ARROW recallZ-0629-2022 · Teleflex Medical Inc · initiated Jan 11, 2022
Terminated
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Teleflex Medical Inc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The firm received reports indicating PTD tip separation during use.
FDA root cause category: Process change control
What the company is doing
Arrow International LLC/Teleflex notified their consignees by letter on 01/11/2022. The letter explained the issue and requested that the consignee cease use and distribution. Distributors were directed to notify their customers. The firm is seeking return of the affected product. The following instructions were provided: Action list number 1 Medical facilities 1. We request that you check your inventory for product within the scope of this FSCA. Users should cease use and distribution of impacted product and quarantine immediately. 2. If you do not have stock in scope of this FSCA mark the according checkbox on the Acknowledgement Form (Appendix 1) and return the form to the fax number or e-mail address mentioned below. 3. If you do have stock in scope of this FSCA, complete the Acknowledgement Form (Appendix 1) and fax it to 1-855-419-8507, Attn: Customer Service or e-mail to [email protected]. This will allow us to document the amount of product you have on hand for return. A customer service representative will contact you with a Return Goods Authorization (RGA) Number and will provide instructions for the return of products to Teleflex Medical. Action list number 2 Distributors 1. Provide this field safety notice to all customers who have received product in scope of this FSCA. Your customer is then required to complete the Acknowledgement Form and return to you. 2. We request that you check your inventory for product within the scope of this FSCA. Cease use and distribution of impacted product and quarantine immediately. You may then return all product in scope to Teleflex. 3. As a distributor, you are then required to confirm to Teleflex that you have completed the field activity outlined above. Upon completion of your actions, please forward the completed Acknowledgement Form to Customer Service. Important - Please ensure you only list batch numbers in scope of this Field Safety Notice when completing this form. 4. If you have further distr
Recall details
- Recall number
- Z-0629-2022
- Status
- Terminated
- Initiated by the company
- Jan 11, 2022
- Posted by the FDA
- Feb 15, 2022
- Recalling firm
- Teleflex Medical Inc
- Quantity
- 14583 units
- Product code
- QEW
- FDA submissions
- K011056
- Distribution
- Worldwide distribution: US Nationwide and the countries of AU, BE, CA, CL, CZ, DE, ES, FR, GB, IL, IT, KR, KW, NL, PA, PH, PT, SA, TR.
Product description
Arrow-Trerotola PTD Set (Percutaneous Thrombolytic Device), REF PT-65509 (finished good kit) The Arrow-Trerotola Percutaneous Thrombolytic Device (PTD) Catheter, in conjunction with the Arrow Rotator Drive Unit, permits mechanical declotting of native arterio-venous (AV) fistulae and synthetic dialysis grafts.
Code information (models, lots, serial numbers)
Show the full code information (4k characters)
Lot Number / UDI: 13F20A0286 / (01)00801902013156(17)211231(10)13F20A0286; Lot Number / UDI: 13F20A0426 / (01)00801902013156(17)211231(10)13F20A0426; Lot Number / UDI: 13F20A0640 / (01)00801902013156(17)220131(10)13F20A0640; Lot Number / UDI: 13F20B0141 / (01)00801902013156(17)220131(10)13F20B0141; Lot Number / UDI: 13F20B0276 / (01)00801902013156(17)220228(10)13F20B0276; Lot Number / UDI: 13F20B0353 / (01)00801902013156(17)220228(10)13F20B0353; Lot Number / UDI: 13F20B0424 / (01)00801902013156(17)220228(10)13F20B0424; Lot Number / UDI: 13F20C0352 / (01)00801902013156(17)220228(10)13F20C0352; Lot Number / UDI: 13F20C0425 / (01)00801902013156(17)220331(10)13F20C0425; Lot Number / UDI: 13F20C0426 / (01)00801902013156(17)220331(10)13F20C0426; Lot Number / UDI: 13F20D0122 / (01)00801902013156(17)220331(10)13F20D0122; Lot Number / UDI: 13F20D0123 / (01)00801902013156(17)220430(10)13F20D0123; Lot Number / UDI: 13F20D0124 / (01)00801902013156(17)220331(10)13F20D0124; Lot Number / UDI: 13F20D0403 / (01)00801902013156(17)220430(10)13F20D0403; Lot Number / UDI: 13F20E0204 / (01)00801902013156(17)220430(10)13F20E0204; Lot Number / UDI: 13F20F0232 / (01)00801902013156(17)220531(10)13F20F0232; Lot Number / UDI: 13F20F0389 / (01)00801902013156(17)220630(10)13F20F0389; Lot Number / UDI: 13F20G0353 / (01)00801902013156(17)220630(10)13F20G0353; Lot Number / UDI: 13F20H0729 / (01)00801902013156(17)220831(10)13F20H0729; Lot Number / UDI: 13F20J0378 / (01)00801902013156(17)220831(10)13F20J0378; Lot Number / UDI: 13F20J0546 / (01)00801902013156(17)220831(10)13F20J0546; Lot Number / UDI: 13F20J0771 / (01)00801902013156(17)220831(10)13F20J0771; Lot Number / UDI: 13F20K0503 / (01)00801902013156(17)220930(10)13F20K0503; Lot Number / UDI: 13F20K0504 / (01)00801902013156(17)220930(10)13F20K0504; Lot Number / UDI: 13F20K0630 / (01)00801902013156(17)220930(10)13F20K0630; Lot Number / UDI: 13F20K0631 / (01)00801902013156(17)220930(10)13F20K0631; Lot Number / UDI: 13F20M0175 / (01)00801902013156(17)221130(10)13F20M0175; Lot Number / UDI: 13F20M0176 / (01)00801902013156(17)221130(10)13F20M0176; Lot Number / UDI: 13F20M0177 / (01)00801902013156(17)221130(10)13F20M0177; Lot Number / UDI: 13F20M0178 / (01)10801902013153(17)221130(10)13F20M0178; Lot Number / UDI: 13F20M0181 / (01)00801902013156(17)221231(10)13F20M0181; Lot Number / UDI: 13F21A0352 / (01)00801902013156(17)221231(10)13F21A0352; Lot Number / UDI: 13F21B0159 / (01)00801902013156(17)230131(10)13F21B0159; Lot Number / UDI: 13F21B0222 / (01)00801902013156(17)230131(10)13F21B0222; Lot Number / UDI: 13F21B0519 / (01)10801902013153(17)230228(10)13F21B0519; Lot Number / UDI: 13F21C0364 / (01)00801902013156(17)230228(10)13F21C0364; Lot Number / UDI: 13F21C0365 / (01)00801902013156(17)230228(10)13F21C0365; Lot Number / UDI: 13F21D0462 / (01)00801902013156(17)230331(10)13F21D0462; Lot Number / UDI: 13F21E0079 / (01)00801902013156(17)230430(10)13F21E0079; Lot Number / UDI: 13F21E0259 / (01)00801902013156(17)230430(10)13F21E0259; Lot Number / UDI: 13F21E0260 / (01)00801902013156(17)230430(10)13F21E0260; Lot Number / UDI: 13F21E0556 / (01)00801902013156(17)230430(10)13F21E0556; Lot Number / UDI: 13F21F0641 / (01)00801902013156(17)230531(10)13F21F0641; Lot Number / UDI: 13F21G0227 / (01)00801902013156(17)230630(10)13F21G0227; Lot Number / UDI: 13F21G0401 / (01)00801902013156(17)230630(10)13F21G0401; Lot Number / UDI: 13F21G0772 / (01)00801902013156(17)230630(10)13F21G0772; Lot Number / UDI: 13F21H0447 / (01)10801902194074(17)230731(10)13F21H0447; Lot Number / UDI: 13F21K0827 / (01)10801902194074(17)231031(10)13F21K0827; Lot Number / UDI: 13F21K0898 / (01)10801902194074(17)231031(10)13F21K0898; Lot Number / UDI: 13F21L0227 / (01)20801902194071(17)231031(10)13F21L0227; Lot Number / UDI: 13F21L0507 / (01)10801902194074(17)231031(10)13F21L0507; Lot Number / UDI: 13F21L0898 / (01)10801902194074(17)231031(10)13F21L0898
Devices named in this recall
2 devices in the FDA device registry whose identifier appears in the recall's code information.
- ARROW IPN035902Teleflex Incorporated · CATHETER, EMBOLECTOMY1 recall12 reports
- ARROW IPN920364Teleflex Incorporated · Catheter, embolectomy1 recall0 reports
Product lines
Other devices under the same FDA submission
9 devices cleared under K011056 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- ARROW IPN036253Teleflex Incorporated · CATHETER, EMBOLECTOMY1 recall1 report
- ARROW IPN036254Teleflex Incorporated · CATHETER, EMBOLECTOMY1 recall1 report
- ARROW IPN036249Teleflex Incorporated · CATHETER, EMBOLECTOMY1 recall0 reports
- ARROW IPN036250Teleflex Incorporated · CATHETER, EMBOLECTOMY1 recall0 reports
- ARROW IPN036252Teleflex Incorporated · CATHETER, EMBOLECTOMY1 recall0 reports
- ARROW IPN036255Teleflex Incorporated · CATHETER, EMBOLECTOMY1 recall0 reports
- ARROW IPN920363Teleflex Incorporated · Catheter, embolectomy1 recall0 reports
- ARROW IPN920365Teleflex Incorporated · Catheter, embolectomy1 recall0 reports
- ARROW IPN914602Teleflex Incorporated · Catheter, embolectomy1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 89390).
- Z-0623-2022 · TerminatedArrow-Trerotola Over-The-Wire PTD Kit, Percutaneous Thrombolytic Device, REF PT-12709-WC, containing components: PT-12709-CA, PT-12709-C, The Arrow-Trerotola Percutaneous Thro…
- Z-0624-2022 · TerminatedArrow-Trerotola Over-The-Wire PTD Kit, Percutaneous Thrombolytic Device, REF PT-65709-HFWC containing components: PT-65709-C, PT-03009, PT-65709-HFS The Arrow-Trerotola Percuta…
- Z-0625-2022 · TerminatedArrow-Trerotola Over-The-Wire PTD Kit, Percutaneous Thrombolytic Device, REF PT-65709-W containing component: PT-65709-C The Arrow-Trerotola Percutaneous Thrombolytic Device (PT…
- Z-0626-2022 · TerminatedArrow-Trerotola Over-The-Wire PTD Kit, Percutaneous Thrombolytic Device, REF PT-65709-WC containing components: PT-65709-C, PT-03009, PT-65709-HS The Arrow-Trerotola Percutane…
- Z-0627-2022 · TerminatedArrow-Trerotola PTD Set (Percutaneous Thrombolytic Device), PT-45509 containing components: PT-45509-S6, PT-45509-T The Arrow-Trerotola Percutaneous Thrombolytic Device (PTD) C…
- Z-0628-2022 · Terminated5 Fr. X 65 cm Arrow-Trerotola PTD Kit, Percutaneous Thrombolytic Device, REF PT-65509-HFC containing components: PT-03000, PT-65509-S (PT-65509-S2 alternative), PT-01000-D (PT-010…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page