EnVeo PRO recallZ-2042-2021 · Medtronic Heart Valves Division · initiated Jun 11, 2021
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Medtronic Heart Valves Division or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Due to reports of actuator separation for various delivery catheter systems. Affected products were manufactured before July 11, 2020.
FDA root cause category: Device Design
What the company is doing
On 06/09/2021, the Recalling Firm sent an "URGENT MEDICAL DEVICE RECALL" Notification via UPS 2-day (US Customers) and regionally approved methods (OUS Customer) to customers informing them that it has been identified that were is the potential for actuator separation with delivery catheter systems manufactured before 07/11/2020. Customer are instructed to: 1. Identify and quarantine all unused affected product manufactured before 11-JUL-2020.The enclosed Customer Notification Detail Report includes affected product for their facility. 2. Return all unused affected product in their inventory to Recalling Firm. Contact Customer Service at 800-854-3570 to initiate a product return. Their local Representative can assist them as necessary in initiating the return of this product. 3. Complete the enclosed Customer Confirmation Form and email it to [email protected]. 4. Forward this notice to all who need to be aware within their organization and to any organization where the affected product may have been transferred. For any questions or assistance, contact Customer Service at 800-854-3570
Recall details
- Recall number
- Z-2042-2021
- Status
- Open, Classified
- Initiated by the company
- Jun 11, 2021
- Recalling firm
- Medtronic Heart Valves Division
- Quantity
- Total of all Delivery Systems (Globally) = 67,339 units
- Product code
- NPT
- Distribution
- U.S.: AL, AK, AR, AZ, CA, CT, DE, DC, FL, GA, HI, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NE, NH, NJ, NM, NV, NY, NC, ND, OH, OK, OR, PA, Puerto Rico, SC, SD, TN, TX, UT, VA, VT, WA, WI, and WV. O.U.S.: Albania, Argentina, Australia, Austria, Belarus, Belgium, Brazil, BULGARIA, Canada, Chile, CHINA, Colombia, COSTA RICA, Croatia, Curacao, CYPRUS, Czech Republic, Denmark, DOMINICAN REPUBLIC, Egypt, ESTONIA, FINLAND, FRANCE, FRENCH POLYNESIA, GEORGIA, Germany, Greece, GUATEMALA, HONDURAS, Hong Kong, Hungary, ICELAND, India, INDONESIA, IRAN (ISLAMIC REPUBLIC OF), Ireland, Israel, Italy, Japan, JORDAN, KAZAKHSTAN, KOREA, REPUBLIC OF, KUWAIT, LATVIA, LEBANON, LITHUANIA, Luxembourg, MACEDONIA (THE FORMER YUGOSLAV REPUBLIC OF), Malaysia, MALTA, Mexico, MOLDOVA (REPUBLIC OF), Montenegro, Netherlands, New Zealand, Norway, OMAN, Panama, Peru, Poland, Portugal, QATAR, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, SLOVAKIA (Slovak Republic), SLOVENIA, South Africa, Spain, Sweden, Switzerland, SYRIAN ARAB REPUBLIC, Taiwan (Republic of China), THAILAND, Turkey, UNITED ARAB EMIRATES, United Kingdom, URUGUAY, UZBEKISTAN, VENEZUELA and Viet Nam.
Product description
CoreValve Evolut PRO Delivery Catheter System, REF: ENVPRO-14; ENVPRO-16;
Code information (models, lots, serial numbers)
Show the full code information (8k characters)
Model Number/Description: ENVPRO-14-US/18FR DELIV SYS ENVPRO-14-US GTIN:00643169987203 Lot Numbers: 0009679690; 0009688651; 0009690887; 0009693142; 0009716570; 0009718679; 0009724715; 0009735112; 0009747816; 0009747817; 0009769716; 0009780304; 0009780312; 0009813675; 0009819798; 0009833186; 0009833187; 0009835584; 0009884743; 0009903576; 0009909651; 0009911878; 0009916760; 0009674373; Model Number/Description: ENVPRO-14/18FR DELIV SYS ENVPRO-14 GTIN: 00643169987258 Lot Numbers 0009674373; 0009677041; 0009681843; 0009686572; 0009696943; 0009702501; 0009708314; 0009720707; 0009730833; 0009732864; 0009735111; 0009741277; 0009745282; 0009751537; 0009751539; 0009753928; 0009753930; 0009759086; 0009763571; 0009765704; 0009765705; 0009769796; 0009771781; 0009778377; 0009780306; 0009780315; 0009797642; 0009801629; 0009801632; 0009815902; 0009815904; 0009819795; 0009824816; 0009827178; 0009827180; 0009831363; 0009831364; 0009839737; 0009843479; 0009846259; 0009872399; 0009872400; 0009875230; 0009877165; 0009879171; 0009882762; 0009882764; 0009885245; 0009888373; 0009894468; 0009894469; 0009896248; 0009899816; 0009902780; 0009903577; 0009905476; 0009907434; 0009916758; 0009924441; 0009930619; 0009930620; 0009932436; 0009935871; 0009940329; 0009941896; 0009943765; 0009943766; 0009950526; 0009956449; 0009959205; 0009959208; 0009968628; 0009971361; 0009973690; 0009974884; 0009974886; 0009979351; 0009981603; 0009981605; 0009992252; 0009992254; 0009999409; 0010001291; 0010004099; 0010004100; 0010007364; 0010007368; 0010010945; 0010014267; 0010017314; 0010017315; 0010021500; 0010021501; 0010024637; 0010026718; 0010028933; 0010030671; 0010032826; 0010032828; 0010037869; 0010037870; 0010037876; 0010040826; 0010040828; 0010044464; 0010044465; 0010051010; 0010051348; 0010054636; 0010058015; 0010061411; 0010061412; 0010063249; 0010063250; 0010081656; 0010083965; 0010087639; 0010091721; 0010091723; 0010097384; 0010097385; 0010099753; 0010099756; 0010100832; 0010100833; 0010100834; 0010100835; 0010105882; 0010105884; 0010108127; 0010108133; 0010108134; 0010108135; 0010120343; 0010124203; 0010126036; 0010129931; 0010131801; 0010133987; 0010136245; 0010136247; 0010139633; 0010139634; 0010143248; 0010143249; 0010143251; 0010145133; 0010148991; 0010153126; 0010153127; 0010153129; 0010157415; 0010161319; 0010163782; 0010169460; 0010169463; 0010169465; 0010175290; 0010175291; 0010175292; 0010184723; 0010184724; 0010184725; 0010188540; 0010194379; 0010194380; 0010197789; 0010199885; 0010204027; 0010209066; 0010211940; 0010211941; 0010220455; 0010245329; 0010248119; 0010249598; 0010252313; 0010253923; 0010255302; 0010256429; 0010256431; 0010257814; 0010258779; 0010260159; 0010261391; 0010262934; 0010268205; 0010268207; 0010269513; 0010271607; 0010273066; 0010273067; 0010275572; 0010275573; 0010278032; 0010278033; 0010281654; 0010282945; 0010285882; 0010285884; 0010286952; 0010286953; 0010288572; 0010288573; 0010288575; 0010289863; 0010297581; 0010297582; 0010301885; Model Number/Description: ENVPRO-16-US/20FR DELIV SYS ENVPRO-16-US GTIN:00643169987210 Lot Numbers: 0009674372; 0009677040; 0009679689; 0009681842; 0009688649; 0009688650; 0009688652; 0009693140; 0009693141; 0009693143; 0009700507; 0009702502; 0009708313; 0009710330; 0009716568; 0009716569; 0009716571; 0009718678; 0009722417; 0009722418; 0009724713; 0009724714; 0009729071; 0009730831; 0009732861; 0009732863; 0009737234; 0009737235; 0009741276; 0009743440; 0009751538; 0009751540; 0009753926; 0009753929; 0009759085; 0009771780; 0009774180; 0009778376; 0009780303; 0009780308; 0009780311; 0009780314; 0009788697; 0009788698; 0009801633; 0009810569; 0009810570; 0009811909; 0009813678; 0009819800; 0009821182; 0009822855; 0009823656; 0009829625; 0009833190; 0009837505; 0009839738; 0009839739; 0009839740; 0009845304; 0009867880; 0009867881; 0009872402; 0009877163; 0009879660; 0009880134; 0009909647; 0009909649; 0009914677; 0009916004; 0009916759; 0009916762; 0009916764; 0009922784; 0009981601; 0010041604; 0010148988; 0010172263; 0010172264; Model Number/Description: ENVPRO-16-C/20FR DELIV SYS ENVPRO-16-C GTIN:00643169987319 Lot Numbers: 0009696941; 0009738986; 0009771779; 0009803583; 0009863079; Model Number/Description: ENVPRO-16/20FR DELIV SYS ENVPRO-16 GTIN:00643169987265 Lot Numbers: 0009674371; 0009677038; 0009679691; 0009679692; 0009684327; 0009686569; 0009686570; 0009686571; 0009696942; 0009698697; 0009702500; 0009702503; 0009708312; 0009710331; 0009730832; 0009732862; 0009737236; 0009738987; 0009741278; 0009743439; 0009745281; 0009747814; 0009747815; 0009753924; 0009753927; 0009758219; 0009759084; 0009759087; 0009763374; 0009763376; 0009774179; 0009778375; 0009780302; 0009780305; 0009780307; 0009780310; 0009788694; 0009788696; 0009788699; 0009788700; 0009795297; 0009795298; 0009795299; 0009797643; 0009801628; 0009801631; 0009805724; 0009808236; 0009811908; 0009813676; 0009813677; 0009815903; 0009815905; 0009819799; 0009823654; 0009823655; 0009827179; 0009829369; 0009829370; 0009829624; 0009831669; 0009833189; 0009837504; 0009880131; 0009880135; 0009898071; 0009898072; 0009903578; 0009909648; 0009909650; 0009911877; 0009916005; 0009916761; 0009916763; 0009922783; 0009924443; 0009926465; 0009926466; 0009930616; 0009930617; 0009935878; 0009946124; 0009948223; 0009950524; 0009950527; 0009953807; 0009953808; 0009956448; 0009959204; 0009959206; 0009959207; 0009959209; 0009965514; 0009968626; 0009968627; 0009968629; 0009971360; 0009971362; 0009971364; 0009973689; 0009973693; 0009973694; 0009979348; 0009981602; 0009981604; 0009989097; 0009989098; 0009992249; 0009992251; 0009994564; 0009994566; 0009999410; 0009999411; 0010001292; 0010004101; 0010007363; 0010007367; 0010007370; 0010010943; 0010010944; 0010014262; 0010014263; 0010014265; 0010017316; 0010017317; 0010021502; 0010021503; 0010024639; 0010024640; 0010026719; 0010028932; 0010032824; 0010032827; 0010035760; 0010035761; 0010037867; 0010037868; 0010037871; 0010037872; 0010037877; 0010040827; 0010044462; 0010044463; 0010047949; 0010051008; 0010051009; 0010057224; 0010057225; 0010058013; 0010058014; 0010058019; 0010058020; 0010063248; 0010081654; 0010081655; 0010087625; 0010087642; 0010091719; 0010091720; 0010093573; 0010093574; 0010095465; 0010097381; 0010097383; 0010099752; 0010099755; 0010100831; 0010103512; 0010103515; 0010108121; 0010108128; 0010108137; 0010108139; 0010108142; 0010118094; 0010118095; 0010120341; 0010124201; 0010124205; 0010126032; 0010126035; 0010129929; 0010129932; 0010139636; 0010142028; 0010143246; 0010145130; 0010145132; 0010145135; 0010148989; 0010153121; 0010157412; 0010161314; 0010161315; 0010163780; 0010163781; 0010163783; 0010163784; 0010171864; 0010171865; 0010171867; 0010175289; 0010182732; 0010182734; 0010182737; 0010187014; 0010187015; 0010188539; 0010188542; 0010188543; 0010194378; 0010197791; 0010199875; 0010199876; 0010204026; 0010204028; 0010209063; 0010209065; 0010209067; 0010211939; 0010215734; 0010215735; 0010217153; 0010217154; 0010217155; 0010220453; 0010220454; 0010222008; 0010222010; 0010225045; 0010225046; 0010245327; 0010245328; 0010245330; 0010248117; 0010248118; 0010249599; 0010249600; 0010252314; 0010252315; 0010253925; 0010255303; 0010255304; 0010256430; 0010256432; 0010257815; 0010258780; 0010260160; 0010261392; 0010262935; 0010262937; 0010268206; 0010269510; 0010269511; 0010271608; 0010271609; 0010273068; 0010273069; 0010275574; 0010275575; 0010278034; 0010278035; 0010279049; 0010279052; 0010279053; 0010279934; 0010279935; 0010281655; 0010281656; 0010282946; 0010285883; 0010285885; 0010286955; 0010291594; 0010291596; 0010297111; 0010297583; 0010300584; 0010300585; 0010301886; 0010303396;
Devices named in this recall
2 devices in the FDA device registry whose identifier appears in the recall's code information.
- EnVeo™ PRO ENVPRO-16-USMedtronic, Inc. · Aortic valve, prosthesis, percutaneously delivered1 recall23 deaths197 reports
- EnVeo™ PRO ENVPRO-14-USMedtronic, Inc. · Aortic valve, prosthesis, percutaneously delivered1 recall4 deaths55 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 88113).
- Z-2043-2021 · Open, ClassifiedCoreValve Evolut PRO PLUS Delivery Catheter System, REF: D-EVPROP2329US; D-EVPROP34US;
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page