Skip to content

Regard, TOTAL HIP - SPRINGFIELD, Item No. 8800297010, Case… recallZ-0693-2022 · ROi CPS LLC · initiated Jan 13, 2022

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from ROi CPS LLC or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Products are mislabeled with the incorrect product identification on individual kit labeling.

FDA root cause category: Under Investigation by firm

What the company is doing

The firm emailed a recall notification to customers on 01/13/2022 tited "ROi CPS, LLC Recall-Custom Procedure Kit Recall-Kits Containing the Incorrect Product Label." The recall notice states that the reason for recall is due to the individual kits being incorrectly labeled, while the outer box is labeled correctly. The outer box was labeled with Product No. 880297010, while the inner kit bags were labeled as Product No. 880261009. Customers are directed to review their inventory to determine if recalled kits are in stock and quarantine any identified recalled product. The firm instructs customers to open the outer carton to examine the packaging of individual bags for incorrect labeling; do not open sterile barrier packaging. Cases containing the correct inner packaging can be released. Customers are asked to complete the recall response form attached to the recall notification indicating any affected product on hand. Complete forms are to be emailed back to [email protected] so product return can be arranged.

Recall details

Recall number
Z-0693-2022
Status
Terminated
Initiated by the company
Jan 13, 2022
Recalling firm
ROi CPS LLC
Quantity
50 kits
Product code
OJH
Distribution
Missouri

Product description

Regard, TOTAL HIP - SPRINGFIELD, Item No. 8800297010, Case Quantity: 2, Sterile

Code information (models, lots, serial numbers)

Item No. 8800297010; Lot No. 90328; UDI 10194717113164

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page