BV Endura with Software Release 2.3- A mobile, diagnostic… recallZ-0355-2022 · Philips North America Llc · initiated Nov 12, 2021
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Philips North America Llc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Instructions for Use do not specify the maximum surface temperature of the X-ray tank (assembly housing the X-ray tube, beam filter & cooling oil) and the image intensifier/detector of the system as required an result in burns
FDA root cause category: Device Design
What the company is doing
Philips issued an Urgent Medical Device Correction letter in the USA on 11/12/21 via Certified mailing United States Postal Service. Recalling outside the US is done through the local Philips organizations in the respective countries Letter states reason for recall, health risk and action to take: Pay attention to the Heat Indication Level Icons shown in the Viewing Station to know the temperature of the X-Ray Tank. o Follow the instructions provided in Annex I of this letter for a safe use of the system. o Circulate this notice to all users of this device so that they are aware of the issue. o Place this Field safety Notice with the documentation of the of your system o Transfer this Field Safety notice within your organization or to any organization, if the affected devices have been transferred. o Complete the enclosed Customer Reply Form and send it back to confirm that users have reviewed and understood this Urgent Field Safety Notice. 5. Actions planned by Philips IGT Systems to correct the problem Philips is informing customers through this Urgent Field Safety Notice and providing the Addendum to the Instructions For Use of the BV Endura Release 2.3, BV Pulsera Release 2.3 and Veradius Unity Release 2.1 systems.
Recall details
- Recall number
- Z-0355-2022
- Status
- Open, Classified
- Initiated by the company
- Nov 12, 2021
- Recalling firm
- Philips North America Llc
- Quantity
- 22 units US
- Product code
- OWB
- FDA submissions
- K010435
- Distribution
- Worldwide distribution - US Nationwide and the countries of Afghanistan, Algeria, American Samoa, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belgium, Bolivia, Brazil, Bulgaria, Cameroon, Chile, China, Colombia, Cook Islands, Costa Rica, Croatia, Cuba, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Estonia, Faroe Islands, Finland, France, French Guiana, Georgia, Germany, Ghana, Gibraltar, Greece, Guadeloupe, Guatemala, Guyana, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Korea, Republic of Kuwait, Lao People's Democratic Republic, Latvia, Lebanon, Libya, Lithuania, Luxembourg, Macao, Macedonia, Malaysia, Maldives Mali, Martinique, Mauritius, Mexico, Moldova, Monaco, Mongolia, Morocco, Myanmar, Namibia, Nepal, Netherlands, New Zealand, Nicaragua, Nigeria, Norway, Oman, Pakistan, Palestine, State of Panama, Papua New Guinea, Peru, Philippines, Poland, Portugal, Puerto Rico, Qatar, R¿union Romania, Russian Federation, Rwanda, Saudi Arabia, Senegal, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sudan, Sweden, Switzerland, Taiwan, Tanzania, Thailand, Tunisia, Turkey, Turkmenistan, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Uzbekistan, Venezuela, Viet Nam, Yemen, Zambia, Zimbabwe.
Product description
BV Endura with Software Release 2.3- A mobile, diagnostic X-ray image acquisition and viewing system. Model Number: 718075
Code information (models, lots, serial numbers)
Show the full code information (1k characters)
Serial Number: 173 105 5301 4950 5360 4651 4891 873 1633 160 543 562 204 5090 5091 5092 4840 5073 125 383 384 5121 UDI: (01)00884838055490(21)173 (01)00884838055490(21)105 (01)00884838055490(21)5301 (01)00884838055490(21)4950 (01)00884838055490(21)5360 (01)00884838055490(21)4651 (01)00884838055490(21)4891 (01)00884838055490(21)873 (01)00884838055490(21)1633 (01)00884838055490(21)160 (01)00884838055490(21)543 (01)00884838055490(21)562 (01)00884838055490(21)204 (01)00884838055490(21)5090 (01)00884838055490(21)5091 (01)00884838055490(21)5092 (01)00884838055490(21)4840 (01)00884838055490(21)5073 (01)00884838055490(21)125 (01)00884838055490(21)383 (01)00884838055490(21)384 (01)00884838055490(21)5121
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- BV EnduraPhilips Medical Systems Nederland B.V. · Interventional Fluoroscopic X-Ray System1 recall1 report
Other products in the same recall event
The FDA groups related products into one recall event (event 89129).
- Z-0356-2022 · Open, ClassifiedBV Pulsera with Software Release 2.3-A mobile, diagnostic X-ray image acquisition and viewing system. Model Number: 718095
- Z-0357-2022 · Open, ClassifiedVeradius Unity with Software Release 2.1- Amobile, diagnostic X-ray imaging and viewing system. . Model Number: 718132
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.