Skip to content

OMNI K2 Hip recallZ-0237-2022 · OmniLife Science · initiated Oct 1, 2021

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from OmniLife Science or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The device is missing grit blasting on the proximal surface of the femoral stem, which may lead to the possibility of product revision in the event of undersizing.

FDA root cause category: Process control

What the company is doing

Affected agents and representatives were notified by email, with initial communication sent via a quarantine form on 10/1/21. Additional notices titled "Urgent Medical Device Recall" were emailed out on October 15, 2021. Customers were instructed to immediately quarantine lot 38933 and return the product to OMNI. Customers were provided with a response form and asked to complete and return the form to the recalling company by FAX to 508-819-3390 or my email at [email protected]. For further information contact Christina Rovaldi, RA/QA Manager at 774-226-1847 or 800-448-6664. Company contact hours are Monday through Friday, 8:00 AM to 4:30 PM, Eastern Time.

Recall details

Recall number
Z-0237-2022
Status
Terminated
Initiated by the company
Oct 1, 2021
Recalling firm
OmniLife Science
Quantity
7
Product code
LPH
FDA submissions
K041950
Distribution
US Distribution to states of: Colorado and Virginia.

Product description

OMNI K2 Hip Stem, Manufacturer's Product Code 130006 The OMNI K2 Hip Stem is intended for use as the femoral component of a primary total hip replacement. This femoral hip stem is intended for uncemented fixation and single use implantation.

Code information (models, lots, serial numbers)

UDI 00841690102496 Lot 38933

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page