Ingenia recallZ-2341-2021 · Philips North America Llc · initiated Jul 30, 2021
Terminated
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Philips North America Llc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Pressure relief valve 42 psi installed in the magnet assembly instead of a 3.75 psi valve, may result in release of helium gas and cause injury
FDA root cause category: Nonconforming Material/Component
What the company is doing
Philips North America issued U.S. customers letters via certified mail using a 3rd Party service. Philips Markets Organizations are responsible for distributing the letters outside of the US. Letter states reason for recall , health risk and action to take: The Philips Ingenia 1.5T, Ingenia 1.5T Evolution, Ingenia 3.0T, and Ingenia Elition X MR systems can continue to fulfill clinical use as normal. However, in case of technical problems related to off cooling of the magnet, magnet pressure or Helium boil-off, please stop use and contact your local Philips service representative immediately and reference FCO78100530. A Philips Field Service Engineer (FSE) will contact you to schedule an on-site visit to resolve the issue by inspecting the potentially affected system(s) to confirm if the non-conforming pressure relief (42 psi) has been installed (reference FCO78100530). If your system has a 42 psi valve, the Phillips FSE will replace the valve with the correct 3.75 psi valve. you need any further information or support concerning this issue, please contact the Customer Care Solutions Center 1-800-722-9377 and reference FCO78100530.
Recall details
- Recall number
- Z-2341-2021
- Status
- Terminated
- Initiated by the company
- Jul 30, 2021
- Recalling firm
- Philips North America Llc
- Quantity
- 2 units US; 16 units OUS
- Product code
- LNH
- FDA submissions
- K162931
- Distribution
- Worldwide distribution - US Nationwide distribution in the states of AL, AR, AZ, CA, CO, MD, MD, NH, NY, OH, TN, TX, VA and the countries of Canada, Mexico, Australia, Austria, Chile, China, Czech Republic, Denmark, France, Germany, India, Indonesia, Italy, Japan, Korea, Republic of Lebanon, Netherlands, Romania, Russian Federation, Saudi Arabia, Spain, Turkey, Ukraine, United Kingdom.
Product description
Ingenia 1.5T Evoluton- Magnetic Resonance Medical Electrical Systems indicated for use as a diagnostic device. The system can produce cross-sectional images, spectroscopic images and/or spectra in any orientation of the internal structure of the head, body or extremities Model Number: 781315
Code information (models, lots, serial numbers)
S/N and UDI: US: 61044 (01)00884838099715(21)61044; 61047 (01)00884838099715(21)61047. OUS: 61028 (01)00884838099715 61008 (01)00884838099715 61025 (01)00884838099715 61026 (01)00884838099715 61034 (01)00884838099715 61035 (01)00884838099715 61042 (01)00884838099715 61046 (01)00884838099715 61040 (01)00884838099715 61041 (01)00884838099715 61018 (01)00884838099715 61017 (01)00884838099715 61032 (01)00884838099715 61043 (01)00884838099715 61033 (01)00884838099715 61064 (01)00884838099715
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Ingenia Ingenia 1.5TPhilips Medical Systems Nederland B.V. · System, nuclear magnetic resonance imaging9 recalls3 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 88461).
- Z-2342-2021 · TerminatedIngenia 1.5T-Magnetic Resonance Medical Electrical Systems indicated for use as a diagnostic device. The system can produce cross-sectional images, spectroscopic images and/or sp…
- Z-2343-2021 · TerminatedIngenia 3.0T-Magnetic Resonance Medical Electrical Systems indicated for use as a diagnostic device. The system can produce cross-sectional images, spectroscopic images and/or sp…
- Z-2344-2021 · TerminatedIngenia Elition X-Magnetic Resonance Medical Electrical Systems indicated for use as a diagnostic device. The system can produce cross-sectional images, spectroscopic images and/…
- Z-2345-2021 · TerminatedIngenia CX 3.0T-Magnetic Resonance Medical Electrical Systems indicated for use as a diagnostic device. The system can produce cross-sectional images, spectroscopic images and/or…
- Z-2347-2021 · TerminatedIngenia Elition S-Magnetic Resonance Medical Electrical Systems indicated for use as a diagnostic device. The system can produce cross-sectional images, spectroscopic images and/…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page