Skip to content

EVOS recallZ-0363-2022 · Smith & Nephew, Inc. · initiated Nov 1, 2021

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Smith & Nephew, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Testing vendor notified Smith & Nephew that previous MRI scanning conditions provided in Exponents reports were erroneous. Erroneous data from the reports was incorporated in the released IFU 81110294 EVOS Large Fragment Plating System.

FDA root cause category: Under Investigation by firm

What the company is doing

The firm initiated the correction by email and letter on 1 NOV 2021. The notice explained the problem and requested that the consignees check their inventory and complete the Inventory Correction Certificate form. Distributors were directed to notify their customers of the field action and ensure that these actions are carried out.

Recall details

Recall number
Z-0363-2022
Status
Terminated
Initiated by the company
Nov 1, 2021
Recalling firm
Smith & Nephew, Inc.
Quantity
38 units
Product code
HRS
FDA submissions
K201918
Distribution
US Nationwide distribution in the states of CO, FL, GA, MA, MN, MO, NC, NJ, NV, NY, OH, OK, PA, TN, TX, WA.

Product description

smith&Nephew EVOS MEDIAL DISTAL FEMUR PLATE, L, 5H, 90MM, REF 72573101

Code information (models, lots, serial numbers)

Batch Number: 20HM10249, 20HM10250, 20HM10251, 20HM22783, UDI (01) 00885556752999 (17) 310630 (10) 12345678 (11) 2210630 (21) 87654321

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other devices under the same FDA submission

5 devices cleared under K201918 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 89031).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page