Leica Microsystems M220 F12 Microscope Systems recallZ-0412-2022 · Leica Microsystems, Inc. · initiated Nov 3, 2021
Terminated
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Leica Microsystems, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
In May 2021, Leica Microsystems received one (1) complaint pertaining to the M220 F12 Surgical Microscope system. There was no patient injury reported for this complaint. However, the identified defect carries some potential injury risk. In the event that this defect may occur, the M220 optics carrier may unintentionally drop into the surgical field, risking contact with the patient.
FDA root cause category: Process control
What the company is doing
Leica Microsystems is issuing this field safety notice as a corrective action for all Leica M220 F12 Surgical Microscopes that were manufactured since market introduction in January 2009 until March 9th, 2017. Leica Microsystems is initiating a field safety corrective action to provide a retention fixture to be added on all Leica M220 F12 Surgical Microscopes manufactured before March 9th, 2017 to ensure patient safety and customer satisfaction. A Leica Microsystems representative will provide the consignee with a retention fixture as an additional means of securing the M220 Optics Carrier. Prior to each use of the M220 F12, consignees should perform the following mandatory inspection: If the gap between the black adapter and the white housing at the end of the swing arm becomes bigger than 3mm or a free rotation of the optics carrier is possible for more than 360: Immediately stop using the device. If you notice any anomalies with your M220 F12 Microscope System before the additional retention fixture is attached to your instrument, please contact a Leica representative for relevant repair of your instrument. If the mandatory inspection performed prior using your M220 F12 does not result in an observed potential defect, you can continue to use the system pending fixation of the retention fixture. Consignees must acknowledge receipt of this Field Safety Notice with the digital acknowledgment form by either scanning the QR code in your mobile phone or with the following link: https://forms.office.com/r/kZfZnwpxbA As an alternative you can also acknowledge receipt of this Field Safety Notice by completing, signing, dating and returning the attached Acknowledgement Form via an email to the following address: [email protected] This Field Safety Notice must be distributed to all those who need to be aware within your organization or to any organization where the potentially affected devices have been transferred to. Please maintain awarene
Recall details
- Recall number
- Z-0412-2022
- Status
- Terminated
- Initiated by the company
- Nov 3, 2021
- Recalling firm
- Leica Microsystems, Inc.
- Quantity
- 2,800
- Product code
- FSO
- Distribution
- Worldwide distribution - US Nationwide distribution in the states of CA, DC, FL, GA, IL, IN, MA, MI, MD, NY, OH, OR, PA, TN, TX, VA, WI, and Puerto Rico. The countries of Albania, Algeria, Argentina, Armenia, Australia, Austria, Bangladesh, Belgium, Brazil, Bulgaria, Burundi, Canada, Chile, China, Colombia, Congo, Costa Rica, Dominican Republic, Ecuador, Egypt, France, Germany, Georgia, Great Britain, Greece, Hong Kong, India, Indonesia, Iraq, Ireland, Italy, Japan, South Korea, Kuwait, Latvia, Lebanon, Libya, Malaysia, Mexico, Morocco, Myanmar, Netherlands, New Zealand, Nicaragua, Nigeria, Norway, Pakistan, Palestine, Peru, Philippines, Poland, Portugal, North Macedonia, Romania, Russia, Saudi Arabia, Senegal, Serbia, Singapore, Slovakia, South Africa, Spain, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Thailand, Turkey, Ukraine, United Arab Emirates, Uruguay, Vietnam.
Product description
Leica Microsystems M220 F12 Microscope Systems
Code information (models, lots, serial numbers)
Show the full code information (1k characters)
Part #: 10448367 UDI #: 07630003552188 U.S. Serial #'s: 180709002, 191109001, 261016002, 250815001, 280813001, 251111006, 61209001, 260810003, 240815001, 171116001, 190815001, 151209004, 261016001, 70715001, 60713002, 140510004, 180712001, 300910003, 161013001, 270612001, 140510005, 250512002, 10712002, 130913001, 51112001, 290414001, 260311001, 180515001, 270813001, 231014001, 110614001, 210814001, 180515002, 111212001, 270812001, 100614001, 171214002, 290514003, 101212001, 291013001, 50614002, 81012001, 30514001, 101212002, 40611002, 50614003, 171116002, 200314001, 170212001, 30311002, 150211001, 130913002, 60210005, 100212001, 290509001, 111113001, 120615003, 100212002, 300411006, 110315001, 210612002, 100212003, 210612001, 30614001, 20816001, 100212004, 260816001, 260816002, 291113001, 100212005, 21109002, 260811001, 100810004, 90112002 Canada Serial #'s: 260311010, 240414001, 70414002
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- M220 F12 000000000010448367Leica Microsystems (Schweiz) AG · MICROSCOPE, SURGICAL, GENERAL & PLASTIC SURGERY1 recall0 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page