Power Processor 3K, 3K High Speed Stockyards recallZ-0117-2022 · Beckman Coulter Inc. · initiated Jul 28, 2021

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Beckman Coulter Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Missing splash guard which may expose users to hazardous materials. .

FDA root cause category: Process control

What the company is doing

During the week of 08/30/2021, the Recalling Firm sent an "URGENT MEDICAL DEVICE RECALL" Notification via its Field Action Notification process. On 09/16/2021, the Recalling Firm communicated a updated/revised "URGENT MEDICAL DEVICE RECALL" Notification letter which informed customer that, as a result of internal investigations, the Recalling Firm has discovered that the track-level splash guard has not been consistently provided to all customers. This guard is required for safe operation of the Power Processor system and could potentially cause user injuries if the splash guard is not in place. Customer are being informed that a Recalling Firm Field Service representative will check their system for the splash guard. if the splash guard is not present, the Field Service representative will order and install it. Customers are instructed to: 1. Continue to follow all safety warnings and system operation information provided in the Instructions for Use. 2. Share this information with their laboratory staff and to retain this notification as part of their laboratory Quality System documentation. 3. If customers have forwarded any of the affected products to another laboratory, provide a copy of this notification to them. 4. To confirm receipt of this Notification, customer are asked to respond within 10 days in one of the following ways: -Electronically, if you received this communication via email. -Manually, complete and return the enclosed Response Form. Any questions, contact the Recalling Firm's Customer Support Center or their local representative via website http://www.beckmancoulter.com Update 12/15/2021: On 12/13/2021, the firm sent an "URGENT MEDICAL DEVICE RECALL" Letter informing of additional products being added to the original Recall initiated in 08/2021 for the missing splash guards. A Recalling Firm's Representative will check the systems and if a splash guard is not present, they will order and install it. In addition, the Recalling Firm is a

Recall details

Recall number
Z-0117-2022
Status
Open, Classified
Initiated by the company
Jul 28, 2021
Posted by the FDA
Oct 14, 2021
Recalling firm
Beckman Coulter Inc.
Quantity
Total Units=271 units [258 units (3K and 3K HS); 13 units (5K added 12/15/2021)]
Product code
CDD
FDA submissions
K110413
Distribution
U.S. Nationwide Distribution in the states of: CA, GA, IA, ID, IL, IN, LA, MA, MI, MS, NJ, NY, OH, OR, PA, SC, SD, TN, TX, WA, and WV O.U.S.: Canada, China, Spain, France, United Kingdom, Israel, Italy, South Korea, Kuwait, Lebanon, Mexico, Malaysia, Netherlands, Portugal, Qatar, Saudi Arabia, Thailand, Turkey, and Taiwan. Additional O.U.S. added 12/15/2021 - South Africa

Product description

Power Processor 3K, 3K High Speed Stockyards Module Nameplate: PP 3K Stockyard (Refrig) 9707 REF A88116

Code information (models, lots, serial numbers)

Show the full code information (1k characters)

Catalog Number: A88116 - 3K HS Refrigerated Stockyard UDI: 15099590591960 Serial Numbers: 90HF1304006; 90HF1306009; 90HF1303005; 90HF1508005; 90HF1302005; 90HF1610005; 90HF1806001; 90HF1209005; 90HF1303007; 90HF1708006; 90HF1205001; 90HF1805002; 90HF1402005; 90HF1503001; 90HF1104002; 90HF1208004; 90HF1208005; 90HF1203001; 90HF1012004; 90HF1101002; 90HF1702003; 90HF1402004; 90HF1303008; 90HF1210001; 90HF1207001; 90HF1305007; 90HF1305004; 90HF1610001; 90HF1505002; 90HF1307005; 90HF1610002; 90HF1508003; 90HF1210002; 90HF1306001; 90HF1206002; 90HF1208002; 90HF1305008; 90HF1009002; 90HF1101001; Catalog Number: 6915556 - Power Processor 3K Stockyard (Stockyard 3000/R Kit for PWR PROC) Serial Numbers: 90AM0805001; 90AM0902009; 90AM0902005; 90AM0806003; 90AM0801003; 90AM0802002; 90AM0803001; 90AM0912002; 90AM1003002; 90AM0905010; 90AM0801004; 90AM06007008; 90AM060703; 90AM060704; 90AM0710002; 90AM0905007; 90AM0710008; 90AM0903002; 90AM0808002; 90AM0711004; 90AM0711005; 90AM0810001; 90AM0901002; 90AM0905006; 90AM0904004; 90AM0904005; 90AM0803002; 90AM0710004; Amended - additional Products and Codes (12/15/2021): B05520 - Stockyard, 5K HS, Refrigerated Serial Numbers: 90HG1104002; 90HG1101001; 90HG1105002; 90HG1104001; 90HG1101003; 90HG1101002; 90HG1202004; 90HG1102005; 90HG1101004; 90HG1206002; A88115 - 5K HS Refrig Stockyard Serial Numbers: 90HG0911001 90HG0911003 90HG0911002

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Other devices under the same FDA submission

1 device cleared under K110413 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page