Skip to content

CODMAN recallZ-2224-2021 · Integra LifeSciences Corp. · initiated Jun 1, 2021

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Integra LifeSciences Corp. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Packaging-pouches with wrinkles across the Tyvek side may compromise the integrity of the pouch's sterile barrier

FDA root cause category: Process change control

What the company is doing

Integra notified consignees (Hospitals, Government facilities, and Distributors) by letter via FedX 2 Day Mail on June 1, 2021 and email to Distributor. Letter states reason for recall, health risk and action to take: We request that you take the following actions: 1. Customers a. If you do have units of the affected product listed in Table 1, remove it immediately from service. b. If you do have affected product, check the box on the enclosed form I do have affected product. Record the total quantity of the affected product that you have. c. If you do not have affected product, check the box, I do not have affected product. d. Complete the attached Acknowledgement Form and return to [email protected] or FAX to 1-609-750-4220. e. Keep a copy of the form for your records. f. When your form is received, and it is noted that you have affected product, Customer Service will contact you and provide a Return Material Authorization (RMA) number and directions to return the affected product. A replacement order will also be placed for the quantity noted on the form. 2. Distributors a. If you do have the affected product listed in Table 1, remove the product from further distribution. b. Complete the attached form. If you do not have affected product, check the box, I do not have affected product. c. If you do have affected product, check the box I do have affected product. Record the total quantity of affected product you have. d. Complete the attached Acknowledgement Form and return to [email protected] or FAX to 1-609-750-4220. e. Keep a copy of the form for your records. f. Please check your customer traceability records for shipments of above catalog numbers. g. Forward a copy of the enclosed Field Safety Notice to any of your customers that have purchased the above catalog and lot number. h. When the form from you and your customer is received, and it is noted that you or your customer has affected product, Customer Service will co

Recall details

Recall number
Z-2224-2021
Status
Terminated
Initiated by the company
Jun 1, 2021
Recalling firm
Integra LifeSciences Corp.
Quantity
906 units
Product code
HBO
Distribution
US Nationwide Distribution - Foreign Distribution: Hong Kong

Product description

CODMAN RANEY SCALP CLIPS-Disposable cylindrical clips designed to provide hemostasis to the scalp wound edge. Part Number: 201037

Code information (models, lots, serial numbers)

Lot Numbers: W2102045, W2102046, W2102048 and W2102049  GTIN# 10381780511625

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Sources and updates