Electri-Cool II Portable Cold Therapy Unit, Model Number… recallZ-0496-2022 · Gentherm Medical, LLC · initiated Nov 22, 2021
Class II Terminated
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Gentherm Medical, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
This recall has been initiated to provide a labeling update with revised instruction how to operate the unit with Cold Therapy pads.
FDA root cause category: Device Design
What the company is doing
URGENT Medical Device Recall notification letters dated 11/19/21 were sent to customers. Pad Instructions: 1. Read and understand operation manual before connecting paCI to Etectri-Cool II. 2. Read and understand pad application instructions accompanying GENTHERM pad. 3. Connect hose to unit by inserting connectors on hose into connectors on unit. Connectors should "click" as they lock together. 4. Apply pad to patient and attach pad to hose by inserting connectors on pad into hose connectors. Normal use of the pad is to be applied in direct contact with the patient. Pad may be used over or under a patient, in the flat state or wrapped around patient. 5. Turn unit "ON" and set to temperature. 6. After pad has filled, check water level in unit. Add sterile water or water that has been passed through a filter less than or equal to 0.22 microns, only if necessary. 7. When therapy needs to be paused, disconnect the pad from the Electri-Cool II and allow the unit to continue to run. NOTE: Reconnect the pad when ready to resume therapy. 8. When therapy is no longer needed disconnect pad from hose by pressing metal clips. Dispose of pad. 9. Tum unit "OFF" at the power switch and unplug the power cord from the outlet. INSTRUCTIONS TO CUSTOMERS: 1) Immediately examine your inventory and add this letter and/or Appendix B to Electri-Cool II manual. 2) Ensure that all users are informed of the contents of this letter. If you have further distributed this product, please provide those accounts with a copy of this notice. 3) Please complete and return the enclosed response form as soon as possible to acknowledge receipt of this notification and to inform Gentherm Medical, LLC that you have performed and completed the requested actions. Return the form by e-mail to [email protected], or mail to: Gentherm Medical, LLC 12011 Mosteller Road Cincinnati, OH 45241 This recall should be carried out to the user level. Your assistance is appreciated and necessary to reduce potent
Recall details
- Recall number
- Z-0496-2022
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Nov 22, 2021
- Recalling firm
- Gentherm Medical, LLC
- Quantity
- 484 devices
- Product code
- ILO
- Distribution
- Worldwide distribution - US Nationwide and the countries of Canada, Chile, Hong Kong, Korea, UAE.
Product description
Electri-Cool II Portable Cold Therapy Unit, Model Number 767, Part Number 86193
Code information (models, lots, serial numbers)
UDI- 10613031861938; Affected Serial Numbers are: (164-76703084 Through 213-76703694)
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Electri-Cool ll 767Gentherm Medical, LLC · Pack, Hot Or Cold, Water Circulating1 recall0 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page