Skip to content

Potassium Hydroxide 10% (KOH 10%), Cat. No. 300145 recallZ-2450-2021 · Gibson Bioscience · initiated Jul 1, 2021

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Gibson Bioscience or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

A lot of KOH 10% was made with Hydrogen Peroxide 3% instead of KOH 10%.

FDA root cause category: Mixed-up of materials/components

What the company is doing

On July 1, 2021, the firm, Microbiologics, distributed "Urgent Medical Device Recall " letters to affected customers through mail and email. Customers were informed of the product issue. Customers were instructed to discard all affected inventory immediately. In addition, they were to complete and return the customer response form to [email protected], and keep the recall letter for their records. Customers may contact Microbiologics if a replacement is needed. Please contact Microbiologics' Recall Support Team at 320.229.7073 or [email protected] with questions or concerns. Collect calls may be made.

Recall details

Recall number
Z-2450-2021
Status
Terminated
Initiated by the company
Jul 1, 2021
Recalling firm
Gibson Bioscience
Quantity
11 kits
Product code
JTO
Distribution
US Domestic distribution to NY, KY, ME, NC, NY, CO, CA, and SC.

Product description

Potassium Hydroxide 10% (KOH 10%), Cat. No. 300145 Potassium Hydroxide 10% (KOH 10%) is intended for use in the demonstration of yeast and fungal elements in wet clinical preparations. Potassium Hydroxide digests the proteinaceous material in the specimen, leaving the fungal cell wall intact for direct microscopic observation.

Code information (models, lots, serial numbers)

Lot 21137116, UDI 10634853000633

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.