OXY-1 System, Model Number 001-0500-001. Used during… recallZ-0550-2022 · Abiomed, Inc. · initiated Dec 14, 2021

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Abiomed, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The firm has received complaints of the OXY-1 System Console experiencing a power interruption while in use, which disrupts therapy delivered to a patient on support. Disruption of therapy could lead to prolonged hypoxia.

FDA root cause category: Software design (manufacturing process)

What the company is doing

The firm began notifying customers by phone starting December 14, 2021. The firm followed up by sending Urgent Medical Device Recall letters to customers on January 20, 2022. The firm is requesting that customers return their devices. The firm does not recommend placing new patients on support. For patients currently on support with an affected system, the firm recommends physicians use their clinical judgement on whether it is appropriate to continue a patient on the OXY-1 System or to transition that patient to the next step in their clinical journey. The letter included instructions on how to return impacted product. If you have questions on returning product, If you have any questions, please contact Diane Welsh from Abiomed s customer service team at (978) 646-1467 or email [email protected]. The firm's Clinical field staff will be available to coordinate boxing and shipping of the OXY-1 system for return to Abiomed. If you have questions regarding this recall, please contact Ralph Barisano directly at (978) 882-8068.

Recall details

Recall number
Z-0550-2022
Status
Terminated
Initiated by the company
Dec 14, 2021
Posted by the FDA
Feb 3, 2022
Recalling firm
Abiomed, Inc.
Quantity
42 Units
Product code
DTZ
FDA submissions
K200109
Distribution
US distribution to FL, GA, IN, MD, NC, PA, TX.

Product description

OXY-1 System, Model Number 001-0500-001. Used during cardiopulmonary bypass. OXY-1 System consists of Console, Pump Drive, E Drive, and Pole Mount. The disposable Pump Lung Unit PLU part number 001-0400-100 is sold separately and is not part of this recall.

Code information (models, lots, serial numbers)

UDI 00860001797909 Software Version Number V-1.3.15.0, V-1.4.6.0 and V.1.4.8.0 Serial Numbers 20200231-04 20200249-01 20200249-05 20200309-02 20200339-01 20200339-03 20200352-01 20200442-01 20200442-02 20200442-05 20210768-01 20210795-02 20210870-01 20210880-01 20210890-01 20210894-01 20210895-01 20210896-01 20210897-01 20210795-04

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.