Medtronic Sterile single-use blades that are components of… recallZ-0158-2022 · Medtronic Xomed, Inc. · initiated Aug 31, 2021
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Medtronic Xomed, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
There was an increase in complaints related to tip breaks and wobble/vibration.
FDA root cause category: Process control
What the company is doing
The recalling firm issued letters dated 8/13/2021 on 8/31/2021 via USPS. The letter explained the reason for recall, risk to patient, and the actions to be taken. The consignee was requested to review their inventory of Medtronic blades against the list of affected products enclosed with the letter and to quarantine them. All unused, affected devices are to be returned to Medtronic by completing the Customer Confirmation form and returning it via FAX or email. Once completed, Customer Care will provide return instructions and issue a replacement for the returned product.
Recall details
- Recall number
- Z-0158-2022
- Status
- Open, Classified
- Initiated by the company
- Aug 31, 2021
- Posted by the FDA
- Oct 21, 2021
- Recalling firm
- Medtronic Xomed, Inc.
- Quantity
- 274,991 total devices
- Product code
- HBE
- FDA submissions
- K081475
- Distribution
- Worldwide distribution - US Nationwide, including PR. There was also government/military distribution and the countries of Australia, Austria, Bahrain, Belgium, Canada, Chile, Croatia, Czech Republic, Denmark, Dominican Republic, Estonia, Finland, France, Germany, Greece, Guatemala, Hong Kong, Iceland, India, Iran, Ireland, Israel, Italy, Jamaica, Japan, Kenya, Republic of Korea, Kuwait, Latvia, Libya, Luxembourg, Macao, Malaysia, Mauritius, Mongolia, Namibia, Netherlands, New Zealand, Norway, Oman, Pakistan, Panama, Philippines, Poland, Portugal, Reunion, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Thailand, Trinidad and Tobago, Turkey, Ukraine, United Arab Emirates, United Kingdom, and Viet Nam.
Product description
Medtronic Sterile single-use blades that are components of the StraightShot M Series Handpiece and Fusion ENT Navigation Software, and the Midas Rex Microsaw and Triton System with the following product numbers: (1) REF 40A2500, Nucleus Removal Angled Blade, 25 cm x 4.0 mm; (2) REF 40R2000, Nucleus Removal Set, 20 cm x 4.0 mm; (3) REF 40R2500, Nucleus Removal Set, 25 cm x 4.0 mm; (4) REF 40S2000, Nucleus Removal Straight Blade, 20 cm x 4.0 mm; (5) REF 40S2500, Nucleus Removal Straight Blade, 25 cm x 4.0 mm; (6) REF 45A2000, Nucleus Removal Angled Blade, 20 cm x 4.5 mm; (7) REF 45A2500, Nucleus Removal Angled Blade, 25 cm x 4.5 mm; (8) REF 45R2000, Nucleus Removal Set, 20 cm x 4.5 mm; (9) REF 45R2500, Nucleus Removal Set, 25 cm x 4.5 mm; (10) REF 45S2000, Nucleus Removal Straight Blade, 20 cm x 4.5 mm; (11) REF 45S2500, Nucleus Removal Straight Blade, 25 cm x 4.5 mm.
Code information (models, lots, serial numbers)
Show the full code information (1k characters)
(1) REF 40A2500, GTIN/UPN 00763000033835, Lot/serial number 0220002825; (2) REF 40R2000, GTIN/UPN 00763000033866, Lot/serial number 0218179458; (3) REF 40R2500, GTIN/UPN 00763000033798, Lot/serial number 0218299192; (4) REF 40S2000, GTIN/UPN 00763000033873, Lot/serial number 0218179459; (5) REF 40S2500, GTIN/UPN 00763000043193, Lot/serial number 0220226674; (6) REF 45A2000, GTIN/UPN 00763000033880, Lot/serial number 0220443669; (7) REF 45A2500, GTIN/UPN 00763000043223, Lot/serial number 0218323602; (8) REF 45R2000, GTIN/UPN 00763000033897, Lot/serial numbers 0218284087, 0218292178, 0218973633; (9) REF 45R2500, GTIN/UPN 00763000033811, Lot/serial number 0218297618; (10) REF 45S2000, GTIN/UPN 00763000033903, Lot/serial number 0218299193; (11) REF 45S2500, GTIN/UPN 00763000043216, Lot/serial number 0218297619.
Devices named in this recall
11 devices in the FDA device registry whose identifier appears in the recall's code information.
- N/A 40R2500Medtronic Xomed, Inc. · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)1 recall0 reports
- N/A 45R2500Medtronic Xomed, Inc. · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)1 recall0 reports
- N/A 40A2500Medtronic Xomed, Inc. · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)1 recall0 reports
- N/A 40R2000Medtronic Xomed, Inc. · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)1 recall0 reports
- N/A 40S2000Medtronic Xomed, Inc. · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)1 recall0 reports
- N/A 45A2000Medtronic Xomed, Inc. · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)1 recall0 reports
- N/A 45R2000Medtronic Xomed, Inc. · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)1 recall0 reports
- N/A 45S2000Medtronic Xomed, Inc. · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)1 recall0 reports
- N/A 40S2500Medtronic Xomed, Inc. · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)1 recall0 reports
- N/A 45S2500Medtronic Xomed, Inc. · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)1 recall0 reports
- N/A 45A2500Medtronic Xomed, Inc. · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)1 recall0 reports
Other devices under the same FDA submission
19 devices cleared under K081475 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- Indigo™ 1845000Medtronic Xomed, Inc. · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)165 reports
- Indigo 1845020Medtronic Xomed, Inc. · DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE105 reports
- Midas Rex® EM210Medtronic Xomed, Inc. · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)64 reports
- Midas Rex® EC300RFMedtronic Xomed, Inc. · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)1 recall1 report
- NA EC300RFMedtronic Xomed, Inc. · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)1 recall0 reports
- MIDAS REX® EM210RFMedtronic Xomed, Inc. · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)48 reports
- Midas Rex® Stylus Touch® EM210RFMedtronic Xomed, Inc. · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)36 reports
- Indigo 1845010Medtronic Xomed, Inc. · DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE33 reports
- Midas Rex® EM210Medtronic Xomed, Inc. · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)11 reports
- NA TN40RCDMedtronic Xomed, Inc. · DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE11 reports
- Midas Rex® EF201RFMedtronic Powered Surgical Solutions · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)7 reports
- NA TN30RCDMedtronic Xomed, Inc. · DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE6 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 88586).
- Z-0156-2022 · Open, ClassifiedMedtronic sterile single-use blades that are components of the StraightShot M Series Handpiece and Fusion ENT Navigation Software, and the Midas Rex Microsaw and Triton System with…
- Z-0157-2022 · Open, ClassifiedMedtronic sterile single-use blades that are components of the StraightShot M Series Handpiece and Fusion ENT Navigation Software, and the Midas Rex Microsaw and Triton System with…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page