Scorpion recallZ-2501-2021 · Argon Medical Devices, Inc · initiated Jul 8, 2021

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Argon Medical Devices, Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

As a result of design changes, sheaths have exhibited cracking/breaking at the tips.

FDA root cause category: Device Design

What the company is doing

On 07/08/2021, the firm's sales representative/clinical Specialist contacted via telephone to inform the single customer that the Recalling Firm was pausing evaluations of the affected device and was requesting return of the device. The customer was also informed that the Recalling Firm was in the process of implementing a few changes to improve upon the product. The product was place on hold at the customer's facility and the Recalling Firm's sales representative was picked up and returned to the Recalling Firm on 07/21/2021.

Recall details

Recall number
Z-2501-2021
Status
Open, Classified
Initiated by the company
Jul 8, 2021
Recalling firm
Argon Medical Devices, Inc
Quantity
20 units
Product code
DYB
FDA submissions
K202141
Distribution
U.S. Nationwide distribution in the state of WA. O.U.S.: N/A

Product description

Scorpion Portal Vein Access Set - Product Usage: used to gain access to the hepatic vein and guide a sharp puncture tool (0.040 Stylet or 17ga Needle) through the parenchyma.

Code information (models, lots, serial numbers)

Item Numbers: TPS005 (Stylet); TPS006 (Needle); UDI Code: 00886333225552 Lot Number: 11369609

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Sources and updates