Endurant IIs recallZ-2136-2021 · Medtronic Vascular, Inc. · initiated Jun 7, 2021

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Medtronic Vascular, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

During product complaint investigation, it was determined that during stent graft deployment, the radiopaque (RO) marker bond detached from the distal end of the graft cover.

FDA root cause category: Under Investigation by firm

What the company is doing

On 06/07/2021, the firm initiated a verbal notification regarding an "URGENT MEDICAL DEVICE RECALL" which informed customers that Earlier this week, Medtronic identified the potential for a specific subset of 20Fr Endurant II/IIs Stent Graft Systems to have the potential for the Radiopaque Marker to detach during deployment. There is evidence that devices that used one batch of graft cover subassemblies has a higher potential to exhibit this failure mode. This is an acute failure of the delivery system, there are no concerns with previously implanted stent grafts. We are requesting that you immediately refrain from using this specific subset of products and place it in quarantine until further notice. You will be receiving an official recall notice with instructions for product return for credit or replacement. Your Sales Representative will assist you with this process. We apologize for the inconvenience; however, this action is necessary to ensure the utmost patient safety with the use of our products. On 06/18/2021, the firm sent an "URGENT MEDICAL DEVICE RECALL" Notification Letter via UPS 2-Day delivery that restated the verbal notification. The Customer Notification Letter instructed customers to: 1)Return all unused affected devices via their sales representative or calling customer service at 1-888-283-7868 or email [email protected]. 2)Complete the enclosed Customer Confirmation Form and email to [email protected]. 3)Forward this notice to all those who need to be aware within their organization. For questions or assistance, contact Customer Service at 1-888-283-7868.

Recall details

Recall number
Z-2136-2021
Status
Open, Classified
Initiated by the company
Jun 7, 2021
Recalling firm
Medtronic Vascular, Inc.
Quantity
21 devices
Product code
MIH
FDA submissions
P100021
Distribution
U.S.: GA, MA, MD, NY, SC, TX, WA, and WV O.U.S.: Argentina, Australia, Germany, Greece, Israel, Italy, Netherlands, Poland, South Africa, Spain, Switzerland and United Kingdom.

Product description

Medtronic Endurant IIs Stent Graft System Medtronic Endurant IIs Stent Graft System REF ESBF3614C103EE and ESBF3214C103E. Used to treat infrarenal abdominal aortic or aortoiliac aneurysms using an endovascular approach.

Code information (models, lots, serial numbers)

Endurant IIs Stent Graft System: Model (REF):ESBF3214C103E GTIN: 00643169439962 Serial Numbers: V29780773; V29780771; V29780759; Model (REF): ESBF3614C103EE GTIN: 00643169780460 Serial Numbers: V29781932; V29781933; V29781934; V29781935; V29781936; V29781937; V29781938; V29781939; V29781940; V29781941; V29781942; V29781943; V29781944; V29781945; V29781946; V29781947; V29781948; V29781951;

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other devices under the same FDA submission

169 devices cleared under P100021 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 88158).