Skip to content

728332 IQon Spectral CT-Computed Tomography X-ray system recallZ-2317-2021 · Philips North America Llc · initiated Jun 8, 2021

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Philips North America Llc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

When setting patient weight unit preferences to pounds, SynchRight P3T (Personalized Patient Protocol) software will result in increased contrast volume, resulting in increased contrast volume recommended for the patient.

FDA root cause category: Software design

What the company is doing

Philips issued Urgent Medical Device Correction letter dated 6/8/21 to consignees via certified mail. Letter states reason for recall, health risk and action to take: following steps: Go to Preferences> Regional Settings > Weight units, select the Weight units to Kg. Step 1: Verify that the patient weight unit is set to Kg, rather than Lbs. Step 2: Change the weight unit to Kg if weight unit is set to Lbs, referring to the screenshot below Step 3: Log out and then log back in to make the changes effective. Please share this notice with all those who need to be aware within your organization. Please complete and return the attached response form to Philips Healthcare promptly upon receipt and no later than 30 days from receipt, via email to [email protected] The software update is planned to be released in the 3rd quarter of 2021. Once the software update is released, a Philips representative will contact you to schedule a software update to permanently resolve the issue. For North America, contact the Customer Care Solutions Center (1-800-722-9377).

Recall details

Recall number
Z-2317-2021
Status
Terminated
Initiated by the company
Jun 8, 2021
Recalling firm
Philips North America Llc
Product code
JAK
FDA submissions
K193454
Distribution
US Nationwide distribution in the states of AZ, FL, LA, MN, NE, NV, NY, PA.

Product description

728332 IQon Spectral CT-Computed Tomography X-ray system

Code information (models, lots, serial numbers)

Serial Number/UDI: 60001 (01)00884838059542(21)60001, 60005 (01)00884838059542(21)60005, 60015 (01)00884838059542(21)60015, 60017 (01)00884838059542(21)60017, 860026 (01)00884838059542(21)860026, 860035 (01)00884838059542(21)860035, 860053 (01)00884838059542(21)860053 860074 (01)00884838059542(21)860074, 860108 (01)00884838059542(21)860108, 860109 (01)00884838059542(21)860109. Software Version 4.7.7 SynchRight Option

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Other products in the same recall event

The FDA groups related products into one recall event (event 88336).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page