NobelReplace recallZ-0658-2022 · Nobel Biocare Usa Llc · initiated Dec 17, 2021

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Nobel Biocare Usa Llc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Due to 16 complaints related to drivers not engaging in implants. According to the firm's investigations, the incorrect offset gauge was potentially used during production of the implants. As a result, this could potentially lead to aborted or a delay in surgical procedure.

FDA root cause category: Under Investigation by firm

What the company is doing

On 01/12/2022, the firm sent an "URGENT MEDICAL DEVICE RECALL" Letter via UPS to customers informing them that specific batches/lots of devices, the dimensions of the hexagon connection part are potentially not fully conforming to specification and may not fully engage in the implant. Customers are instructed to: 1. Inspect their stock and quarantine all affected devices. 2. Complete the Customer Acknowledgement Form and return to the Recalling Firm via email or fax within 5 days of receipt of this notice. 3. Return all affected stock on hand to the Recalling Firm using the shipping label provided. 4. Ensure relevant staff members are informed of this recall, and if you further supplied or transferred any of the affected products to another facility or organization, let these facilities know of the recall and provide them with a copy of this notice. For questions, contact your local customer support representative,

Recall details

Recall number
Z-0658-2022
Status
Terminated
Initiated by the company
Dec 17, 2021
Recalling firm
Nobel Biocare Usa Llc
Quantity
702 pieces
Product code
DZE
FDA submissions
K073142
Distribution
Worldwide distribution - U.S. Nationwide distribution in the states of AL, AR, AZ, CA, CO, FL, GA, HI, ID, IL, IN, KS, KY, MA, ME, MI, MN, MS, NC, NJ, NM, NV, NY, OH, OR, PA, SC, TN, TX, UT, VA, VT, WA and WI. The countries of Canada, Mexico and Turkey.

Product description

NobelReplace Conical Connection RP 4.3x11.5 mm REF 36707 STERILE R

Code information (models, lots, serial numbers)

Catalog Number: 36707 UDI Code: (01)07332747160529(10)13116495(11)210720(17)260618; (01)07332747160529(10)13116496(11)210720(17)260618; (01)07332747160529(10)13116497(11)210720(17)260618; Lot Numbers: 13116495, 13116496, 13116497