00884838020054(21)100302964 100091075  100120976  100147661  100032534  MX0000332 (01)00884838020054(21)MX0000332 201028104 (01)00884838020054(21)201028104 100130950  100360674 (01)00884838020054(21)100360674 100094723  201032131 (01)00884838020054(21)201032131 100347825 (01)00884838020054(21)100347825 100043722  100114570  100266417 (01)00884838020054(21)100266417 100043812  100312855 (01)00884838020054(21)100312855 100066482  100066913  100057973  100240908 (01)00884838020054(21)100240908 100360608 (01)00884838020054(21)100360608 100132982  100172584  100105301  MX0000811 (01)00884838020054(21)MX0000811 100049922  201014454  100045057  100119310  100279355 (01)00884838020054(21)100279355 100256638 (01)00884838020054(21)100256638 100379488 (01)00884838020054(21)100379488 100230368 (01)00884838020054(21)100230368 100012543  100058126  100095885  100072279  100074017  100028855  100272829 (01)00884838020054(21)100272829 100027744  100015201  100058120  100134489  100065687  100192989  100266252 (01)00884838020054(21)100266252 201030363 (01)00884838020054(21)201030363 100080380  100060798  100010101  100073350  100050880  201019491  100034592  100059844  100292155 (01)00884838020054(21)100292155 100065690  100240306 (01)00884838020054(21)100240306 100092442  100084643  100063471  100091560  100184278  201008243  100051577  100292954 (01)00884838020054(21)100292954 100095009  100092540  100170409  100251069 (01)00884838020054(21)100251069 100184876  201022432 (01)00884838020054(21)201022432 100249314 (01)00884838020054(21)100249314 100188585  100141803  100028067  100265833 (01)00884838020054(21)100265833 100112787  100238524 (01)00884838020054(21)100238524 100023629  100191117  100023474  100060967  100066569  100148673  100092302  100042111  201029330 (01)00884838020054(21)201029330 100094822  100150330  100084010  100113272  100378839 (01)00884838020054(21)100378839 100104483  100043449  100082989  100132958  100090020  100031686  100029521  100084770  100042021  100080984  100091092  100096788  100175175  100235999 (01)00884838020054(21)100235999 201029470 (01)00884838020054(21)201029470 100015985  100162248  100235912 (01)00884838020054(21)100235912 100168717  100284772 (01)00884838020054(21)100284772 100082777  100071612  100385452 (01)00884838020054(21)100385452 100029517  100319179 (01)00884838020054(21)100319179 100035364  100141536  100033471  100344538 (01)00884838020054(21)100344538 201000384  100330907 (01)00884838020054(21)100330907 100209178 (01)00884838020054(21)100209178 100229296 (01)00884838020054(21)100229296 201019552  100185513  100186030  100047820  100009102  100015661  100251121 (01)00884838020054(21)100251121 100283321 (01)00884838020054(21)100283321 100005114  100066315  100292966 (01)00884838020054(21)100292966 100146517  100267865 (01)00884838020054(21)100267865 201026954 (01)00884838020054(21)201026954 100043440  100037836  100060802  100038505  100078562  100037657  100031685  100147672  201035397 (01)00884838020054(21)201035397 100258290 (01)00884838020054(21)100258290 100108292  100118814  100101557  100057971  100302866 (01)00884838020054(21)100302866 100041202  100058679  100057990  201030455 (01)00884838020054(21)201030455 100054197  100029524  100012045  100073413  100063498  100081182  100081568  201019506  100072280  100065053  100303070 (01)00884838020054(21)100303070 100239144 (01)00884838020054(21)100239144 100193608  100031689  100143204  100034694  100179762  100273519 (01)00884838020054(21)100273519 100077060  100048293  100071616  201033026 (01)00884838020054(21)201033026 100066726  100059847  100162238  100062768  100088753  100101568  100056764  100033470  100097539  100035601  100030461  100105810  100174144  100063885  100090005  100103817  100151611  201017722  100120829  100308332 (01)00884838020054(21)100308332 100258815 (01)00884838020054(21)100258815 100028566  100005396  100080859  100105828  100061374  100227575 (01)00884838020054(21)100227575 100366009 (01)00884838020054(21)100366009 100356232 (01)00884838020054(21)100356232 100357485 (01)00884838020054(21)100357485 100044776  100043442  100029669  100351203 (01)00884838020054(21)100351203 100356963 (01)00884838020054(21)100356963 100236545 (01)00884838020054(21)100236545 100246393 (01)00884838020054(21)100246393 201025663 (01)00884838020054(21)201025663 100099969  MX0000233 (01)00884838020054(21)MX0000233 MX0000368 (01)00884838020054(21)MX0000368 100079012  100278291 (01)00884838020054(21)100278291 100079000  100099766  100038336  100048253  100035738  100023434  100041906  100415118 (01)00884838020054(21)100415118 100064441  100312749 (01)00884838020054(21)100312749 100120430  100148191  100365899 (01)00884838020054(21)100365899 100242919 (01)00884838020054(21)100242919 100090018  100150604  100278299 (01)00884838020054(21)100278299 100045096  100236459 (01)00884838020054(21)100236459 201002834  100301490 (01)00884838020054(21)100301490 100187690  100150052  100041751  100009973  100344568 (01)00884838020054(21)100344568 100051744  100049919  100175176  100121323  100012861  100161316  100092045  100057962  100334852 (01)00884838020054(21)100334852 100058800  100126907  100246068 (01)00884838020054(21)100246068 100150665  100271325 (01)00884838020054(21)100271325 100093992  100130367  100288718 (01)00884838020054(21)100288718 100230446 (01)00884838020054(21)100230446 100021591  201028355 (01)00884838020054(21)201028355 100125239  100040868  100065834  100135922  100253808 (01)00884838020054(21)100253808 100087926  100051222  100150324  100074107  100042475  100121152  100286599 (01)00884838020054(21)100286599 100067204  100072285  100043350  100171351  100095332  100114921  100057861  100185673  100266685 (01)00884838020054(21)100266685 100293443 (01)00884838020054(21)100293443 MX0000238 (01)00884838020054(21)MX0000238 100331096 (01)00884838020054(21)100331096 100331180 (01)00884838020054(21)100331180 100240670 (01)00884838020054(21)100240670 100293958 (01)00884838020054(21)100293958 100088708  100275526 (01)00884838020054(21)100275526 100143160  100275236 (01)00884838020054(21)100275236 100194038 (01)00884838020054(21)100194038 100151427  100168250  100168622  100135783  100195058 (01)00884838020054(21)100195058 100188258  100064686  100059973  100005495  100043439  100295535 (01)00884838020054(21)100295535 100086410  100085637  100006536  100116545  100136237  100321031 (01)00884838020054(21)100321031 100179120  100164145  100121422  100146430  100062493  100299044 (01)00884838020054(21)100299044 100194457  100048295  100082338  100038502  100280731 (01)00884838020054(21)100280731 100101547  100029856  100072198  100062868  100331195 (01)00884838020054(21)100331195 100047308  100187253  100379299 (01)00884838020054(21)100379299 100037769  100062986  201009743  100047183  100261542 (01)00884838020054(21)100261542 100068667  100047288  100088497  100041912  100229287 (01)00884838020054(21)100229287 100134980  100134790  100172336  100057010  100186433  100097542  100064438  100021504  100045032  100317647 (01)00884838020054(21)100317647 100034160  100021836  100159749  100047287  100249025 (01)00884838020054(21)100249025 100187691  100267319 (01)00884838020054(21)100267319 100031559  100153032  100313539 (01)00884838020054(21)100313539 100006213  100171610  201022885 (01)00884838020054(21)201022885 100040173  201025712 (01)00884838020054(21)201025712 100109734  100232319 (01)00884838020054(21)100232319 100121691  100147641  100238964 (01)00884838020054(21)100238964 100170127  100099354  100255853 (01)00884838020054(21)100255853 100223794 (01)00884838020054(21)100223794 100063393  100089004  100226064 (01)00884838020054(21)100226064 100082340  100010317  100088401  100155080  201014575  201024359 (01)00884838020054(21)201024359 100149437  100409568 (01)00884838020054(21)100409568 201024303 (01)00884838020054(21)201024303 100086137  100122252  100223642 (01)00884838020054(21)100223642 100037640  201024282 (01)00884838020054(21)201024282 100086194  100028055  100073101  201015094  100118785  100061151  100025790  100069330  100334142 (01)00884838020054(21)100334142 100157691  100185654  100105448  100310876 (01)00884838020054(21)100310876 100301495 (01)00884838020054(21)100301495 100276891 (01)00884838020054(21)100276891 100309098 (01)00884838020054(21)100309098 201014424  100104852  100092307  100181279  100220406 (01)00884838020054(21)100220406 100168705  100006528  100182319  201025673 (01)00884838020054(21)201025673 100054314  100096083  100057539  100238137 (01)00884838020054(21)100238137 100192675  100120651  100172589  100180562  100023417  100023232  100299584 (01)00884838020054(21)100299584 100030827  100261747 (01)00884838020054(21)100261747 100272460 (01)00884838020054(21)100272460 201017686  100063013  100218795 (01)00884838020054(21)100218795 100012694  100047381  100047382  100062883  100088746  100324163 (01)00884838020054(21)100324163 201027220 (01)00884838020054(21)201027220 100034697  100034995  100086615  100049797  100224968 (01)00884838020054(21)100224968 100080993  201008497  100044788  100297480 (01)00884838020054(21)100297480 100070679  201000381  100014079  201027348 (01)00884838020054(21)201027348 100118787  100152929  100135503  100158627  100086650  100033326  100027739  100066854  100156075  100321931 (01)00884838020054(21)100321931 100377146 (01)00884838020054(21)100377146 100230719  100119582  201031729 (01)00884838020054(21)201031729 100011302  100141943  100006603  100023238  100330574 (01)00884838020054(21)100330574 100047307  100155060  100034979  100073835  100341094 (01)00884838020054(21)100341094 100157491  100016355  100065697  100262209 (01)00884838020054(21)100262209 100028482  100117761  100006941  100026966  100066178  100026828  100028476  100027422  100024475  100085371  100113868  201034495 (01)00884838020054(21)201034495 100409926 (01)00884838020054(21)100409926 MX0001199 (01)00884838020054(21)MX0001199 100113860  100101195  100313756 (01)00884838020054(21)100313756 100295529 (01)00884838020054(21)100295529 100093789  100246120 (01)00884838020054(21)100246120 100066581  100297483 (01)00884838020054(21)100297483 100066100  100135284  100121708  100086648  100073509  100113054  100032482  100275980 (01)00884838020054(21)100275980 100164291  100064784  100048647  100227589 (01)00884838020054(21)100227589 100027749  100114949  100087819  100334369 (01)00884838020054(21)100334369 100382512 (01)00884838020054(21)100382512 100320594 (01)00884838020054(21)100320594 100313743 (01)00884838020054(21)100313743 100120822  100378826 (01)00884838020054(21)100378826 100059956  100141152  100240716 (01)00884838020054(21)100240716 100246461 (01)00884838020054(21)100246461 100152770  100248825 (01)00884838020054(21)100248825 100131669  100235921 (01)00884838020054(21)100235921 100088608  100068592  100004546  100075675  100048498  MX0001029 (01)00884838020054(21)MX0001029 100029840  100094726  100073833  100054303  100063004  100236027 (01)00884838020054(21)100236027 100069013  100050757  100043431  100042570  100062357  100218811 (01)00884838020054(21)100218811 100265685 (01)00884838020054(21)100265685 100083739  100150664  100340614 (01)00884838020054(21)100340614 100004997  100112866  100004542  100047719  100044803  100390066 (01)00884838020054(21)100390066 201010452  100106893  100302368 (01)00884838020054(21)100302368 100032471  100359425 (01)00884838020054(21)100359425 100241825 (01)00884838020054(21)100241825 100048256  100095120  100283162 (01)00884838020054(21)100283162 100013791  100094890  100092434  100195050 (01)00884838020054(21)100195050 100302583 (01)00884838020054(21)100302583 100318179 (01)00884838020054(21)100318179 100080061  100172522  201002590  100044618  100062882  100049352  100232302 (01)00884838020054(21)100232302 201021064  100084013  100059724  100028470  100154215  100270684 (01)00884838020054(21)100270684 100176309  100131473  100304029 (01)00884838020054(21)100304029 100024299  100246582 (01)00884838020054(21)100246582 100066578  100065703  201016744  100143128  201025704 (01)00884838020054(21)201025704 201025598 (01)00884838020054(21)201025598 201010639  100351240 (01)00884838020054(21)100351240 100102513  100235909 (01)00884838020054(21)100235909 100014282  100041757  100033455  100075690  100246581 (01)00884838020054(21)100246581 100023721  100042671  100149602  100009517  100130526  100074092  100142048  100004031  100302580 (01)00884838020054(21)100302580 100064333  100041755  100042481  100007972  100178181  100148066  100045099  100058794  100255972 (01)00884838020054(21)100255972 201002796  100283355 (01)00884838020054(21)100283355 201000837  100008138  100165440  100121320  100112871  100094735  201002835  100091090  100082096  100333811 (01)00884838020054(21)100333811 100119209  100055120  100185262  100299287 (01)00884838020054(21)100299287 100356940 (01)00884838020054(21)100356940 100040181  201024768 (01)00884838020054(21)201024768 100143208  100118802  100009939  100050812  100298084 (01)00884838020054(21)100298084 100191144  100331194 (01)00884838020054(21)100331194 100349414 (01)00884838020054(21)100349414 100303494 (01)00884838020054(21)100303494 100265837 (01)00884838020054(21)100265837 100095092  100005599  100109421  100024832  100028344  100051618  100041075  100072195  100194448  100094936  100148677  100337476 (01)00884838020054(21)100337476 100150328  100071648  100266407 (01)00884838020054(21)100266407 100275754 (01)00884838020054(21)100275754 100148006  100158343  100016751  100258818 (01)00884838020054(21)100258818 100231061 (01)00884838020054(21)100231061 100186424  201001645  MX0001033 (01)00884838020054(21)MX0001033 100133093  100011328  100113271  100379344 (01)00884838020054(21)100379344 100093878  100016528  201031330 (01)00884838020054(21)201031330 100131170  100051574  201001686  100303800 (01)00884838020054(21)100303800 100149365  201034533 (01)00884838020054(21)201034533 100059952  100289913 (01)00884838020054(21)100289913 100261402 (01)00884838020054(21)100261402 100011271  100321034 (01)00884838020054(21)100321034 100093097  201031096 (01)00884838020054(21)201031096 100330563 (01)00884838020054(21)100330563 100321411 (01)00884838020054(21)100321411 100104724  201019556  100161306  100313201 (01)00884838020054(21)100313201 100319184 (01)00884838020054(21)100319184 201014577  100263575 (01)00884838020054(21)100263575 100174647  100091076  100045118  100040321  100074501  100104746  100029684  100099348  100162182  100019586  100072853  100049289  100104176  100041382  100219461 (01)00884838020054(21)100219461 100030612  100194307 (01)00884838020054(21)100194307 100141908  100146616  100114574  100103903  100098328  201015055  100008064  100080055  100184935  100141151  100321030 (01)00884838020054(21)100321030 100151261  100094893  201002554  100376321 (01)00884838020054(21)100376321 100129531  100100458  100021832  100271329 (01)00884838020054(21)100271329 100032480  100313205 (01)00884838020054(21)100313205 100317516 (01)00884838020054(21)100317516 100016741  201014801  100308413 (01)00884838020054(21)100308413 100084648  100072762  100047833  100093241  100037840  100030457  100079232  201026061 (01)00884838020054(21)201026061 100042652  100160324  100038601  100028868  201013195  100160571  201022517 (01)00884838020054(21)201022517 100017108  100141905  100267095 (01)00884838020054(21)100267095 100042239  100094899  100041918  100065811  100301611 (01)00884838020054(21)100301611 100059963  100165031  100038517  201019540  100313200 (01)00884838020054(21)100313200 201009840  201019632  100069024  100104379  100109740  100101669  100095636  100065162  100246095 (01)00884838020054(21)100246095 100226211 (01)00884838020054(21)100226211 100069460  100101552  100102516  100035611  100093715  100096086  100035435  100066486  100182917  100084209  100093899  100287018 (01)00884838020054(21)100287018 100080132  100172596  100356227 (01)00884838020054(21)100356227 100109731  201021047  100263576 (01)00884838020054(21)100263576 100088719  100158858  100048646  100087729  100009516  100075678  100120692  201018025  100168616  100009560  MX0001282 (01)00884838020054(21)MX0001282 100184947  100073400  100026608  100131419  100283908 (01)00884838020054(21)100283908 100086414  100150050  100120436  100193230  100023426  100231683 (01)00884838020054(21)100231683 100006702  100263123 (01)00884838020054(21)100263123 100162173  100012868  100147666  100366526 (01)00884838020054(21)100366526 100032466  100152913  100405716 (01)00884838020054(21)100405716 100085041  100133402  100012013  100097820  100048191  100141127  100146937  100325532 (01)00884838020054(21)100325532 100122265  100109638  100266148 (01)00884838020054(21)100266148 100303075 (01)00884838020054(21)100303075 100275316 (01)00884838020054(21)100275316 100113480  100094902  100149266  201015963  100039431  100038926  100188504  100039407  100188604  100034823  100098502  100072440  100272251 (01)00884838020054(21)100272251 100254975 (01)00884838020054(21)100254975 100295496 (01)00884838020054(21)100295496 100255292 (01)00884838020054(21)100255292 201026729 (01)00884838020054(21)201026729 100358924 (01)00884838020054(21)100358924 201000812  100090444  100364111 (01)00884838020054(21)100364111 100188448  100071618  100295209 (01)00884838020054(21)100295209 201029414 (01)00884838020054(21)201029414 100179763  100059719  100130366  100063470  201026989 (01)00884838020054(21)201026989 100031668  100141800  100237788 (01)00884838020054(21)100237788 100085816  100082652  201017931  100112564  201017003  100064349  100177838  100126519  100060789  100341090 (01)00884838020054(21)100341090 201001581  100182554  100043824  100238926 (01)00884838020054(21)100238926 100030481  100126512  201035555 (01)00884838020054(21)201035555 100025991  100258929 (01)00884838020054(21)100258929 100072203  100042033  100063898  100081325  100021509  100193625  100308031 (01)00884838020054(21)100308031 100092035  100026943  201025122 (01)00884838020054(21)201025122 100065168  100271321 (01)00884838020054(21)100271321 100337334 (01)00884838020054(21)100337334 100265704 (01)00884838020054(21)100265704 100112656  100295495 (01)00884838020054(21)100295495 100240321 (01)00884838020054(21)100240321 100038802  100067313  100179135  100102211  100102210  100086960  100066312  100303488 (01)00884838020054(21)100303488 100037046  100219469 (01)00884838020054(21)100219469 100007639  100160180  100185636  100236227 (01)00884838020054(21)100236227 100255157 (01)00884838020054(21)100255157 100057739  100066441  100081565  100074008  100098850  100061143  100369254 (01)00884838020054(21)100369254 100245567 (01)00884838020054(21)100245567 100230208 (01)00884838020054(21)100230208 100230214 (01)00884838020054(21)100230214 100021949  100236039 (01)00884838020054(21)100236039 100040164  100101197  100330231 (01)00884838020054(21)100330231 100193008  100109283  100060791  100133416  100095557  100041758  100050633  201003311  100113086  100182908  100094536  100100485  100312925 (01)00884838020054(21)100312925 100195186 (01)00884838020054(21)100195186 100380041 (01)00884838020054(21)100380041 100224094 (01)00884838020054(21)100224094 100055003  100042116  100224977 (01)00884838020054(21)100224977 100246093 (01)00884838020054(21)100246093 100028851  100040320  100014423  100048204  201029993 (01)00884838020054(21)201029993 100064423  100177871  100056772  100118886  100042650  100054201  100060709  100120425  201027365 (01)00884838020054(21)201027365 100028944  100273497 (01)00884838020054(21)100273497 100028950  100034411  100082333  100135924  100035427  100157487  100007041  100091067  100031678  100109218  100057998  100037768  100222615 (01)00884838020054(21)100222615 100087268  100029519  100029678  100102514  100029509  201015058  100029690  100227577 (01)00884838020054(21)100227577 100299596 (01)00884838020054(21)100299596 100229562 (01)00884838020054(21)100229562 100140437  100230444 (01)00884838020054(21)100230444 100048636  100066307  100057871  100186214  100010165  100172524  100109736  100112659  201008826  100010382  100250719 (01)00884838020054(21)100250719 100181999  100182620  100331541 (01)00884838020054(21)100331541 100043346  100154405  100017113  100245839 (01)00884838020054(21)100245839 100004085  100035421  100240305 (01)00884838020054(21)100240305 100267137 (01)00884838020054(21)100267137 100034384  100308633 (01)00884838020054(21)100308633 100093248  100191900  100010517  100082082  100081555  100034700  100093794  100035829  100088750  100081556  201027069 (01)00884838020054(21)201027069 100082085  100044785  100044624  100321103 (01)00884838020054(21)100321103 100081191  100310588 (01)00884838020054(21)100310588 100034684  100105945  100066842  100158398  100066228  201035577 (01)00884838020054(21)201035577 100240130 (01)00884838020054(21)100240130 100105302  100182589  100158467  201003072  100091038  100290548 (01)00884838020054(21)100290548 201022554 (01)00884838020054(21)201022554 100157685  100157507  100149587  100323688 (01)00884838020054(21)100323688 100188860  100042468  100057963  100052079  100265436 (01)00884838020054(21)100265436 100246127 (01)00884838020054(21)100246127 100072843  100026620  100073353  100168982  100135274  100039408  201016621  100148657  100259394 (01)00884838020054(21)100259394 100109268  100223781 (01)00884838020054(21)100223781 100372400 (01)00884838020054(21)100372400 100032536  100068393  201003183  100295215 (01)00884838020054(21)100295215 100082095  100249308 (01)00884838020054(21)100249308 100072281  100049349  100090296  100005141  100069116  100049908  100019431  100230486 (01)00884838020054(21)100230486 100124740  100281722 (01)00884838020054(21)100281722 100051223  100184907  100052280  100087275  100113092  100068382  100114937  100093098  100142052  100131425  100075573  100098762  100042110  100061150  100288829 (01)00884838020054(21)100288829 100119698  100121832  100130800  100130268  201014657  100117224  100117200  100175403  100062635  100023728  201011043  100043833  100117740  100134874  100076883  100190707  100068572  201009846  100094051  100060696  100027878  100170121  100170109  100159165  100069117  100045031  100073348  100102759  100327223 (01)00884838020054(21)100327223 201032016 (01)00884838020054(21)201032016 100194710 (01)00884838020054(21)100194710 100302369 (01)00884838020054(21)100302369 100030460  100095334  100141507  100034039  100063602  100008649  100275527 (01)00884838020054(21)100275527 100057523  100066318  100037651  100085374  100096873  100085363  100037832  100097609  100172587  100037833  201009784  100149311  201006557  100067202  100034162  100047722  100085375  100379369 (01)00884838020054(21)100379369 100085367  100295633 (01)00884838020054(21)100295633 100299585 (01)00884838020054(21)100299585 100272103 (01)00884838020054(21)100272103 100250575 (01)00884838020054(21)100250575 100062642  100300900 (01)00884838020054(21)100300900 100299448 (01)00884838020054(21)100299448 100289849 (01)00884838020054(21)100289849 100031667  100066905  100187244  100030834  201038152 (01)00884838020054(21)201038152 100263673 (01)00884838020054(21)100263673 100299300 (01)00884838020054(21)100299300 100006365  100037868  100131483  100073175  201001308  100048648  100047181  100322246 (01)00884838020054(21)100322246 100280732 (01)00884838020054(21)100280732 100062995  100331190 (01)00884838020054(21)100331190 100141745  201029415 (01)00884838020054(21)201029415 100306793 (01)00884838020054(21)100306793 100157681  100063415  100253619 (01)00884838020054(21)100253619 201022503 (01)00884838020054(21)201022503 100184877  201026966 (01)00884838020054(21)201026966 100392865 (01)00884838020054(21)100392865 100010514  100114580  100263197 (01)00884838020054(21)100263197 100316710 (01)00884838020054(21)100316710 100161352  100240668 (01)00884838020054(21)100240668 100057022  100149316  100140073  100288399 (01)00884838020054(21)100288399 100066474  100024472  100029500  100065698  100025788  100251067 (01)00884838020054(21)100251067 100043732  201032024 (01)00884838020054(21)201032024 201032088 (01)00884838020054(21)201032088 100101277  100283104 (01)00884838020054(21)100283104 100194732 (01)00884838020054(21)100194732 100081558  100047388  100050003  100016012  100118898  100020219  100141923  100048278  100062890  100121707  100064991  100110114  100065228  100041749  100042392  100371791 (01)00884838020054(21)100371791 100112577  201034617 (01)00884838020054(21)201034617 201034622 (01)00884838020054(21)201034622 201019550  100148007  100065700  100031000  100240662 (01)00884838020054(21)100240662 MX0001821 (01)00884838020054(21)MX0001821 100303069 (01)00884838020054(21)100303069 100103239  100113098  201029515 (01)00884838020054(21)201029515 100093442  100068595  201030025 (01)00884838020054(21)201030025 100092059  100273693 (01)00884838020054(21)100273693 100158649  100193806  100427792  100121330  100006895  100266679 (01)00884838020054(21)100266679 100107656  100368889 (01)00884838020054(21)100368889  OUS (International) System Serial Number/UDI Code: 100410528 (01)00884838089297(21)100410528 100342563  100261002 (01)00884838009851(21)100261002 100391108 (01)00884838089297(21)100391108 100345521  100027294  201033669 (01)00884838009851(21)201033669 100078694  100029332  100378456 (01)00884838009851(21)100378456 100143739  100144969  100243519 (01)00884838009851(21)100243519 100341392 (01)00884838009851(21)100341392 201029945 (01)00884838009851(21)201029945 100221160 (01)00884838009851(21)100221160 100368048 (01)00884838009851(21)100368048 100071881  100132030  100087603  100345113  100343567  100026763  100343580  100125173  100257206 (01)00884838009851(21)100257206 100020049  100250344 (01)00884838025776(21)100250344 100142326  201029762 (01)00884838009844(21)201029762 100170107  100021689  100060108  100169767  100030132  100248621 (01)00884838009851(21)100248621 100026671  100230020 (01)00884838009844(21)100230020 100094352  100406417 (01)00884838089297(21)100406417 100027974  100103663  201004367  100256852 (01)00884838009851(21)100256852 100158086  100267691 (01)00884838009851(21)100267691 100132108  100004667  100007329  100036438  100352131 (01)00884838009851(21)100352131 100175815  100139917  100189486  100071130  100172257  201013817  201002709  100078696  100003246  100086773  100118670  100088051  100189489  201002273  100247417 (01)00884838009851(21)100247417 100346921 (01)00884838009851(21)100346921 100176338  100278791 (01)00884838009851(21)100278791 100285001 (01)00884838009851(21)100285001 100190085  100006111  100230029 (01)00884838009851(21)100230029 100244201 (01)00884838009844(21)100244201 100179385  100239534 (01)00884838009851(21)100239534 201029776 (01)00884838009851(21)201029776 100110677  100243224 (01)00884838009851(21)100243224 100348087 (01)00884838009851(21)100348087 100144874  100179382  100229633 (01)00884838009844(21)100229633 100378459 (01)00884838089297(21)100378459 100410514  100315894  100126161  100133620  100094393  100144981  100230026 (01)00884838009851(21)100230026 100233964 (01)00884838009851(21)100233964 100015334  201022999 (01)00884838009851(21)201022999 100244590 (01)00884838009851(21)100244590 100028670  201029115 (01)00884838009851(21)201029115 100006114  100118695  100176847  100110846  100371670 (01)00884838009851(21)100371670 100189477  100189480  100078367  100169650  100172258  100381925 (01)00884838009851(21)100381925 201010821  100166085  100043299  100045681  201029918 (01)00884838009851(21)201029918 100234781 (01)00884838009851(21)100234781 100219997 (01)00884838009851(21)100219997 100082433  100104065  100024697  100011191  100259879 (01)00884838009851(21)100259879 201033640 (01)00884838009851(21)201033640 100237375 (01)00884838009851(21)100237375 100247101 (01)00884838009851(21)100247101 100289690  100189819  100254748 (01)00884838009851(21)100254748 100116378  201033644 (01)00884838009851(21)201033644 100237374 (01)00884838009844(21)100237374 201002710  100104055  100278758 (01)00884838009851(21)100278758 100145123  100232758 (01)00884838009851(21)100232758 100055938  100395579 (01)00884838089297(21)100395579 201010183  100237630 (01)00884838009844(21)100237630 201001547  100124977  100124980  201007601  100315184 (01)00884838009851(21)100315184 201010864  100176022  100046415  100092492  100260037 (01)00884838009851(21)100260037 100186765  100043298  100152333  100163590  100039840  100278756 (01)00884838009851(21)100278756 100110675  100233967 (01)00884838009851(21)100233967 100259870 (01)00884838009851(21)100259870 100274760 (01)00884838009851(21)100274760 100351158  100081755  100279007 (01)00884838009851(21)100279007 100045726  100135544  100257671 (01)00884838009851(21)100257671 100219976 (01)00884838009851(21)100219976 100097395  201031754 (01)00884838009851(21)201031754 100170696  100070576  100144980  100029455  100174449  100029075  100139896  100136805  100110672  100243918 (01)00884838009851(21)100243918 100226891 (01)00884838009851(21)100226891 100388081 (01)00884838009851(21)100388081 100219290 (01)00884838009851(21)100219290 100015329  100026097  100027970  100239303 (01)00884838009851(21)100239303 100138231  100406482 (01)00884838009851(21)100406482 100097980  100292345 (01)00884838009851(21)100292345 100415676  201029884 (01)00884838009851(21)201029884 100277916 (01)00884838009851(21)100277916 100092187  100348161 (01)00884838009844(21)100348161 100029460  100055947  100312444 (01)00884838009851(21)100312444 100286657 (01)00884838009851(21)100286657 100026680  201032697 (01)00884838009851(21)201032697 201032759 (01)00884838009851(21)201032759 100152551  100282843 (01)00884838009851(21)100282843 100369435 (01)00884838009851(21)100369435 100341862 (01)00884838089297(21)100341862 100089812  100190080  100007194  100026666  100343575  100007411  100290662 (01)00884838009851(21)100290662 100019394  100343690  100406654 (01)00884838089297(21)100406654 100378670 (01)00884838089297(21)100378670 100259999 (01)00884838009851(21)100259999 100358186  100345848 (01)00884838089297(21)100345848 100375915 (01)00884838089297(21)100375915 100003612  100343639  100345599  100345547 (01)00884838089297(21)100345547 100042753  100343610  100343599  100037544  100314705  100345527  100354459  100375828  100343611  100404856 (01)00884838089297(21)100404856 100375832  100003339  100251169 (01)00884838009844(21)100251169 100247260 (01)00884838009851(21)100247260 100011042  100011044  100266744 (01)00884838009844(21)100266744 100003160  100271022 (01)00884838009851(21)100271022 100029459  100003243  100456025 (01)00884838009851(21)100456025 100064711  100055828  100013968  100026094  100137503  100247108 (01)00884838009851(21)100247108 100252969 (01)00884838009851(21)100252969 100189485  201002380  100065597  100037555  100055935  100231067 (01)00884838009851(21)100231067 100149705  100098083  100159650  100274445 (01)00884838009851(21)100274445 100004673  100045725  100160838  100065594  100102051  100350166 (01)00884838009851(21)100350166 100190008  100027301  100114025  100164088  100039849  100346973 (01)00884838009851(21)100346973 100219281 (01)00884838009851(21)100219281 100057375  100237625 (01)00884838009844(21)100237625 100110942  100100996  100368016 (01)00884838009851(21)100368016 100095586  100080824  100319297 (01)00884838025776(21)100319297 201027752 (01)00884838025776(21)201027752 100333778 (01)00884838025776(21)100333778 100125719  100120746  100289711  100081706  100226369 (01)00884838009851(21)100226369 100005815  100315391 (01)00884838009844(21)100315391 100252774 (01)00884838009844(21)100252774 201032861 (01)00884838009851(21)201032861 100176321  100357773 (01)00884838009851(21)100357773 100132408  100092646  100174964  100005813  201029728 (01)00884838009844(21)201029728 100026665  100063151  100274692 (01)00884838009851(21)100274692 100169364  100282846 (01)00884838009851(21)100282846 100239390 (01)00884838009851(21)100239390 100286304 (01)00884838009851(21)100286304 100231171 (01)00884838009844(21)100231171 100021402  100278998 (01)00884838009851(21)100278998 100279463 (01)00884838009851(21)100279463 100019389  100013971  100237192 (01)00884838009851(21)100237192 100149718  100060588  100020745  100361234 (01)00884838009851(21)100361234 100237302 (01)00884838009851(21)100237302 100174967  100391367 (01)00884838009844(21)100391367 100252765 (01)00884838009851(21)100252765 201029778 (01)00884838009851(21)201029778 100082434  201010835  100064708  100367183 (01)00884838009851(21)100367183 100114463  201001571  100029469  100030113  100053793  100069521  100014773  100084821  100228994 (01)00884838009851(21)100228994 100163725  100277890 (01)00884838009851(21)100277890 201015439  201029822 (01)00884838009851(21)201029822 201025393 (01)00884838009851(21)201025393 201025923 (01)00884838009851(21)201025923 100071127  100415765 (01)00884838009851(21)100415765

Devices named in this recall

4 devices in the FDA device registry whose identifier appears in the recall's code information.