The Steripath Gen2 Blood Collection System recallZ-0318-2022 · Magnolia Medical Technologies, Inc. · initiated Nov 2, 2021
Terminated
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Magnolia Medical Technologies, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Due to an uptick of complaints associated with premature needle retraction of blood collection system.
FDA root cause category: Process control
What the company is doing
On 11/02/2021, the Recalling Firm sent an "URGENT: MEDICAL DEVICE RECALL" Notification Letter to customers via email informing them that the firm has received a number of product complaints associated with its blood collecting systems where the retractable feature of the butterfly needles may prematurely activate. The product complaints have been associated with retractable feature being activated within the product packaging. Customers are instructed to: -Determine the quantity of the affected products remaining at their facility and communicated this to the Recalling Firm's Sales or Clinical team members. -The Recalling Firm will replace any of the affected products at no cost to the customer. -Notify all their staff of the recall and if the affected products have been further distributed, to contact/notify those customers/organization of this recall.
Recall details
- Recall number
- Z-0318-2022
- Status
- Terminated
- Initiated by the company
- Nov 2, 2021
- Recalling firm
- Magnolia Medical Technologies, Inc.
- Quantity
- 137,200 devices
- Product code
- JKA
- FDA submissions
- K192247
- Distribution
- U.S. Nationwide distribution in the states of MD, WV, DE, VA, DC, IN, MI, MA, SC, NY, IL, TX, AL, OK, GA, NE, FL, MO, CA, CO, NM, AR, WY, KS, PA, WA, MT, and AZ.
Product description
The Steripath Gen2 Blood Collection System Model Number/Description 2700-21-EN / Steripath Gen2 Blood Collection System - 21G, Wide Neck 2700-23-EN / Steripath Gen2 Blood Collection System - 23G, Wide Neck 27BD-21-EN / Steripath Gen2 Blood Collection System - 21G, Long Neck 27BD-23-EN / Steripath Gen2 Blood Collection System - 23G, Long Neck
Code information (models, lots, serial numbers)
Show the full code information (1k characters)
Model Number/Part Number/Description: 2700-21-EN/P00157-001/Steripath Gen2 Blood Collection System - 21G, Wide Neck UDI Sales Pack: 10858366007013 UDI Sterile Pack: 00858366007016 Lot Numbers: 9447412 9447043 9446903 9447120 9446991 9447413 2700-23-EN/P00158-001/Steripath Gen2 Blood Collection System - 23G, Wide Neck UDI Sales Pack: 10858366007020 UDI Sterile Pack: 00858366007023 Lot Numbers: 9446807 9447124 5005432 9446708 9447121 27BD-21-EN/P00166-000/Steripath Gen2 Blood Collection System - 21G, Long Neck UDI Sales Pack: 10858366007037 UDI Sterile Pack: 00858366007030 Lot Numbers: 5005328 9446696 9446547 9446994 27BD-23-EN/P00167-000/Steripath Gen2 Blood Collection System - 23G, Long Neck UDI Sales Pack: 10858366007044 UDI Sterile Pack: 00858366007047 Lot Numbers: 9447122
Devices named in this recall
4 devices in the FDA device registry whose identifier appears in the recall's code information.
- SteriPath Gen2, 21G 2700-21-ENMagnolia Medical Technologies, Inc. · Tubes, Vials, Systems, Serum Separators, Blood Collection1 recall0 reports
- SteriPath Gen2, 23G 2700-23-ENMagnolia Medical Technologies, Inc. · Tubes, Vials, Systems, Serum Separators, Blood Collection1 recall0 reports
- SteriPath Gen2, 21G, BD 27BD-21-ENMagnolia Medical Technologies, Inc. · Tubes, Vials, Systems, Serum Separators, Blood Collection1 recall0 reports
- SteriPath Gen2, 23G, BD 27BD-23-ENMagnolia Medical Technologies, Inc. · Tubes, Vials, Systems, Serum Separators, Blood Collection1 recall0 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗