Power Processor 3K, 3K High Speed Stockyards recallZ-0117-2022 · Beckman Coulter Inc. · initiated Jul 28, 2021
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Beckman Coulter Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Missing splash guard which may expose users to hazardous materials. .
FDA root cause category: Process control
What the company is doing
During the week of 08/30/2021, the Recalling Firm sent an "URGENT MEDICAL DEVICE RECALL" Notification via its Field Action Notification process. On 09/16/2021, the Recalling Firm communicated a updated/revised "URGENT MEDICAL DEVICE RECALL" Notification letter which informed customer that, as a result of internal investigations, the Recalling Firm has discovered that the track-level splash guard has not been consistently provided to all customers. This guard is required for safe operation of the Power Processor system and could potentially cause user injuries if the splash guard is not in place. Customer are being informed that a Recalling Firm Field Service representative will check their system for the splash guard. if the splash guard is not present, the Field Service representative will order and install it. Customers are instructed to: 1. Continue to follow all safety warnings and system operation information provided in the Instructions for Use. 2. Share this information with their laboratory staff and to retain this notification as part of their laboratory Quality System documentation. 3. If customers have forwarded any of the affected products to another laboratory, provide a copy of this notification to them. 4. To confirm receipt of this Notification, customer are asked to respond within 10 days in one of the following ways: -Electronically, if you received this communication via email. -Manually, complete and return the enclosed Response Form. Any questions, contact the Recalling Firm's Customer Support Center or their local representative via website http://www.beckmancoulter.com Update 12/15/2021: On 12/13/2021, the firm sent an "URGENT MEDICAL DEVICE RECALL" Letter informing of additional products being added to the original Recall initiated in 08/2021 for the missing splash guards. A Recalling Firm's Representative will check the systems and if a splash guard is not present, they will order and install it. In addition, the Recalling Firm is a
Recall details
- Recall number
- Z-0117-2022
- Status
- Open, Classified
- Initiated by the company
- Jul 28, 2021
- Posted by the FDA
- Oct 14, 2021
- Recalling firm
- Beckman Coulter Inc.
- Quantity
- Total Units=271 units [258 units (3K and 3K HS); 13 units (5K added 12/15/2021)]
- Product code
- CDD
- FDA submissions
- K110413
- Distribution
- U.S. Nationwide Distribution in the states of: CA, GA, IA, ID, IL, IN, LA, MA, MI, MS, NJ, NY, OH, OR, PA, SC, SD, TN, TX, WA, and WV O.U.S.: Canada, China, Spain, France, United Kingdom, Israel, Italy, South Korea, Kuwait, Lebanon, Mexico, Malaysia, Netherlands, Portugal, Qatar, Saudi Arabia, Thailand, Turkey, and Taiwan. Additional O.U.S. added 12/15/2021 - South Africa
Product description
Power Processor 3K, 3K High Speed Stockyards Module Nameplate: PP 3K Stockyard (Refrig) 9707 REF A88116
Code information (models, lots, serial numbers)
Show the full code information (1k characters)
Catalog Number: A88116 - 3K HS Refrigerated Stockyard UDI: 15099590591960 Serial Numbers: 90HF1304006; 90HF1306009; 90HF1303005; 90HF1508005; 90HF1302005; 90HF1610005; 90HF1806001; 90HF1209005; 90HF1303007; 90HF1708006; 90HF1205001; 90HF1805002; 90HF1402005; 90HF1503001; 90HF1104002; 90HF1208004; 90HF1208005; 90HF1203001; 90HF1012004; 90HF1101002; 90HF1702003; 90HF1402004; 90HF1303008; 90HF1210001; 90HF1207001; 90HF1305007; 90HF1305004; 90HF1610001; 90HF1505002; 90HF1307005; 90HF1610002; 90HF1508003; 90HF1210002; 90HF1306001; 90HF1206002; 90HF1208002; 90HF1305008; 90HF1009002; 90HF1101001; Catalog Number: 6915556 - Power Processor 3K Stockyard (Stockyard 3000/R Kit for PWR PROC) Serial Numbers: 90AM0805001; 90AM0902009; 90AM0902005; 90AM0806003; 90AM0801003; 90AM0802002; 90AM0803001; 90AM0912002; 90AM1003002; 90AM0905010; 90AM0801004; 90AM06007008; 90AM060703; 90AM060704; 90AM0710002; 90AM0905007; 90AM0710008; 90AM0903002; 90AM0808002; 90AM0711004; 90AM0711005; 90AM0810001; 90AM0901002; 90AM0905006; 90AM0904004; 90AM0904005; 90AM0803002; 90AM0710004; Amended - additional Products and Codes (12/15/2021): B05520 - Stockyard, 5K HS, Refrigerated Serial Numbers: 90HG1104002; 90HG1101001; 90HG1105002; 90HG1104001; 90HG1101003; 90HG1101002; 90HG1202004; 90HG1102005; 90HG1101004; 90HG1206002; A88115 - 5K HS Refrig Stockyard Serial Numbers: 90HG0911001 90HG0911003 90HG0911002
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Power Processor Sample Handling SystemBeckman Coulter, Inc. · Analyzer, Chemistry (Photometric, Discrete), For Clinical Use1 recall2 reports
Other devices under the same FDA submission
1 device cleared under K110413 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- Power LinkBeckman Coulter, Inc. · Analyzer, Chemistry (Photometric, Discrete), For Clinical Use1 report
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page