QC Sets and Panels: KWIK-STIK 2 Pack recallZ-0589-2022 · Microbiologics Inc · initiated Feb 18, 2020
Terminated
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Microbiologics Inc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Tests returning incorrect susceptible results. High Level Gentamicin Synergy value was out of specification (sensitive) when analyzed on the Vitek 2 AST-GP67 card, and these lots failed the disk diffusion method used for product release.
FDA root cause category: Process design
What the company is doing
Microbiologics notified customers on about 02/18/2020 via letter titled "Urgent Medical Device Recall." Instructions included to inspect inventory for affected products. If found, instructions included the following: 1. REVIEW lab procedures to understand how thisinformation affects your usage; 2. USE OR DISCARD depending on your lab procedures and how this information affects your usage; 3. COMPLETE the response form provided; 4. RETURN the response form to [email protected]; 5. KEEP this letter for your records; 6. CONTACT Microbiologics if a replacement kit is needed. Questions or concerns can be directed to Technical Support team at 320.229.7045 or [email protected].
Recall details
- Recall number
- Z-0589-2022
- Status
- Terminated
- Initiated by the company
- Feb 18, 2020
- Posted by the FDA
- Feb 8, 2022
- Recalling firm
- Microbiologics Inc
- Quantity
- 3 units
- Product code
- JTR
- FDA submissions
- K861022
- Distribution
- Worldwide Distribution. US nationwide including Puerto Rico and Guam, Australia, Austria, Belgium, Brazil, Canada, Chile, China, Colombia, Costa Rica, Croatia/Hrvatska, Curacao, Denmark, Ecuador, Egypt, El Salvador, Finland, France, Germany, Guatemala, Hong Kong, India, Ireland, Israel, Italy, Japan, Democratic People's Republic of Korea, Republic of Korea, Kuwait, Lebanon, Malaysia, Netherlands, New Zealand, Panama, Peru, Poland, Portugal, Qatar, Romania, Saudi Arabia, Singapore, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Trinidad and Tobago, Turkey, United Arab Emirates, United Kingdom, and Vietnam.
Product description
QC Sets and Panels, KWIK-STIK Blood Culture Identification (BCID) Verification Panel (Live Culture). KWIK-STIK kit that contains Catalog Number 0959P. In Vitro Diagnostic Control microorganisms
Code information (models, lots, serial numbers)
Catalog Number: 5229P; Lot Numbers (Expiration Date): 5229-01 (07/31/2020), 5229-02 (09/30/2020), 5229-03 (09/30/2020); UDI: 70845357038899
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- QC Sets and Panels: KWIK-STIK™ 2 Pack 5229PMicrobiologics Inc. · Multi-analyte controls unassayed1 recall0 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 89401).
- Z-0585-2022 · TerminatedKWIK-STIK, Quality control kit for culture media, sold as: a. KWIK-STIK 2-pack Enterococcus faecalis derived from ATCC 51299 b. KWIK-STIK 6-pack Enterococcus faecalis derive…
- Z-0586-2022 · TerminatedLYFO-DISK Enterococcus faecalis derived from ATCC 51299 In Vitro Diagnostic Control microorganisms
- Z-0587-2022 · TerminatedQC Sets and Panels, KWIK-STIK AST-GP (4 Strains) QC Set. KWIK-STIK kit that contains Catalog Number 0959P. In Vitro Diagnostic Control microorganisms
- Z-0588-2022 · TerminatedQC Sets and Panels, KWIK-STIK AST-GP (6 Strains) QC Set. KWIK-STIK kit that contains Catalog Number 0959P. In Vitro Diagnostic Control microorganisms
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page