Access hsTnI High Sensitivity Troponin I, REF B52699 recallZ-0032-2022 · Beckman Coulter Inc. · initiated Aug 9, 2021
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Beckman Coulter Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Through customer feedback, testing a sample with cardiac troponin I concentration that are >270,000 pg/mL may cause carryover into the hsTnl reagent pack. The carryover is caused by sample-to-pack carryover predominantly into the particle well of the reagent pack by way of the probe. The particle well then becomes contaminated by the high sample from the probe. The extent of the carryover is dependent on sample concentration and frequency of high concentration samples.
FDA root cause category: Unknown/Undetermined by firm
What the company is doing
During the week of 08/09/2021, the firm sent an "URGENT MEDICAL DEVICE RECALL" Notification Letter via mail or email informing customer of possible carryover with in-use, open (punctured) hsTnI reagent packs, and the impact of carryover on patient samples that are tested from the same reagent pack as a sample with a high cardiac troponin (cTnI) concentration >270,000 pg/mL (ng/L), or the carryover may also impact a different reagent pack. Clinical carryover can affect the results of all subsequent samples that are tested from the affected pack. Customer are instructed to: - If an hsTnI result >270,000 pg/mL (ng/L) are observed, perform the following steps: 1. Remove and discard all open Access hsTnI reagent packs. For customers in the United States, if replacement products are needed, complete the attached Replacement Order Form and email to [email protected] or fax to (866) 294-7850 OR Call Client Services at (800) 526-3821. 2. Load a single Access hsTnI reagent pack. 3. Run the current low level hsTnI QC on all reagent pipettors configured for hsTnI to verify that there is no further carryover. NOTE: UniCel DxI operators can test all configured reagent pipettors by setting up a QC file as outlined in Appendix A (included with the recall notice). 4. If the QC result is within the laboratory s defined ranges for each pipettor configured, repeat each positive hsTnI sample that was tested after the >270,000 pg/mL (ng/L) cTnI sample and then continue normal operation. Load additional reagent packs if it is appropriate for the laboratory s testing requirements. 5. If the QC result are not within the acceptable range, contact the Recalling Firm's Customer Technical Support for further assistance. Additional instructions: - Download the most current version of the hsTnI reagent -nstructions for Use (IFU) from the Recalling Firm's website. Update laboratory procedures as appropriate. -Share this information with laboratory staff and retain this not
Recall details
- Recall number
- Z-0032-2022
- Status
- Open, Classified
- Initiated by the company
- Aug 9, 2021
- Recalling firm
- Beckman Coulter Inc.
- Quantity
- 293,640 kits
- Product code
- MMI
- FDA submissions
- K172783, K172787
- Distribution
- U.S.: AL, AR, AZ, CA, CO, CT, DE, DC, FL, GA, Guam, HI, IA, ID, IL, IN, KS, KY, LA, ME, MD, MI, MN, MS, MO, NC, ND, NE, NV, NJ, NY, OH, OK, OR, PA, Puerto Rico, RI, SC, SD, TN, TX, VA, WA, WI, WV, and WY O.U.S.: Algeria, Andorra, Australia, Austria, Bahrain, Bangladesh, Belgium, Bosnia and Herzegovina, Brazil, Brunei, Bulgaria, Canada, China, Croatia, Cura¿ao, Czech Republic, Denmark, Egypt, Estonia, Ethiopia, Finland, France, Georgia, Germany, Ghana, Gibraltar, Greece, Hong Kong, Hungary, India, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Kuwait, Lebanon, Libya, Lithuania, Malaysia, Mayotte, Mexico, Morocco, Myanmar, Namibia, Nepal, Netherlands, New Zealand, Niger, Nigeria, Norway, Oman, Philippines, Poland, Portugal, Puerto Rico, Qatar, Romania, Russia, San Marino, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sri Lanka, Swaziland, Switzerland, Taiwan, Thailand, Tunisia, Turkey, United Arab Emirates, United Kingdom, Vietnam, Yemen, and Zimbabwe
Product description
Access hsTnI High Sensitivity Troponin I, REF B52699 Access Immunoassay System REAGENT PACK
Code information (models, lots, serial numbers)
Show the full code information (2k characters)
Catalog Number:B52699 Lot Number / UDI Code: 921491 / (01)15099590693183(17)201130(11)191201(10)921491; 921638 / (01)15099590693183(17)201231(11)200101(10)921638; 921631 / (01)15099590693183(17)210131(11)200201(10)921631; 921750 / (01)15099590693183(17)210131(11)200201(10)921750; 921805 / (01)15099590693183(17)210228(11)200229(10)921805; 921831 / (01)15099590693183(17)210331(11)200331(10)921831; 921943 / (01)15099590693183(17)210430(11)200430(10)921943; 922000 / (01)15099590693183(17)210531(11)200531(10)922000; 922039 / (01)15099590693183(17)210630(11)200630(10)922039; 922040 / (01)15099590693183(17)210731(11)200731(10)922040; 922062 / (01)15099590693183(17)210731(11)200731(10)922062; 922145 / (01)15099590693183(17)210831(11)200831(10)922145; 922146 / (01)15099590693183(17)210930(11)200930(10)922146; 922338 / (01)15099590693183(17)210930(11)200930(10)922338; 922440 / (01)15099590693183(17)211031(11)201031(10)922440; 922545 / (01)15099590693183(17)211031(11)201031(10)922545; 922547 / (01)15099590693183(17)211130(11)201130(10)922547; 922713 / (01)15099590693183(17)211231(11)201231(10)922713; 922714 / (01)15099590693183(17)211231(11)201231(10)922714; 922722 / (01)15099590693183(17)220131(11)210131(10)922722; 922897 / (01)15099590693183(17)220131(11)210131(10)922897; 124235 / (01)15099590693183(17)220228(11)210228(10)124235; 124339 / (01)15099590693183(17)220228(11)210228(10)124339; 922898 / (01)15099590693183(17)220228(11)210228(10)922898; 124340 / (01)15099590693183(17)220331(11)210331(10)124340; 124348 / (01)15099590693183(17)220430(11)210430(10)124348; 124353 / (01)15099590693183(17)220430(11)210430(10)124353; 124688 / (01)15099590693183(17)220430(11)210430(10)124688; 124354 / (01)15099590693183(17)220531(11)210531(10)124354; 124355 / (01)15099590693183(17)220630(11)210630(10)124355; 124356 / (01)15099590693183(17)220630(11)210630(10)124356; 124357 / (01)15099590693183(17)220630(11)210630(10)124357;
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Access hsTnI B52669Beckman Coulter, Inc. · Immunoassay Method, Troponin Subunit2 recalls170 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page