Artis icono floor with software VE20C-diagnostic imaging… recallZ-0517-2022 · Siemens Medical Solutions USA, Inc · initiated Oct 12, 2021
Terminated
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Siemens Medical Solutions USA, Inc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Four potential software issues with Artis pheno and Artis icono systems with software VE20C: 1. Updated calibration data not saved with measurement after scene+/-; 2. No x-ray possible, system shutdown/restart might be required during intervention; 3. Corrupted Image during Roadmap; 4. Unintended shutdown of Imaging System with UPS (Uninterruptable Power Supply) option
FDA root cause category: Software design
What the company is doing
Siemens issued A Customer Safety Advisory Notice dated 10/12/21 to affected customers via AX047/21/S. Additionally, a software update will be initiated for all affected users via Update Instruction AX046/21/S. Letter states reason for recall, health risk and action to take: To avoid inaccurate measurement values, repeat the distance measurements after changing the Auto TOD Calibration or Distance Calibration of the scene. Additionally, before closing the image after performing the calibration, create a screenshot/secondary capture image for post review. The Operator Manual of the affected systems will be updated according to the information mentioned above. This is regarded as an interim solution until the next software update is available. The Operator Manual update will draw attention to the issue. We do not consider it necessary to re-examine any patients in relation to the issues (issue 1 to 4) described above. If measurements have already been performed in the past for diagnostics, please verify the results and diagnostic evaluation if applicable. Siemens will correct the software error via Update Instruction AX046/21/S. Our service organization will contact you shortly to arrange a date to perform this corrective action. Please feel free to contact our service organization for an earlier appointment at 1-800-888-7436.
Recall details
- Recall number
- Z-0517-2022
- Status
- Terminated
- Initiated by the company
- Oct 12, 2021
- Recalling firm
- Siemens Medical Solutions USA, Inc
- Quantity
- 23 units
- Product code
- OWB
- FDA submissions
- K193326
- Distribution
- US Nationwide distribution.
Product description
Artis icono floor with software VE20C-diagnostic imaging angiography system Model: 11327700
Code information (models, lots, serial numbers)
Serial Numbers: 170319 170026 170343 170305 170307 170331 170309 170310 170023 170304 170328 170334 170336 170025 170346 170051 170333 170332 170312 170027 170303 170322 170053 UDI: 04056869149325
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- ARTIS icono floor 11327700Siemens Healthcare GmbH · System, x-ray, tomography, computed15 recalls24 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 89260).
- Z-0515-2022 · TerminatedArtis pheno with software VE20C- diagnostic imaging angiography system Model: 10849000
- Z-0516-2022 · TerminatedArtis icono biplane with software VE20C-diagnostic imaging angiography system Model: 11327600
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page