Skip to content

VirtuTRAX Instrument Navigator recallZ-0397-2022 · Civco Medical Instruments Co. Inc. · initiated Oct 8, 2021

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Civco Medical Instruments Co. Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

There is potential that the VirtuTRAX device will not tighten onto the instrument to be used.

FDA root cause category: Device Design

What the company is doing

The firm distributed an "URGENT: MEDICAL DEVICE RECALL" notice on 10/21/2021 by email. The notice explained the problem and requested acknowledgement of the notice. Entities further distributing the products were directed to notify those to whom the devices were distributed. The firm will be replacing the units.

Recall details

Recall number
Z-0397-2022
Status
Terminated
Initiated by the company
Oct 8, 2021
Recalling firm
Civco Medical Instruments Co. Inc.
Quantity
18 boxes (5 units/box)
Product code
IYO
FDA submissions
K092619
Distribution
Worldwide distribution - US Nationwide and the countries of Belgium, Canada, China (Mainland), China (Taiwan), Denmark, France, Germany, Italy, Japan, Netherlands, Singapore, South Korea, Spain, Thailand.

Product description

VirtuTRAX Instrument Navigator, Sensor with 10 17 GA VirtutTRAX

Code information (models, lots, serial numbers)

Part Number 610-1080; UDI 00841436101936; Lot Numbers: A109102, A109175, A121363, A129087, A129088, A131042, A133436, A134220, A138099, A139369, A140674, A141818

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other devices under the same FDA submission

11 devices cleared under K092619 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 88906).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page