BioFire Blood Culture Identification 2 (BCID2) Panel recallZ-0312-2022 · BioFire Diagnostics, LLC · initiated Oct 11, 2021
Terminated
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from BioFire Diagnostics, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Due to an increased risk of false positive Acinetobacter calcoaceticus- baumannii complex results. The cause may be as a due to non-viable contamination in the blood culture vials.
FDA root cause category: Material/Component Contamination
What the company is doing
On or about 10/10/2021, the Recalling Firm sent an "URGENT: MEDICAL DEVICE CORRECTION" Notification Letter to U.S. customer via overnight FedEx to inform them that there is an increased risk of false positive Acinetobacter calcoaceticus- baumannii complex results when the blood culture identification panels are used with some types of blood culture vials. The increased risk are potential due to an increased level of nucleic acid from non-viable Acinetobacter calcoaceticus- baumannii complex targets in the blood culture vials. Affected Blood Culture Media vials are: 442023 - BD BACTEC Plus Aerobic medium in plastic vials; 442021 - BD BACTEC Lytic Anaerobic medium in plastic vials; 442192 - BD BACTEC Plus Aerobic medium in glass vials; 442022 - BD BACTEC Plus Anaerobic medium in plastic vials; 442020 - BD BACTEC Peds Plus mediums in plastic vials; Customers are instructed to: 1) If the BCID2 Panel is used to test BD BACTEC" blood culture vials with an expiration date of April 30, 2022 or earlier, positive results for Acinetobacter calcoaceticus- baumannii complex should be confirmed by another method prior to reporting the test results. 2) Complete the accompanying Field Action Acknowledgment of Receipt Form and return to the Recalling Firm via one of the listed return methods so that the Recalling Firm may acknowledge their receipt of this notification. Any questions or concerns, contact customer support department at [email protected] or via telephone by dialing 1.800.735.6544 and selecting option 5 for Product Technical Support.
Recall details
- Recall number
- Z-0312-2022
- Status
- Terminated
- Initiated by the company
- Oct 11, 2021
- Recalling firm
- BioFire Diagnostics, LLC
- Quantity
- Total 14,140 Kits [11,155 kits (U.S.) and 2,985 kits (O.U.S.)]
- Product code
- PEN
- FDA submissions
- K193519
- Distribution
- Worldwide distribution - U.S. Nationwide distribution in the states of AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV and WY. The Countries of Algeria, Argentina, Australia, Austria, Bahrain, Belarus, Belgium, Bulgaria, Chile, Colombia, Costa Rica, Croatia, Czech Republic, Dominican Rep/, Ecuador, Egypt, France, Georgia, Germany, Greece, Guam, Guatemala, Hong Kong, Hungary, India, Israel Italy, Kuwait, Latvia, Macedonia, Malaysia, Mexico, Netherlands, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russian Fed., Saudi Arabia, Serbia, Shanghai, Slovenia, South Africa, Spain, Sweden, Switzerland, Thailand, Tunisia, Turkey, United Kingdom, Uruguay, Utd/Arab Emir/.
Product description
BioFire Blood Culture Identification 2 (BCID2) Panel Part No: RFIT-ASY-0147 UDI: 00815381020338 Blood Culture Catalog No./Description: 442023/BD BACTEC Plus Aerobic medium in plastic vials 442021/BD BACTEC Lytic Anaerobic medium in plastic vials 442192//BD BACTEC Plus Aerobic medium in glass vials 442022/BD BACTEC Plus Anaerobic medium in plastic vials 442020/BD BACTEC Peds Plus mediums in plastic vials All lots with an expiry date prior to
Code information (models, lots, serial numbers)
Blood Culture Identification 2 (BCID2) Panel Part Number: RFIT-ASY-0147 UDI Code: 00815381020338 Blood Culture Media Catalog No./Description: 442023/BD BACTEC Plus Aerobic medium in plastic vials 442021/BD BACTEC Lytic Anaerobic medium in plastic vials 442192//BD BACTEC Plus Aerobic medium in glass vials 442022/BD BACTEC Plus Anaerobic medium in plastic vials 442020/BD BACTEC Peds Plus mediums in plastic vials All BCID2 Panel lots BD Blood Culture Media vials with expiration date of April 30, 2022 or earlier.
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- BioFire® Blood Culture Identification 2 (BCID2) Panel RFIT-ASY-0147Biofire Diagnostics, LLC · Gram-Positive Bacteria And Their Resistance Markers6 recalls6 deaths555 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page