Skip to content

MR MAGNETOM Systems with RT Image Suite with syngo MR… recallZ-2447-2021 · Siemens Medical Solutions USA, Inc · initiated Aug 16, 2021

Class II Open, Classified All recalls posted in September 2021

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Siemens Medical Solutions USA, Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Potential for data loss when using the contouring or patient marking workflow when editing a structure set after saving and reopening a study, resulting in an incomplete structure set sent to the treatment planning system (TPS).

FDA root cause category: Software design

What the company is doing

Siemens has initiated three (3) Customer Safety Advisory Notices (CSANs) to affected customers via SY052/21/S, CT042/21/S and MR019/21/S by email or hand delivery by the Siemens Customer Service Engineer on 8/16/21. Temporary workarounds are available for some of the software versions and are described in the CSANs. Additionally, a solution to eliminate the root cause of this problem is being distributed via SY051/21/S (for SY and CT) and MR021/21/S for MR to all affected customers free of charge. The service organization will contact you to arrange a date to perform this corrective action free of charge. Contact the service organization for an earlier appointment at 1-800-888-7436. ***UPDATE: 01/31/22 Additionally, a solution to eliminate the root cause of this problem is being distributed via CT049/21S, CT061/21/S, CT062/21/S, and CT088/21/S for CT. ***UPDATE: 03/12/22 Status Report Additional 13 S/N's added

Recall details

Recall number
Z-2447-2021
Recall class
Class II
Status
Open, Classified
Initiated by the company
Aug 16, 2021
Posted by the FDA
Sep 10, 2021
Recalling firm
Siemens Medical Solutions USA, Inc
Quantity
2 systems
Product code
LNH
FDA submissions
K203443
Distribution
US Nationwide distribution.

Product description

MR MAGNETOM Systems with RT Image Suite with syngo MR VA20A, VA30A and VA31A with license SW-Key CT_RT_IMA_SUITE_SYNTHETIC_CT Material Numbers: 11060815

Code information (models, lots, serial numbers)

Serial Numbers: 176302 175681

Other devices under the same FDA submission

1 device cleared under K203443 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page