Endurant II recallZ-0291-2022 · Medtronic Vascular, Inc. · initiated Oct 15, 2021
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Medtronic Vascular, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Due to, during stent graft deployment, the spindle may detach from the distal end of the spindle hypotube.
FDA root cause category: Under Investigation by firm
What the company is doing
On or about 10/15/2021, the firm initiated a global verbal notification to customers informing them that is the potential for a specific subset of 18Fr Stent Graft Systems to have the potential for the spindle to detach from the spindle hypotube during deployment. On 10/18/2021, the firm sent an "URGENT MEDICAL DEVICE RECALL" Notification Letter via UPS 2-day delivery. Customers are instructed to take the following actions: 1) Return all unused affected devices to the Recalling Firm via their sales representative or calling customer service at 1-888-283-7868 or email [email protected]. 2) Complete the enclosed Customer Confirmation Form and email to [email protected]. 3) Forward this Recall notice to all those who need to be aware within their organization to include physician implanters (US only). Additionally, if any affected devices have been distributed to other organizations, forward this notice to those entities. For questions or assistance - contact their Field Representative or Customer Service at 888-283-7868 or email [email protected].
Recall details
- Recall number
- Z-0291-2022
- Status
- Open, Classified
- Initiated by the company
- Oct 15, 2021
- Recalling firm
- Medtronic Vascular, Inc.
- Quantity
- 511 devices
- Product code
- MIH
- FDA submissions
- P100021
- Distribution
- Worldwide distribution - U.S. Nationwide distribution in the states of CA, FL, GA, MN, MO, NC, NJ, NM, NV, NY, PA, SC, TN, TX, VA, and WV. The countries of Austria, Belgium, Bulgaria, China, Cyprus, Czech Republic, Denmark, Finland, France, Germany, Hungary, Ireland, Italy, Japan, Jordan, Lithuania, Luxembourg, Netherlands, Norway, Poland, Portugal, Qatar, South Africa, Spain, Sweden, Switzerland, and United Kingdom.
Product description
Endurant II Stent Graft System REF/Description: ETBF2513C145EE STENT GRAFT ETBF2513C145EE ENDUR II BIF ETBF2820C166EE STENT GRAFT ETBF2820C166EE ENDUR II BIF ETBF2820C166EE STENT GRAFT ETBF2820C166EE ENDUR II BIF ETBF2820C166EE STENT GRAFT ETBF2820C166EE ENDUR II BIF ETBF2820C166EE STENT GRAFT ETBF2820C166EE ENDUR II BIF ETBF2820C166EE STENT GRAFT ETBF2820C166EE ENDUR II BIF ETBF2820C166EE STENT GRAFT ETBF2820C166EE ENDUR II BIF ETBF2820C166EE STENT GRAFT ETBF2820C166EE ENDUR II BIF ETCF2828C49EE STENT GRAFT ETCF2828C49EE ENDUR II CUFF ETCF2828C49EE STENT GRAFT ETCF2828C49EE ENDUR II CUFF ETCF2828C49EE STENT GRAFT ETCF2828C49EE ENDUR II CUFF ETUF2514C102EE STENT GRAFT ETUF2514C102EE ENDUR II AUI ETUF2514C102EE STENT GRAFT ETUF2514C102EE ENDUR II AUI ETUF2514C102EE STENT GRAFT ETUF2514C102EE ENDUR II AUI ETUF2514C102EE STENT GRAFT ETUF2514C102EE ENDUR II AUI ETUF2514C102EE STENT GRAFT ETUF2514C102EE ENDUR II AUI ETUF2514C102EE STENT GRAFT ETUF2514C102EE ENDUR II AUI ETBF2313C124EE STENT GRAFT ETBF2313C124EE ENDUR II BIF ETBF2513C145EE STENT GRAFT ETBF2513C145EE ENDUR II BIF ETBF2513C145EE STENT GRAFT ETBF2513C145EE ENDUR II BIF ETBF2513C145EE STENT GRAFT ETBF2513C145EE ENDUR II BIF ETBF2513C145EE STENT GRAFT ETBF2513C145EE ENDUR II BIF ETBF2516C166EE STENT GRAFT ETBF2516C166EE ENDUR II BIF ETBF2516C166EE STENT GRAFT ETBF2516C166EE ENDUR II BIF ETBF2516C166EE STENT GRAFT ETBF2516C166EE ENDUR II BIF ETBF2813C166EE STENT GRAFT ETBF2813C166EE ENDUR II BIF ETBF2813C166EE STENT GRAFT ETBF2813C166EE ENDUR II BIF ETBF2813C166EE STENT GRAFT ETBF2813C166EE ENDUR II BIF ETBF2813C166EE STENT GRAFT ETBF2813C166EE ENDUR II BIF ETBF2813C166EE STENT GRAFT ETBF2813C166EE ENDUR II BIF ETBF2813C166EE STENT GRAFT ETBF2813C166EE ENDUR II BIF ETBF2813C166EE STENT GRAFT ETBF2813C166EE ENDUR II BIF ETBF2813C166EE STENT GRAFT ETBF2813C166EE ENDUR II BIF ETBF2813C166EE STENT GRAFT ETBF2813C166EE ENDUR II BIF ETBF2816C166EE STENT GRAFT ETBF2816C166EE ENDUR II BIF ETBF2816C166EE STENT GRAFT ETBF2816C166EE ENDUR II BIF ETBF2816C166EE STENT GRAFT ETBF2816C166EE ENDUR II BIF ETBF2816C166EE STENT GRAFT ETBF2816C166EE ENDUR II BIF ETBF2816C166EE STENT GRAFT ETBF2816C166EE ENDUR II BIF ETBF2816C166EE STENT GRAFT ETBF2816C166EE ENDUR II BIF ETBF2816C166EE STENT GRAFT ETBF2816C166EE ENDUR II BIF ETBF2516C145EE STENT GRAFT ETBF2516C145EE ENDUR II BIF ETBF2516C145EE STENT GRAFT ETBF2516C145EE ENDUR II BIF ETBF2516C145EE STENT GRAFT ETBF2516C145EE ENDUR II BIF ETBF2516C145EE STENT GRAFT ETBF2516C145EE ENDUR II BIF ETBF2516C145EE STENT GRAFT ETBF2516C145EE ENDUR II BIF ETBF2516C145EE STENT GRAFT ETBF2516C145EE ENDUR II BIF ETBF2516C145EE STENT GRAFT ETBF2516C145EE ENDUR II BIF ETBF2516C145EE STENT GRAFT ETBF2516C145EE ENDUR II BIF ETUF2514C102E STENT GRAFT ETUF2514C102E ENDUR II AUI ETTF2828C70EE STENT GRAFT ETTF2828C70EE ENDUR II TUBE ETTF2828C70EE STENT GRAFT ETTF2828C70EE ENDUR II TUBE ETTF2828C70EE STENT GRAFT ETTF2828C70EE ENDUR II TUBE ETTF2828C70EE STENT GRAFT ETTF2828C70EE ENDUR II TUBE ETTF2828C70EE STENT GRAFT ETTF2828C70EE ENDUR II TUBE ETTF2828C70EE STENT GRAFT ETTF2828C70EE ENDUR II TUBE ETTF2828C70EE STENT GRAFT ETTF2828C70EE ENDUR II TUBE ETTF2828C70EE STENT GRAFT ETTF2828C70EE ENDUR II TUBE ETCF2525C49E STENT GRAFT ETCF2525C49E ENDUR II AOEXT ETCF2828C49E STENT GRAFT ETCF2828C49E ENDUR II AOEXT ETCF2828C49E STENT GRAFT ETCF2828C49E ENDUR II AOEXT ETCF2828C49E STENT GRAFT ETCF2828C49E ENDUR II AOEXT ETUF2814C102E STENT GRAFT ETUF2814C102E ENDUR II AUI ETUF2814C102E STENT GRAFT ETUF2814C102E ENDUR II AUI ETTF2525C70EE STENT GRAFT ETTF2525C70EE ENDUR II TUBE ETTF2525C70EE STENT GRAFT ETTF2525C70EE ENDUR II TUBE ETTF2525C70EE STENT GRAFT ETTF2525C70EE ENDUR II TUBE ETTF2525C70EE STENT GRAFT ETTF2525C70EE ENDUR II TUBE ETTF2525C70EE STENT GRAFT ETTF2525C70EE ENDUR II TUBE ETTF2525C70EE STENT GRAFT ETTF2525C70EE ENDUR II TUBE ETBF2313C166EE STENT GRAFT ETBF2313C166EE ENDUR II BIF ETBF2513C124EE STENT GRAFT ETBF2513C124EE ENDUR II BIF ET
Code information (models, lots, serial numbers)
Show the full code information (34k characters)
Model Number/GTIN Code/Product Description: ETBF2513C145EE 00643169781047 V30584649 STENT GRAFT ETBF2513C145EE ENDUR II BIF ETBF2820C166EE 00643169781177 V30593589 STENT GRAFT ETBF2820C166EE ENDUR II BIF ETBF2820C166EE 00643169781177 V30593590 STENT GRAFT ETBF2820C166EE ENDUR II BIF ETBF2820C166EE 00643169781177 V30593591 STENT GRAFT ETBF2820C166EE ENDUR II BIF ETBF2820C166EE 00643169781177 V30593592 STENT GRAFT ETBF2820C166EE ENDUR II BIF ETBF2820C166EE 00643169781177 V30593593 STENT GRAFT ETBF2820C166EE ENDUR II BIF ETBF2820C166EE 00643169781177 V30593594 STENT GRAFT ETBF2820C166EE ENDUR II BIF ETBF2820C166EE 00643169781177 V30593595 STENT GRAFT ETBF2820C166EE ENDUR II BIF ETCF2828C49EE 00643169780798 V30595328 STENT GRAFT ETCF2828C49EE ENDUR II CUFF ETCF2828C49EE 00643169780798 V30595331 STENT GRAFT ETCF2828C49EE ENDUR II CUFF ETCF2828C49EE 00643169780798 V30595332 STENT GRAFT ETCF2828C49EE ENDUR II CUFF ETUF2514C102EE 00643169780934 V30595382 STENT GRAFT ETUF2514C102EE ENDUR II AUI ETUF2514C102EE 00643169780934 V30595383 STENT GRAFT ETUF2514C102EE ENDUR II AUI ETUF2514C102EE 00643169780934 V30595386 STENT GRAFT ETUF2514C102EE ENDUR II AUI ETUF2514C102EE 00643169780934 V30595389 STENT GRAFT ETUF2514C102EE ENDUR II AUI ETUF2514C102EE 00643169780934 V30595390 STENT GRAFT ETUF2514C102EE ENDUR II AUI ETUF2514C102EE 00643169780934 V30595391 STENT GRAFT ETUF2514C102EE ENDUR II AUI ETBF2313C124EE 00643169780972 V30595456 STENT GRAFT ETBF2313C124EE ENDUR II BIF ETBF2513C145EE 00643169781047 V30595505 STENT GRAFT ETBF2513C145EE ENDUR II BIF ETBF2513C145EE 00643169781047 V30595515 STENT GRAFT ETBF2513C145EE ENDUR II BIF ETBF2513C145EE 00643169781047 V30595516 STENT GRAFT ETBF2513C145EE ENDUR II BIF ETBF2513C145EE 00643169781047 V30595517 STENT GRAFT ETBF2513C145EE ENDUR II BIF ETBF2516C166EE 00643169781085 V30596585 STENT GRAFT ETBF2516C166EE ENDUR II BIF ETBF2516C166EE 00643169781085 V30596586 STENT GRAFT ETBF2516C166EE ENDUR II BIF ETBF2516C166EE 00643169781085 V30596587 STENT GRAFT ETBF2516C166EE ENDUR II BIF ETBF2813C166EE 00643169781115 V30596625 STENT GRAFT ETBF2813C166EE ENDUR II BIF ETBF2813C166EE 00643169781115 V30596626 STENT GRAFT ETBF2813C166EE ENDUR II BIF ETBF2813C166EE 00643169781115 V30596634 STENT GRAFT ETBF2813C166EE ENDUR II BIF ETBF2813C166EE 00643169781115 V30596635 STENT GRAFT ETBF2813C166EE ENDUR II BIF ETBF2813C166EE 00643169781115 V30596637 STENT GRAFT ETBF2813C166EE ENDUR II BIF ETBF2813C166EE 00643169781115 V30596639 STENT GRAFT ETBF2813C166EE ENDUR II BIF ETBF2813C166EE 00643169781115 V30596640 STENT GRAFT ETBF2813C166EE ENDUR II BIF ETBF2813C166EE 00643169781115 V30596641 STENT GRAFT ETBF2813C166EE ENDUR II BIF ETBF2813C166EE 00643169781115 V30596644 STENT GRAFT ETBF2813C166EE ENDUR II BIF ETBF2816C166EE 00643169781146 V30596645 STENT GRAFT ETBF2816C166EE ENDUR II BIF ETBF2816C166EE 00643169781146 V30596646 STENT GRAFT ETBF2816C166EE ENDUR II BIF ETBF2816C166EE 00643169781146 V30596648 STENT GRAFT ETBF2816C166EE ENDUR II BIF ETBF2816C166EE 00643169781146 V30596649 STENT GRAFT ETBF2816C166EE ENDUR II BIF ETBF2816C166EE 00643169781146 V30596650 STENT GRAFT ETBF2816C166EE ENDUR II BIF ETBF2816C166EE 00643169781146 V30596651 STENT GRAFT ETBF2816C166EE ENDUR II BIF ETBF2816C166EE 00643169781146 V30596655 STENT GRAFT ETBF2816C166EE ENDUR II BIF ETBF2516C145EE 00643169781078 V30603296 STENT GRAFT ETBF2516C145EE ENDUR II BIF ETBF2516C145EE 00643169781078 V30603298 STENT GRAFT ETBF2516C145EE ENDUR II BIF ETBF2516C145EE 00643169781078 V30603309 STENT GRAFT ETBF2516C145EE ENDUR II BIF ETBF2516C145EE 00643169781078 V30603310 STENT GRAFT ETBF2516C145EE ENDUR II BIF ETBF2516C145EE 00643169781078 V30603311 STENT GRAFT ETBF2516C145EE ENDUR II BIF ETBF2516C145EE 00643169781078 V30603312 STENT GRAFT ETBF2516C145EE ENDUR II BIF ETBF2516C145EE 00643169781078 V30603313 STENT GRAFT ETBF2516C145EE ENDUR II BIF ETBF2516C145EE 00643169781078 V30603315 STENT GRAFT ETBF2516C145EE ENDUR II BIF ETUF2514C102E 00643169232570 V30608080 STENT GRAFT ETUF2514C102E ENDUR II AUI ETTF2828C70EE 00643169780897 V30609554 STENT GRAFT ETTF2828C70EE ENDUR II TUBE ETTF2828C70EE 00643169780897 V30609556 STENT GRAFT ETTF2828C70EE ENDUR II TUBE ETTF2828C70EE 00643169780897 V30609557 STENT GRAFT ETTF2828C70EE ENDUR II TUBE ETTF2828C70EE 00643169780897 V30609558 STENT GRAFT ETTF2828C70EE ENDUR II TUBE ETTF2828C70EE 00643169780897 V30609559 STENT GRAFT ETTF2828C70EE ENDUR II TUBE ETTF2828C70EE 00643169780897 V30609560 STENT GRAFT ETTF2828C70EE ENDUR II TUBE ETTF2828C70EE 00643169780897 V30609561 STENT GRAFT ETTF2828C70EE ENDUR II TUBE ETTF2828C70EE 00643169780897 V30609564 STENT GRAFT ETTF2828C70EE ENDUR II TUBE ETCF2525C49E 00613994991430 V30616355 STENT GRAFT ETCF2525C49E ENDUR II AOEXT ETCF2828C49E 00613994991447 V30616364 STENT GRAFT ETCF2828C49E ENDUR II AOEXT ETCF2828C49E 00613994991447 V30616365 STENT GRAFT ETCF2828C49E ENDUR II AOEXT ETCF2828C49E 00613994991447 V30616368 STENT GRAFT ETCF2828C49E ENDUR II AOEXT ETUF2814C102E 00643169232594 V30616643 STENT GRAFT ETUF2814C102E ENDUR II AUI ETUF2814C102E 00643169232594 V30616646 STENT GRAFT ETUF2814C102E ENDUR II AUI ETTF2525C70EE 00643169780880 V30616793 STENT GRAFT ETTF2525C70EE ENDUR II TUBE ETTF2525C70EE 00643169780880 V30616794 STENT GRAFT ETTF2525C70EE ENDUR II TUBE ETTF2525C70EE 00643169780880 V30616795 STENT GRAFT ETTF2525C70EE ENDUR II TUBE ETTF2525C70EE 00643169780880 V30616796 STENT GRAFT ETTF2525C70EE ENDUR II TUBE ETTF2525C70EE 00643169780880 V30616798 STENT GRAFT ETTF2525C70EE ENDUR II TUBE ETTF2525C70EE 00643169780880 V30616799 STENT GRAFT ETTF2525C70EE ENDUR II TUBE ETBF2313C166EE 00643169780132 V30617526 STENT GRAFT ETBF2313C166EE ENDUR II BIF ETBF2513C124EE 00643169781030 V30617789 STENT GRAFT ETBF2513C124EE ENDUR II BIF ETBF2513C124EE 00643169781030 V30617798 STENT GRAFT ETBF2513C124EE ENDUR II BIF ETTF2323C70EE 00643169780873 V30619008 STENT GRAFT ETTF2323C70EE ENDUR II TUBE ETTF2323C70EE 00643169780873 V30619017 STENT GRAFT ETTF2323C70EE ENDUR II TUBE ETTF2323C70EE 00643169780873 V30619021 STENT GRAFT ETTF2323C70EE ENDUR II TUBE ETTF2323C70EE 00643169780873 V30619022 STENT GRAFT ETTF2323C70EE ENDUR II TUBE ETTF2828C70EE 00643169780897 V30619025 STENT GRAFT ETTF2828C70EE ENDUR II TUBE ETTF2828C70EE 00643169780897 V30619027 STENT GRAFT ETTF2828C70EE ENDUR II TUBE ETTF2828C70EE 00643169780897 V30619029 STENT GRAFT ETTF2828C70EE ENDUR II TUBE ETTF2828C70EE 00643169780897 V30619030 STENT GRAFT ETTF2828C70EE ENDUR II TUBE ETUF2814C102EE 00643169780941 V30619053 STENT GRAFT ETUF2814C102EE ENDUR II AUI ETUF2814C102EE 00643169780941 V30619054 STENT GRAFT ETUF2814C102EE ENDUR II AUI ETUF2814C102EE 00643169780941 V30619055 STENT GRAFT ETUF2814C102EE ENDUR II AUI ETUF2814C102EE 00643169780941 V30619056 STENT GRAFT ETUF2814C102EE ENDUR II AUI ETUF2814C102EE 00643169780941 V30619057 STENT GRAFT ETUF2814C102EE ENDUR II AUI ETUF2814C102EE 00643169780941 V30619058 STENT GRAFT ETUF2814C102EE ENDUR II AUI ETUF2814C102EE 00643169780941 V30619059 STENT GRAFT ETUF2814C102EE ENDUR II AUI ETUF2814C102EE 00643169780941 V30619060 STENT GRAFT ETUF2814C102EE ENDUR II AUI ETUF2814C102EE 00643169780941 V30619061 STENT GRAFT ETUF2814C102EE ENDUR II AUI ETUF2814C102EE 00643169780941 V30619062 STENT GRAFT ETUF2814C102EE ENDUR II AUI ETUF2814C102EE 00643169780941 V30619063 STENT GRAFT ETUF2814C102EE ENDUR II AUI ETUF2814C102EE 00643169780941 V30619064 STENT GRAFT ETUF2814C102EE ENDUR II AUI ETUF2814C102EE 00643169780941 V30619065 STENT GRAFT ETUF2814C102EE ENDUR II AUI ETUF2814C102EE 00643169780941 V30619066 STENT GRAFT ETUF2814C102EE ENDUR II AUI ETUF2814C102EE 00643169780941 V30619067 STENT GRAFT ETUF2814C102EE ENDUR II AUI ETUF2814C102EE 00643169780941 V30619068 STENT GRAFT ETUF2814C102EE ENDUR II AUI ETUF2814C102EE 00643169780941 V30619070 STENT GRAFT ETUF2814C102EE ENDUR II AUI ETUF2814C102EE 00643169780941 V30619071 STENT GRAFT ETUF2814C102EE ENDUR II AUI ETUF2814C102EE 00643169780941 V30619072 STENT GRAFT ETUF2814C102EE ENDUR II AUI ETBF2813C124EE 00643169781092 V30619093 STENT GRAFT ETBF2813C124EE ENDUR II BIF ETBF2813C124EE 00643169781092 V30619096 STENT GRAFT ETBF2813C124EE ENDUR II BIF ETCF2525C49EE 00763000099206 V30619397 STENT GRAFT ETCF2525C49EE ENDUR II CUFF ETCF2828C49EE 00763000099213 V30619457 STENT GRAFT ETCF2828C49EE ENDUR II CUFF ETCF2828C49EE 00763000099213 V30619458 STENT GRAFT ETCF2828C49EE ENDUR II CUFF ETCF2828C49EE 00763000099213 V30619459 STENT GRAFT ETCF2828C49EE ENDUR II CUFF ETCF2828C49EE 00763000099213 V30619460 STENT GRAFT ETCF2828C49EE ENDUR II CUFF ETCF2828C49EE 00763000099213 V30619465 STENT GRAFT ETCF2828C49EE ENDUR II CUFF ETUF2314C102EE 00763000099343 V30622674 STENT GRAFT ETUF2314C102EE ENDUR II AUI ETUF2314C102EJ 00643169366442 V30623217 STENT GRAFT ETUF2314C102EJ ENDUR II AUI ETTF2525C70EE 00643169780880 V30623451 STENT GRAFT ETTF2525C70EE ENDUR II TUBE ETTF2525C70EE 00643169780880 V30623452 STENT GRAFT ETTF2525C70EE ENDUR II TUBE ETTF2525C70EE 00643169780880 V30623453 STENT GRAFT ETTF2525C70EE ENDUR II TUBE ETTF2525C70EE 00643169780880 V30623459 STENT GRAFT ETTF2525C70EE ENDUR II TUBE ETTF2828C70EE 00643169780897 V30623470 STENT GRAFT ETTF2828C70EE ENDUR II TUBE ETTF2828C70EE 00643169780897 V30623471 STENT GRAFT ETTF2828C70EE ENDUR II TUBE ETBF2516C145EE 00643169781078 V30623502 STENT GRAFT ETBF2516C145EE ENDUR II BIF ETBF2516C145EE 00643169781078 V30623503 STENT GRAFT ETBF2516C145EE ENDUR II BIF ETBF2516C145EE 00643169781078 V30623505 STENT GRAFT ETBF2516C145EE ENDUR II BIF ETBF2516C145EE 00643169781078 V30623506 STENT GRAFT ETBF2516C145EE ENDUR II BIF ETBF2816C145EE 00643169781139 V30623522 STENT GRAFT ETBF2816C145EE ENDUR II BIF ETBF2816C145EE 00643169781139 V30623523 STENT GRAFT ETBF2816C145EE ENDUR II BIF ETBF2816C145EE 00643169781139 V30623524 STENT GRAFT ETBF2816C145EE ENDUR II BIF ETBF2816C145EE 00643169781139 V30623525 STENT GRAFT ETBF2816C145EE ENDUR II BIF ETBF2816C145EE 00643169781139 V30623527 STENT GRAFT ETBF2816C145EE ENDUR II BIF ETBF2816C145EE 00643169781139 V30623539 STENT GRAFT ETBF2816C145EE ENDUR II BIF ETBF2816C145EE 00643169781139 V30623540 STENT GRAFT ETBF2816C145EE ENDUR II BIF ETTF2828C70EJ 00643169215719 V30624649 STENT GRAFT ETTF2828C70EJ ENDUR II AOEXT ETCF2525C49EE 00643169780781 V30625031 STENT GRAFT ETCF2525C49EE ENDUR II CUFF ETCF2525C49EE 00643169780781 V30625032 STENT GRAFT ETCF2525C49EE ENDUR II CUFF ETCF2525C49EE 00643169780781 V30625035 STENT GRAFT ETCF2525C49EE ENDUR II CUFF ETCF2525C49EE 00643169780781 V30625036 STENT GRAFT ETCF2525C49EE ENDUR II CUFF ETCF2525C49EE 00643169780781 V30625037 STENT GRAFT ETCF2525C49EE ENDUR II CUFF ETCF2525C49EE 00643169780781 V30625038 STENT GRAFT ETCF2525C49EE ENDUR II CUFF ETCF2525C49EE 00643169780781 V30625039 STENT GRAFT ETCF2525C49EE ENDUR II CUFF ETCF2525C49EE 00643169780781 V30625040 STENT GRAFT ETCF2525C49EE ENDUR II CUFF ETTF2323C70EE 00643169780873 V30625049 STENT GRAFT ETTF2323C70EE ENDUR II TUBE ETTF2323C70EE 00643169780873 V30625050 STENT GRAFT ETTF2323C70EE ENDUR II TUBE ETTF2323C70EE 00643169780873 V30625052 STENT GRAFT ETTF2323C70EE ENDUR II TUBE ETTF2323C70EE 00643169780873 V30625053 STENT GRAFT ETTF2323C70EE ENDUR II TUBE ETTF2323C70EE 00643169780873 V30625054 STENT GRAFT ETTF2323C70EE ENDUR II TUBE ETTF2323C70EE 00643169780873 V30625055 STENT GRAFT ETTF2323C70EE ENDUR II TUBE ETTF2323C70EE 00643169780873 V30625056 STENT GRAFT ETTF2323C70EE ENDUR II TUBE ETTF2323C70EE 00643169780873 V30625057 STENT GRAFT ETTF2323C70EE ENDUR II TUBE ETTF2323C70EE 00643169780873 V30625058 STENT GRAFT ETTF2323C70EE ENDUR II TUBE ETTF2323C70EE 00643169780873 V30625066 STENT GRAFT ETTF2323C70EE ENDUR II TUBE ETTF2323C70EE 00643169780873 V30625067 STENT GRAFT ETTF2323C70EE ENDUR II TUBE ETTF2323C70EE 00643169780873 V30625068 STENT GRAFT ETTF2323C70EE ENDUR II TUBE ETUF2514C102EE 00643169780934 V30625071 STENT GRAFT ETUF2514C102EE ENDUR II AUI ETUF2514C102EE 00643169780934 V30625073 STENT GRAFT ETUF2514C102EE ENDUR II AUI ETUF2514C102EE 00643169780934 V30625076 STENT GRAFT ETUF2514C102EE ENDUR II AUI ETUF2514C102EE 00643169780934 V30625078 STENT GRAFT ETUF2514C102EE ENDUR II AUI ETUF2514C102EE 00643169780934 V30625079 STENT GRAFT ETUF2514C102EE ENDUR II AUI ETUF2514C102EE 00643169780934 V30625080 STENT GRAFT ETUF2514C102EE ENDUR II AUI ETUF2514C102EE 00643169780934 V30625081 STENT GRAFT ETUF2514C102EE ENDUR II AUI ETUF2514C102EE 00643169780934 V30625082 STENT GRAFT ETUF2514C102EE ENDUR II AUI ETUF2514C102EE 00643169780934 V30625083 STENT GRAFT ETUF2514C102EE ENDUR II AUI ETUF2814C102EE 00643169780941 V30625096 STENT GRAFT ETUF2814C102EE ENDUR II AUI ETUF2814C102EE 00643169780941 V30625097 STENT GRAFT ETUF2814C102EE ENDUR II AUI ETUF2814C102EE 00643169780941 V30625638 STENT GRAFT ETUF2814C102EE ENDUR II AUI ETUF2814C102EE 00643169780941 V30625639 STENT GRAFT ETUF2814C102EE ENDUR II AUI ETUF2814C102EE 00643169780941 V30625641 STENT GRAFT ETUF2814C102EE ENDUR II AUI ETUF2814C102EE 00643169780941 V30625642 STENT GRAFT ETUF2814C102EE ENDUR II AUI ETUF2814C102EE 00643169780941 V30625644 STENT GRAFT ETUF2814C102EE ENDUR II AUI ETUF2814C102EE 00643169780941 V30625645 STENT GRAFT ETUF2814C102EE ENDUR II AUI ETUF2814C102EE 00643169780941 V30625646 STENT GRAFT ETUF2814C102EE ENDUR II AUI ETUF2814C102EE 00643169780941 V30625647 STENT GRAFT ETUF2814C102EE ENDUR II AUI ETUF2814C102EE 00643169780941 V30625648 STENT GRAFT ETUF2814C102EE ENDUR II AUI ETUF2814C102EE 00643169780941 V30625649 STENT GRAFT ETUF2814C102EE ENDUR II AUI ETUF2814C102EE 00643169780941 V30625650 STENT GRAFT ETUF2814C102EE ENDUR II AUI ETUF2814C102EE 00643169780941 V30625651 STENT GRAFT ETUF2814C102EE ENDUR II AUI ETUF2814C102EE 00643169780941 V30625653 STENT GRAFT ETUF2814C102EE ENDUR II AUI ETUF2814C102EE 00643169780941 V30625655 STENT GRAFT ETUF2814C102EE ENDUR II AUI ETUF2814C102EE 00643169780941 V30625656 STENT GRAFT ETUF2814C102EE ENDUR II AUI ETBF2516C124EE 00643169781061 V30625678 STENT GRAFT ETBF2516C124EE ENDUR II BIF ETBF2516C124EE 00643169781061 V30625679 STENT GRAFT ETBF2516C124EE ENDUR II BIF ETBF2516C124EE 00643169781061 V30625681 STENT GRAFT ETBF2516C124EE ENDUR II BIF ETBF2516C124EE 00643169781061 V30625682 STENT GRAFT ETBF2516C124EE ENDUR II BIF ETBF2516C124EE 00643169781061 V30625683 STENT GRAFT ETBF2516C124EE ENDUR II BIF ETBF2516C145EE 00643169781078 V30625688 STENT GRAFT ETBF2516C145EE ENDUR II BIF ETBF2516C145EE 00643169781078 V30625689 STENT GRAFT ETBF2516C145EE ENDUR II BIF ETBF2816C145EE 00643169781139 V30625708 STENT GRAFT ETBF2816C145EE ENDUR II BIF ETBF2816C145EE 00643169781139 V30625709 STENT GRAFT ETBF2816C145EE ENDUR II BIF ETBF2816C145EE 00643169781139 V30625711 STENT GRAFT ETBF2816C145EE ENDUR II BIF ETBF2816C145EE 00643169781139 V30625712 STENT GRAFT ETBF2816C145EE ENDUR II BIF ETBF2816C145EE 00643169781139 V30625713 STENT GRAFT ETBF2816C145EE ENDUR II BIF ETBF2816C145EE 00643169781139 V30625714 STENT GRAFT ETBF2816C145EE ENDUR II BIF ETBF2816C145EE 00643169781139 V30625715 STENT GRAFT ETBF2816C145EE ENDUR II BIF ETBF2816C145EE 00643169781139 V30625716 STENT GRAFT ETBF2816C145EE ENDUR II BIF ETBF2816C145EE 00643169781139 V30625717 STENT GRAFT ETBF2816C145EE ENDUR II BIF ETBF2816C145EE 00643169781139 V30625718 STENT GRAFT ETBF2816C145EE ENDUR II BIF ETBF2816C145EE 00643169781139 V30625720 STENT GRAFT ETBF2816C145EE ENDUR II BIF ETBF2816C145EE 00643169781139 V30625721 STENT GRAFT ETBF2816C145EE ENDUR II BIF ETBF2816C145EE 00643169781139 V30625723 STENT GRAFT ETBF2816C145EE ENDUR II BIF ETBF2816C145EE 00643169781139 V30625724 STENT GRAFT ETBF2816C145EE ENDUR II BIF ETBF2816C145EE 00643169781139 V30625725 STENT GRAFT ETBF2816C145EE ENDUR II BIF ETBF2816C145EE 00643169781139 V30625729 STENT GRAFT ETBF2816C145EE ENDUR II BIF ETBF2816C145EE 00643169781139 V30625730 STENT GRAFT ETBF2816C145EE ENDUR II BIF ETBF2816C145EE 00643169781139 V30625732 STENT GRAFT ETBF2816C145EE ENDUR II BIF ETBF2816C145EE 00643169781139 V30625733 STENT GRAFT ETBF2816C145EE ENDUR II BIF ETBF2816C145EE 00643169781139 V30625734 STENT GRAFT ETBF2816C145EE ENDUR II BIF ETBF2816C145EE 00643169781139 V30625735 STENT GRAFT ETBF2816C145EE ENDUR II BIF ETBF2816C145EE 00643169781139 V30625736 STENT GRAFT ETBF2816C145EE ENDUR II BIF ETBF2816C145EE 00643169781139 V30625737 STENT GRAFT ETBF2816C145EE ENDUR II BIF ETBF2816C145EE 00643169781139 V30625739 STENT GRAFT ETBF2816C145EE ENDUR II BIF ETBF2816C145EE 00643169781139 V30625740 STENT GRAFT ETBF2816C145EE ENDUR II BIF ETBF2816C145EE 00643169781139 V30625741 STENT GRAFT ETBF2816C145EE ENDUR II BIF ETBF2816C145EE 00643169781139 V30625743 STENT GRAFT ETBF2816C145EE ENDUR II BIF ETBF2816C145EE 00643169781139 V30625744 STENT GRAFT ETBF2816C145EE ENDUR II BIF ETBF2820C145EE 00643169781160 V30625768 STENT GRAFT ETBF2820C145EE ENDUR II BIF ETBF2820C145EE 00643169781160 V30625774 STENT GRAFT ETBF2820C145EE ENDUR II BIF ETBF2820C145EE 00643169781160 V30625775 STENT GRAFT ETBF2820C145EE ENDUR II BIF ETBF2820C145EE 00643169781160 V30625776 STENT GRAFT ETBF2820C145EE ENDUR II BIF ETBF2820C145EE 00643169781160 V30625777 STENT GRAFT ETBF2820C145EE ENDUR II BIF ETUF2514C102EE 00763000099350 V30625785 STENT GRAFT ETUF2514C102EE ENDUR II AUI ETUF2814C102EE 00643169780941 V30625830 STENT GRAFT ETUF2814C102EE ENDUR II AUI ETUF2814C102EE 00643169780941 V30625835 STENT GRAFT ETUF2814C102EE ENDUR II AUI ETUF2814C102EE 00643169780941 V30625839 STENT GRAFT ETUF2814C102EE ENDUR II AUI ETUF2814C102EE 00643169780941 V30625840 STENT GRAFT ETUF2814C102EE ENDUR II AUI ETUF2814C102EE 00643169780941 V30625841 STENT GRAFT ETUF2814C102EE ENDUR II AUI ETUF2814C102EE 00643169780941 V30625842 STENT GRAFT ETUF2814C102EE ENDUR II AUI ETUF2814C102EE 00643169780941 V30625843 STENT GRAFT ETUF2814C102EE ENDUR II AUI ETUF2814C102EE 00643169780941 V30625844 STENT GRAFT ETUF2814C102EE ENDUR II AUI ETUF2814C102EE 00643169780941 V30625845 STENT GRAFT ETUF2814C102EE ENDUR II AUI ETUF2814C102EE 00643169780941 V30625846 STENT GRAFT ETUF2814C102EE ENDUR II AUI ETUF2814C102EE 00643169780941 V30625847 STENT GRAFT ETUF2814C102EE ENDUR II AUI ETUF2814C102EE 00643169780941 V30625848 STENT GRAFT ETUF2814C102EE ENDUR II AUI ETUF2814C102EE 00643169780941 V30625849 STENT GRAFT ETUF2814C102EE ENDUR II AUI ETCF2828C49E 00613994991447 V30625864 STENT GRAFT ETCF2828C49E ENDUR II AOEXT ETBF2313C166EE 00643169780996 V30625997 STENT GRAFT ETBF2313C166EE ENDUR II BIF ETBF2313C166EE 00643169780996 V30625998 STENT GRAFT ETBF2313C166EE ENDUR II BIF ETBF2313C166EE 00643169780996 V30626001 STENT GRAFT ETBF2313C166EE ENDUR II BIF ETBF2313C166EE 00643169780996 V30626002 STENT GRAFT ETBF2313C166EE ENDUR II BIF ETBF2313C166EE 00643169780996 V30626003 STENT GRAFT ETBF2313C166EE ENDUR II BIF ETBF2313C166EE 00643169780996 V30626004 STENT GRAFT ETBF2313C166EE ENDUR II BIF ETBF2513C166EE 00643169781054 V30626005 STENT GRAFT ETBF2513C166EE ENDUR II BIF ETBF2513C166EE 00643169781054 V30626006 STENT GRAFT ETBF2513C166EE ENDUR II BIF ETBF2513C166EE 00643169781054 V30626007 STENT GRAFT ETBF2513C166EE ENDUR II BIF ETBF2513C166EE 00643169781054 V30626008 STENT GRAFT ETBF2513C166EE ENDUR II BIF ETBF2513C166EE 00643169781054 V30626009 STENT GRAFT ETBF2513C166EE ENDUR II BIF ETBF2816C124EE 00643169781122 V30626015 STENT GRAFT ETBF2816C124EE ENDUR II BIF ETBF2816C124EE 00643169781122 V30626016 STENT GRAFT ETBF2816C124EE ENDUR II BIF ETBF2816C124EE 00643169781122 V30626019 STENT GRAFT ETBF2816C124EE ENDUR II BIF ETBF2816C124EE 00643169781122 V30626022 STENT GRAFT ETBF2816C124EE ENDUR II BIF ETBF2816C124EE 00643169781122 V30626023 STENT GRAFT ETBF2816C124EE ENDUR II BIF ETBF2816C124EE 00643169781122 V30626024 STENT GRAFT ETBF2816C124EE ENDUR II BIF ETBF2816C166EE 00643169781146 V30626026 STENT GRAFT ETBF2816C166EE ENDUR II BIF ETBF2816C166EE 00643169781146 V30626027 STENT GRAFT ETBF2816C166EE ENDUR II BIF ETBF2816C166EE 00643169781146 V30626028 STENT GRAFT ETBF2816C166EE ENDUR II BIF ETBF2816C166EE 00643169781146 V30626029 STENT GRAFT ETBF2816C166EE ENDUR II BIF ETBF2816C166EE 00643169781146 V30626030 STENT GRAFT ETBF2816C166EE ENDUR II BIF ETBF2816C166EE 00643169781146 V30626031 STENT GRAFT ETBF2816C166EE ENDUR II BIF ETBF2816C166EE 00643169781146 V30626032 STENT GRAFT ETBF2816C166EE ENDUR II BIF ETBF2816C166EE 00643169781146 V30626034 STENT GRAFT ETBF2816C166EE ENDUR II BIF ETBF2816C166EE 00643169781146 V30626035 STENT GRAFT ETBF2816C166EE ENDUR II BIF ETBF2816C166EE 00643169781146 V30626036 STENT GRAFT ETBF2816C166EE ENDUR II BIF ETBF2816C166EE 00643169781146 V30627691 STENT GRAFT ETBF2816C166EE ENDUR II BIF ETBF2816C166EE 00643169781146 V30627692 STENT GRAFT ETBF2816C166EE ENDUR II BIF ETBF2816C166EE 00643169781146 V30627693 STENT GRAFT ETBF2816C166EE ENDUR II BIF ETBF2816C166EE 00643169781146 V30627694 STENT GRAFT ETBF2816C166EE ENDUR II BIF ETBF2816C166EE 00643169781146 V30627695 STENT GRAFT ETBF2816C166EE ENDUR II BIF ETBF2816C166EE 00643169781146 V30627696 STENT GRAFT ETBF2816C166EE ENDUR II BIF ETBF2816C166EE 00643169781146 V30627697 STENT GRAFT ETBF2816C166EE ENDUR II BIF ETBF2816C166EE 00643169781146 V30627698 STENT GRAFT ETBF2816C166EE ENDUR II BIF ETBF2816C166EE 00643169781146 V30627699 STENT GRAFT ETBF2816C166EE ENDUR II BIF ETBF2816C166EE 00643169781146 V30627700 STENT GRAFT ETBF2816C166EE ENDUR II BIF ETUF2514C102E 00643169232570 V30628609 STENT GRAFT ETUF2514C102E ENDUR II AUI ETUF2514C102E 00643169232570 V30628610 STENT GRAFT ETUF2514C102E ENDUR II AUI ETUF2514C102E 00643169232570 V30628614 STENT GRAFT ETUF2514C102E ENDUR II AUI ETTF2525C70EE 00643169780880 V30628777 STENT GRAFT ETTF2525C70EE ENDUR II TUBE ETTF2525C70EE 00643169780880 V30628778 STENT GRAFT ETTF2525C70EE ENDUR II TUBE ETTF2525C70EE 00643169780880 V30628779 STENT GRAFT ETTF2525C70EE ENDUR II TUBE ETTF2525C70EE 00643169780880 V30628780 STENT GRAFT ETTF2525C70EE ENDUR II TUBE ETTF2525C70EE 00643169780880 V30628782 STENT GRAFT ETTF2525C70EE ENDUR II TUBE ETTF2525C70EE 00643169780880 V30628783 STENT GRAFT ETTF2525C70EE ENDUR II TUBE ETTF2525C70EE 00643169780880 V30628784 STENT GRAFT ETTF2525C70EE ENDUR II TUBE ETTF2525C70EE 00643169780880 V30628785 STENT GRAFT ETTF2525C70EE ENDUR II TUBE ETUF2314C102EE 00643169780927 V30628797 STENT GRAFT ETUF2314C102EE ENDUR II AUI ETUF2314C102EE 00643169780927 V30628798 STENT GRAFT ETUF2314C102EE ENDUR II AUI ETUF2314C102EE 00643169780927 V30628799 STENT GRAFT ETUF2314C102EE ENDUR II AUI ETUF2314C102EE 00643169780927 V30628800 STENT GRAFT ETUF2314C102EE ENDUR II AUI ETUF2314C102EE 00643169780927 V30628802 STENT GRAFT ETUF2314C102EE ENDUR II AUI ETUF2314C102EE 00643169780927 V30628804 STENT GRAFT ETUF2314C102EE ENDUR II AUI ETUF2314C102EE 00643169780927 V30628805 STENT GRAFT ETUF2314C102EE ENDUR II AUI ETUF2314C102EE 00643169780927 V30628806 STENT GRAFT ETUF2314C102EE ENDUR II AUI ETUF2314C102EE 00643169780927 V30628808 STENT GRAFT ETUF2314C102EE ENDUR II AUI ETUF2314C102EE 00643169780927 V30628809 STENT GRAFT ETUF2314C102EE ENDUR II AUI ETUF2314C102EE 00643169780927 V30628810 STENT GRAFT ETUF2314C102EE ENDUR II AUI ETUF2314C102EE 00643169780927 V30628811 STENT GRAFT ETUF2314C102EE ENDUR II AUI ETUF2514C102EE 00643169780934 V30628812 STENT GRAFT ETUF2514C102EE ENDUR II AUI ETUF2514C102EE 00643169780934 V30628815 STENT GRAFT ETUF2514C102EE ENDUR II AUI ETUF2514C102EE 00643169780934 V30628816 STENT GRAFT ETUF2514C102EE ENDUR II AUI ETUF2514C102EE 00643169780934 V30628817 STENT GRAFT ETUF2514C102EE ENDUR II AUI ETUF2514C102EE 00643169780934 V30628818 STENT GRAFT ETUF2514C102EE ENDUR II AUI ETUF2514C102EE 00643169780934 V30628819 STENT GRAFT ETUF2514C102EE ENDUR II AUI ETUF2514C102EE 00643169780934 V30628820 STENT GRAFT ETUF2514C102EE ENDUR II AUI ETUF2514C102EE 00643169780934 V30628821 STENT GRAFT ETUF2514C102EE ENDUR II AUI ETUF2514C102EE 00643169780934 V30628822 STENT GRAFT ETUF2514C102EE ENDUR II AUI ETUF2514C102EE 00643169780934 V30628823 STENT GRAFT ETUF2514C102EE ENDUR II AUI ETUF2514C102EE 00643169780934 V30628824 STENT GRAFT ETUF2514C102EE ENDUR II AUI ETUF2514C102EE 00643169780934 V30628825 STENT GRAFT ETUF2514C102EE ENDUR II AUI ETUF2514C102EE 00643169780934 V30628826 STENT GRAFT ETUF2514C102EE ENDUR II AUI ETBF2313C145EE 00643169780989 V30628844 STENT GRAFT ETBF2313C145EE ENDUR II BIF ETBF2313C145EE 00643169780989 V30628845 STENT GRAFT ETBF2313C145EE ENDUR II BIF ETBF2313C145EE 00643169780989 V30628846 STENT GRAFT ETBF2313C145EE ENDUR II BIF ETBF2316C166EE 00643169781023 V30628847 STENT GRAFT ETBF2316C166EE ENDUR II BIF ETBF2316C166EE 00643169781023 V30628857 STENT GRAFT ETBF2316C166EE ENDUR II BIF ETBF2316C166EE 00643169781023 V30628858 STENT GRAFT ETBF2316C166EE ENDUR II BIF ETBF2316C166EE 00643169781023 V30628859 STENT GRAFT ETBF2316C166EE ENDUR II BIF ETBF2316C166EE 00643169781023 V30628860 STENT GRAFT ETBF2316C166EE ENDUR II BIF ETBF2513C145EE 00643169781047 V30628862 STENT GRAFT ETBF2513C145EE ENDUR II BIF ETBF2513C145EE 00643169781047 V30628863 STENT GRAFT ETBF2513C145EE ENDUR II BIF ETBF2513C145EE 00643169781047 V30628864 STENT GRAFT ETBF2513C145EE ENDUR II BIF ETBF2513C145EE 00643169781047 V30628865 STENT GRAFT ETBF2513C145EE ENDUR II BIF ETBF2513C145EE 00643169781047 V30628866 STENT GRAFT ETBF2513C145EE ENDUR II BIF ETBF2516C145EE 00643169781078 V30628883 STENT GRAFT ETBF2516C145EE ENDUR II BIF ETBF2516C145EE 00643169781078 V30628891 STENT GRAFT ETBF2516C145EE ENDUR II BIF ETBF2516C145EE 00643169781078 V30628892 STENT GRAFT ETBF2516C145EE ENDUR II BIF ETBF2516C145EE 00643169781078 V30628893 STENT GRAFT ETBF2516C145EE ENDUR II BIF ETBF2516C145EE 00643169781078 V30628894 STENT GRAFT ETBF2516C145EE ENDUR II BIF ETBF2516C145EE 00643169781078 V30628895 STENT GRAFT ETBF2516C145EE ENDUR II BIF ETBF2516C145EE 00643169781078 V30628896 STENT GRAFT ETBF2516C145EE ENDUR II BIF ETBF2516C145EE 00643169781078 V30628897 STENT GRAFT ETBF2516C145EE ENDUR II BIF ETBF2516C145EE 00643169781078 V30628898 STENT GRAFT ETBF2516C145EE ENDUR II BIF ETBF2816C166EE 00643169781146 V30629151 STENT GRAFT ETBF2816C166EE ENDUR II BIF ETBF2816C166EE 00643169781146 V30629152 STENT GRAFT ETBF2816C166EE ENDUR II BIF ETBF2816C166EE 00643169781146 V30629153 STENT GRAFT ETBF2816C166EE ENDUR II BIF ETBF2816C166EE 00643169781146 V30629154 STENT GRAFT ETBF2816C166EE ENDUR II BIF ETBF2816C166EE 00643169781146 V30629155 STENT GRAFT ETBF2816C166EE ENDUR II BIF ETBF2516C145EE 00763000099497 V30629230 STENT GRAFT ETBF2516C145EE ENDUR II BIF ETTF2323C70EJ 00643169215696 V30629692 STENT GRAFT ETTF2323C70EJ ENDUR II AOEXT ETTF2323C70EJ 00643169215696 V30629693 STENT GRAFT ETTF2323C70EJ ENDUR II AOEXT ETTF2323C70EJ 00643169215696 V30629694 STENT GRAFT ETTF2323C70EJ ENDUR II AOEXT ETCF2828C49EE 00643169780798 V30629776 STENT GRAFT ETCF2828C49EE ENDUR II CUFF ETCF2828C49EE 00643169780798 V30629777 STENT GRAFT ETCF2828C49EE ENDUR II CUFF ETCF2828C49EE 00643169780798 V30629778 STENT GRAFT ETCF2828C49EE ENDUR II CUFF ETCF2828C49EE 00643169780798 V30629779 STENT GRAFT ETCF2828C49EE ENDUR II CUFF ETCF2828C49EE 00643169780798 V30629782 STENT GRAFT ETCF2828C49EE ENDUR II CUFF ETCF2828C49EE 00643169780798 V30629783 STENT GRAFT ETCF2828C49EE ENDUR II CUFF ETCF2828C49EE 00643169780798 V30629784 STENT GRAFT ETCF2828C49EE ENDUR II CUFF ETCF2828C49EE 00643169780798 V30629785 STENT GRAFT ETCF2828C49EE ENDUR II CUFF ETCF2828C49EE 00643169780798 V30629786 STENT GRAFT ETCF2828C49EE ENDUR II CUFF ETCF2828C49EE 00643169780798 V30629788 STENT GRAFT ETCF2828C49EE ENDUR II CUFF ETCF2828C49EE 00643169780798 V30629791 STENT GRAFT ETCF2828C49EE ENDUR II CUFF ETCF2828C49EE 00643169780798 V30629792 STENT GRAFT ETCF2828C49EE ENDUR II CUFF ETCF2828C49EE 00643169780798 V30629793 STENT GRAFT ETCF2828C49EE ENDUR II CUFF ETCF2828C49EE 00643169780798 V30629794 STENT GRAFT ETCF2828C49EE ENDUR II CUFF ETTF2323C70EE 00643169780873 V30629796 STENT GRAFT ETTF2323C70EE ENDUR II TUBE ETTF2323C70EE 00643169780873 V30629798 STENT GRAFT ETTF2323C70EE ENDUR II TUBE ETTF2323C70EE 00643169780873 V30629799 STENT GRAFT ETTF2323C70EE ENDUR II TUBE ETTF2323C70EE 00643169780873 V30629800 STENT GRAFT ETTF2323C70EE ENDUR II TUBE ETTF2323C70EE 00643169780873 V30629801 STENT GRAFT ETTF2323C70EE ENDUR II TUBE ETTF2323C70EE 00643169780873 V30629804 STENT GRAFT ETTF2323C70EE ENDUR II TUBE ETTF2323C70EE 00643169780873 V30629807 STENT GRAFT ETTF2323C70EE ENDUR II TUBE ETTF2323C70EE 00643169780873 V30629808 STENT GRAFT ETTF2323C70EE ENDUR II TUBE ETTF2323C70EE 00643169780873 V30629809 STENT GRAFT ETTF2323C70EE ENDUR II TUBE ETTF2323C70EE 00643169780873 V30629810 STENT GRAFT ETTF2323C70EE ENDUR II TUBE ETBF2313C166EE 00643169780996 V30629910 STENT GRAFT ETBF2313C166EE ENDUR II BIF ETBF2313C166EE 00643169780996 V30629911 STENT GRAFT ETBF2313C166EE ENDUR II BIF ETBF2313C166EE 00643169780996 V30629912 STENT GRAFT ETBF2313C166EE ENDUR II BIF ETBF2313C166EE 00643169780996 V30629913 STENT GRAFT ETBF2313C166EE ENDUR II BIF ETBF2816C145EE 00643169781139 V30630160 STENT GRAFT ETBF2816C145EE ENDUR II BIF ETBF2816C145EE 00643169781139 V30630161 STENT GRAFT ETBF2816C145EE ENDUR II BIF ETBF2816C145EE 00643169781139 V30630162 STENT GRAFT ETBF2816C145EE ENDUR II BIF ETBF2816C145EE 00643169781139 V30630163 STENT GRAFT ETBF2816C145EE ENDUR II BIF ETBF2816C145EE 00643169781139 V30630164 STENT GRAFT ETBF2816C145EE ENDUR II BIF ETBF2816C145EE 00643169781139 V30630165 STENT GRAFT ETBF2816C145EE ENDUR II BIF ETBF2816C145EE 00643169781139 V30630166 STENT GRAFT ETBF2816C145EE ENDUR II BIF ETBF2816C145EE 00643169781139 V30630176 STENT GRAFT ETBF2816C145EE ENDUR II BIF ETBF2813C145EE 00643169781108 V30631612 STENT GRAFT ETBF2813C145EE ENDUR II BIF ETBF2813C145EE 00643169781108 V30631614 STENT GRAFT ETBF2813C145EE ENDUR II BIF ETBF2813C145EE 00643169781108 V30631617 STENT GRAFT ETBF2813C145EE ENDUR II BIF ETBF2813C145EE 00643169781108 V30631619 STENT GRAFT ETBF2813C145EE ENDUR II BIF ETBF2813C145EE 00643169781108 V30631621 STENT GRAFT ETBF2813C145EE ENDUR II BIF ETBF2813C145EE 00643169781108 V30631622 STENT GRAFT ETBF2813C145EE ENDUR II BIF Serial Numbers: V30584649 V30593589 V30593590 V30593591 V30593592 V30593593 V30593594 V30593595 V30595328 V30595331 V30595332 V30595382 V30595383 V30595386 V30595389 V30595390 V30595391 V30595456 V30595505 V30595515 V30595516 V30595517 V30596585 V30596586 V30596587 V30596625 V30596626 V30596634 V30596635 V30596637 V30596639 V30596640 V30596641 V30596644 V30596645 V30596646 V30596648 V30596649 V30596650 V30596651 V30596655 V30603296 V30603298 V30603309 V30603310 V30603311 V30603312 V30603313 V30603315 V30608080 V30609554 V30609556 V30609557 V30609558 V30609559 V30609560 V30609561 V30609564 V30616355 V30616364 V30616365 V30616368 V30616643 V30616646 V30616793 V30616794 V30616795 V30616796 V30616798 V30616799 V30617526 V30617789 V30617798 V30619008 V30619017 V30619021 V30619022 V30619025 V30619027 V30619029 V30619030 V30619053 V30619054 V30619055 V30619056 V30619057 V30619058 V30619059 V30619060 V30619061 V30619062 V30619063 V30619064 V30619065 V30619066 V30619067 V30619068 V30619070 V30619071 V30619072 V30619093 V30619096 V30619397 V30619457 V30619458 V30619459 V30619460 V30619465 V30622674 V30623217 V30623451 V30623452 V30623453 V30623459 V30623470 V30623471 V30623502 V30623503 V30623505 V30623506 V30623522 V30623523 V30623524 V30623525 V30623527 V30623539 V30623540 V30624649 V30625031 V30625032 V30625035 V30625036 V30625037 V30625038 V30625039 V30625040 V30625049 V30625050 V30625052 V30625053 V30625054 V30625055 V30625056 V30625057 V30625058 V30625066 V30625067 V30625068 V30625071 V30625073 V30625076 V30625078 V30625079 V30625080 V30625081 V30625082 V30625083 V30625096 V30625097 V30625638 V30625639 V30625641 V30625642 V30625644 V30625645 V30625646 V30625647 V30625648 V30625649 V30625650 V30625651 V30625653 V30625655 V30625656 V30625678 V30625679 V30625681 V30625682 V30625683 V30625688 V30625689 V30625708 V30625709 V30625711 V30625712 V30625713 V30625714 V30625715 V30625716 V30625717 V30625718 V30625720 V30625721 V30625723 V30625724 V30625725 V30625729 V30625730 V30625732 V30625733 V30625734 V30625735 V30625736 V30625737 V30625739 V30625740 V30625741 V30625743 V30625744 V30625768 V30625774 V30625775 V30625776 V30625777 V30625785 V30625830 V30625835 V30625839 V30625840 V30625841 V30625842 V30625843 V30625844 V30625845 V30625846 V30625847 V30625848 V30625849 V30625864 V30625997 V30625998 V30626001 V30626002 V30626003 V30626004 V30626005 V30626006 V30626007 V30626008 V30626009 V30626015 V30626016 V30626019 V30626022 V30626023 V30626024 V30626026 V30626027 V30626028 V30626029 V30626030 V30626031 V30626032 V30626034 V30626035 V30626036 V30627691 V30627692 V30627693 V30627694 V30627695 V30627696 V30627697 V30627698 V30627699 V30627700 V30628609 V30628610 V30628614 V30628777 V30628778 V30628779 V30628780 V30628782 V30628783 V30628784 V30628785 V30628797 V30628798 V30628799 V30628800 V30628802 V30628804 V30628805 V30628806 V30628808 V30628809 V30628810 V30628811 V30628812 V30628815 V30628816 V30628817 V30628818 V30628819 V30628820 V30628821 V30628822 V30628823 V30628824 V30628825 V30628826 V30628844 V30628845 V30628846 V30628847 V30628857 V30628858 V30628859 V30628860 V30628862 V30628863 V30628864 V30628865 V30628866 V30628883 V30628891 V30628892 V30628893 V30628894 V30628895 V30628896 V30628897 V30628898 V30629151 V30629152 V30629153 V30629154 V30629155 V30629230 V30629692 V30629693 V30629694 V30629776 V30629777 V30629778 V30629779 V30629782 V30629783 V30629784 V30629785 V30629786 V30629788 V30629791 V30629792 V30629793 V30629794 V30629796 V30629798 V30629799 V30629800 V30629801 V30629804 V30629807 V30629808 V30629809 V30629810 V30629910 V30629911 V30629912 V30629913 V30630160 V30630161 V30630162 V30630163 V30630164 V30630165 V30630166 V30630176 V30631612 V30631614 V30631617 V30631619 V30631621 V30631622
Devices named in this recall
4 devices in the FDA device registry whose identifier appears in the recall's code information.
- Endurant® II ETCF2828C49EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls7 deaths44 reports
- Endurant® II ETUF2814C102EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls8 deaths37 reports
- Endurant® II ETUF2514C102EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls3 deaths13 reports
- Endurant® II ETCF2525C49EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls1 death11 reports
Product lines
Other devices under the same FDA submission
166 devices cleared under P100021 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- Endurant® IIs ESBF2814C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls34 deaths254 reports
- Endurant® IIs ESBF3214C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls19 deaths173 reports
- Endurant® IIs ESBF3614C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT1 recall37 deaths214 reports
- Endurant® IIs ESBF2514C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls18 deaths136 reports
- Endurant® IIs ESBF2314C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls9 deaths60 reports
- Endurant® II ETUF3614C102EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls16 deaths36 reports
- Endurant® II ETBF3216C145EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls2 deaths32 reports
- Endurant® II ETCF3636C49EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT1 recall15 deaths77 reports
- Endurant® II ETBF2513C166EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls1 death22 reports
- Endurant® II ETBF2516C145EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls3 deaths17 reports
- Endurant® II ETBF2516C166EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls17 reports
- Endurant® II ETBF2313C124EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls3 deaths7 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 88951).
- Z-0290-2022 · Open, ClassifiedEndurant Stent Graft System REF/Description: ENTF2525C70EE STENT GRAFT ENTF2525C70EE ENDURANT TUBE
- Z-0292-2022 · Open, ClassifiedEndurant IIs Stent Graft System REF/Description: ESBF2814C103EE STENT GRAFT ESBF2814C103EE ENDUR IIS BIF ESBF2814C103EE STENT GRAFT ESBF2814C103EE ENDUR IIS BIF ESBF2814C103EE STEN…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page