Aptio Automation Interface Module to ADVIA Centaur XP/XPT.… recallZ-0193-2022 · Siemens Healthcare Diagnostics, Inc. · initiated Oct 6, 2021

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Siemens Healthcare Diagnostics, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The Aptio Automation Firmware for the ADVIA Centaur XP/XPT Interface Module may lead, in some cases, to an incorrect association of test results to sample ID. Incorrect association of test results could impact clinical interpretations and clinical decisions.

FDA root cause category: Software design

What the company is doing

On October 6, 2021, the firm notified affected customers via an "Urgent Medical Device Correction" letter. Customers were informed of the product issue and provided instructions on actions to be taken by laboratories, which prevent the issue from happening. If you have any questions, please contact your Siemens Healthineers Customer Care Center or your local Siemens Healthineers technical support representative. A new version of the Firmware for the Aptio Automation Interface Module to the ADVIA Centaur XP/XPT will correct the issue. Your service provider will contact you to schedule the firmware upgrade.

Recall details

Recall number
Z-0193-2022
Status
Open, Classified
Initiated by the company
Oct 6, 2021
Recalling firm
Siemens Healthcare Diagnostics, Inc.
Quantity
1142
Product code
CDT
FDA submissions
K043546
Distribution
Worldwide distribution.

Product description

Aptio Automation Interface Module to ADVIA Centaur XP/XPT. Aptio by Siemens Centaur Interface Firmware (FW) CENa_3-1-0-1 and lower. SMN 10703038

Code information (models, lots, serial numbers)

UDI 00630414596754

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.