Outlook recallZ-2546-2021 · B. Braun Medical, Inc. · initiated Aug 10, 2021
Terminated
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from B. Braun Medical, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Leakage of the pump set within the cassette portion of the pump sets, potential for delays in administration of medications, under-delivery or incomplete dosing and bloodstream infections.
FDA root cause category: Under Investigation by firm
What the company is doing
BBraun issued Urgent Medical Device Correction letter to direct end customers and BBMI distributors, dated 8/10/21. Letter states reason for recall, health risk and action to take: Review the Device Recall Notification in its entirety and ensure that all users in your organization of the impacted product, and other concerned persons, are informed about this voluntary product recall. If you are a distributor, please forward this recall notification to your customers. The recall is to be extended to the customer level. 2. Copies of Attachment 02 should be distributed to all personnel and staff who utilize Outlook Pump Sets within their care area to ensure they identify impacted units prior to use. DO NOT use any Impacted units identified by the Instructions. Discard any Identified products to ptevent Inadvertent use. 3. B. Braun does not recommend return of this material at this time as it may lead to shortages of supply which may adversely impact patients. Not all pump sets within the identified lot are impacted. If your facility has both impacted and non-impacted lots of material, you may wish to direct impacted lots to care areas where less critical medications are administered or where risks of exposure to hazardous (e.g. cytotoxic) medications may be decreased. 4. Complete the attached "Product Correction Acknowledgement," form and return it to B. Braun Medical Inc. Quality Assurance department by faxing the form to (610) 849-1197 or e-mail to [email protected] within two (2) weeks of receipt. 5. To receive credit for any units which you have had to discard via this process, please contact customer support by calling 1-800-227-2862. Please be sure to identify your local distributor when contacting customer support so we can ensure that proper credit is provided
Recall details
- Recall number
- Z-2546-2021
- Status
- Terminated
- Initiated by the company
- Aug 10, 2021
- Recalling firm
- B. Braun Medical, Inc.
- Quantity
- 44,866 units
- Product code
- FPA
- FDA submissions
- K083723, K913661, K942988
- Distribution
- Worldwide distribution - US Nationwide distribution and the country of Canada.
Product description
Outlook Pump 15 Drop 0.2 Micron Filtered IV Set with 2 CARESITE Y-Sites: single-use for use with the Outlook Safety Infusion System and/or Horizon NXT Pump for infusion of parenteral fluids Ref: 354214
Code information (models, lots, serial numbers)
Lot Code: 0061752471 0061756801 0061757861 0061766196 0061777798 DI: 04046964182235
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Outlook® 354214B. Braun Medical Inc. · Set, administration, intravascular2 recalls4 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 88531).
- Z-2544-2021 · Terminated15 drops/ml, Priming Volume: 27ml, Length: 129 in. Single-use for use with the Outlook Safety Infusion System and/or Horizon NXT Pump for infusion of parenteral fluid Material ID:…
- Z-2545-2021 · TerminatedOutlook Pump Sets w/3 CARESITE Needle-Free Injection Sites: single-use for use with the Outlook Safety Infusion System and/or Horizon NXT Pump for infusion of parenteral fluids Ma…
- Z-2547-2021 · Terminated60 drops/ml, Priming volume: 25 ml, Length: 120 in.in-single-use for use with the Outlook Safety Infusion System and/or Horizon NXT Pump for infusion of parenteral fluids Ref: 35…
- Z-2548-2021 · Terminated10 drops/ml, Priming volume: 46ml, Length: 130 in.-single-use for use with the Outlook Safety Infusion System and/or Horizon NXT Pump for infusion of parenteral fluids Ref: 35421…
- Z-2549-2021 · TerminatedOutlook Pump Measured Volume Solution Set/ w/ Auto Shut-Off. Single-use for use with the Outlook Safety Infusion System and/or Horizon NXT Pump for infusion of parenteral fluids R…
- Z-2550-2021 · TerminatedET82HP OUTLOOK PUMP EXTENSION SET-single-use for use with the Outlook Safety Infusion System and/or Horizon NXT Pump for infusion of parenteral fluids Ref: 470024
- Z-2551-2021 · TerminatedV1500HP OUTLOOK PUMP SET 118-single-use for use with the Outlook Safety Infusion System and/or Horizon NXT Pump for infusion of parenteral fluids Ref: 470025
- Z-2552-2021 · TerminatedOUTLOOK PUMP IV SET W/UNIVERSAL SPIKE-single-use for use with the Outlook Safety Infusion System and/or Horizon NXT Pump for infusion of parenteral fluids Ref: 474004
- Z-2553-2021 · TerminatedY-Type Blood, Outlook Pump Set, Caresite-single-use for use with the Outlook Safety Infusion System and/or Horizon NXT Pump for infusion of parenteral fluids Ref: 477005
- Z-2554-2021 · TerminatedOUTLOOK PUMP IV SET, 134 IN.-single-use for use with the Outlook Safety Infusion System and/or Horizon NXT Pump for infusion of parenteral fluids Ref: 477009
- Z-2555-2021 · TerminatedPUMP SET, 15 DR, 3 CARESITE, 127 in.single-use for use with the Outlook Safety Infusion System and/or Horizon NXT Pump for infusion of parenteral fluids Ref: 480239
- Z-2556-2021 · TerminatedOUTLOOK PUMP SET, 3 CARESITE LADS, 115 IN-single use, for use with the Outlook Safety Infusion System and/or Horizon NXT Pump for infusion of parenteral fluids Ref: 490361
- Z-2557-2021 · TerminatedOUTLOOK PUMP SET 3 CARESITE LADS 133 IN.-single use, for use with the Outlook Safety Infusion System and/or Horizon NXT Pump for infusion of parenteral fluids Ref: 490564
- Z-2558-2021 · TerminatedOutlook Pump Add-On Set-single use, for use with the Outlook Safety Infusion System and/or Horizon NXT Pump for infusion of parenteral fluids Ref: V7400
- Z-2559-2021 · TerminatedOutlook Pump Primary Administration Set w/ micro Tubing for Epidural Administration-single use, for use with the Outlook Safety Infusion System and/or Horizon NXT Pump for infusion…
- Z-2560-2021 · TerminatedOutlook Pump Blood Administration Set with SAFELINE Injection Site-single use, for use with the Outlook Safety Infusion System and/or Horizon NXT Pump for infusion of parenteral fl…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page