cobas infinity central lab running software version 3.01.03… recallZ-2444-2021 · Roche Diagnostics Operations, Inc. · initiated Jun 24, 2021
Terminated
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Roche Diagnostics Operations, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Under specific circumstances created by the user, the cobas e flow test results could be replaced by an automatic result sent to the lab LIS. This automatic result could be released and interpreted by your Laboratory Information System (LIS) as a false positive or negative result. It should be noted that the "correct" result would still be available on the instrument.
FDA root cause category: Software design
What the company is doing
An Urgent Medical Device Correction notification letter dated 6/24/21 was sent to customers. Actions Required " Follow the Workaround section of this Urgent Medical Device Correction (UMDC), if applicable. " Consult with the physician or pathologist at your facility to determine any clinical implications specific to your patients. " Until the service patch is available, use the workaround described in this UMDC. " Complete all sections of the enclosed faxback form (TP-01337) and fax it to 1-847-457-1389 or email it to [email protected]. " File this UMDC for future reference. Identifying the Issue This issue occurs under specific circumstances and ALL of the conditions listed in the Issue section on page one of this UMDC must be fulfilled in order for the issue to occur. The issue can be identified when: " the current result appears as a repetition in the validation screen, " the Raw result column will appear blank indicating that the current result was not received from the instrument, and " the test alarm REP is triggered that indicates that the test results were repeated. Also, the issue can be detected in the validation screen based on the information that might be shown. The list below may help identify the issue (depending on your system configuration): " The LIS test result differs from the analyzer result. " There is a test alarm indicating the result is a repetition. " There is no analyzer associated with the result. " The main result time differs from some or all of the embedded result times. " The Preliminary result column displays the previous result, which corresponds to the original result sent by the instrument. " The repetition icon indicates the number of repetitions (i.e., 1 or more). Workaround Roche recommends contacting the Roche Support Network Customer Support Center at 1-800-526-2272 to request support for the workaround listed below: The following workaround can be used until the service patches are available to resolve this i
Recall details
- Recall number
- Z-2444-2021
- Status
- Terminated
- Initiated by the company
- Jun 24, 2021
- Recalling firm
- Roche Diagnostics Operations, Inc.
- Quantity
- 2058 licenses worldwide, 79 licenses in US
- Product code
- JQP
- Distribution
- US Nationwide distribution
Product description
cobas infinity central lab running software version 3.01.03 through 3.02.08, Catalog#: 07154003001
Code information (models, lots, serial numbers)
UDI: 04015630936007; All systems running software version 3.01.03 - 3.02.08
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- cobas® infinity central lab 07154003001Roche Diagnostics GmbH · Calculator/data processing module, for clinical use2 recalls31 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page