KWIK-STIK recallZ-0641-2022 · Microbiologics Inc · initiated Mar 25, 2013

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Microbiologics Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Some of the KWIK-STIK Catalog #01023P Campylobacter coli, Lot# 1023-05-7, contained an inner label stating the product is Catalog #0919K Haemophilus influenza. The organism has been confirmed to be C. coli, indicating this is a labeling error and the organism is the correct organism.

FDA root cause category: Labeling mix-ups

What the company is doing

Microbiologics notified customers on 03/25/2013 via letter titled "Important Medical Device Correction". Instructions included to inspect inventory for affected products. If customers have received the product identified within the recall (KWIK-STIK" 01023P, Lot# 1023-05-7), it is recommended that the customer destroys the product. Microbiologics shall replace the destroyed product upon confirmation of destruction. Questions or concerns can be directed to the Customer Service Manager at 320.229.7057.

Recall details

Recall number
Z-0641-2022
Status
Terminated
Initiated by the company
Mar 25, 2013
Recalling firm
Microbiologics Inc
Quantity
10 units
Product code
JTR
FDA submissions
K861022
Distribution
Worldwide distribution - US Nationwide and the countries of Canada, South Africa, France, Netherlands, Mauritius, China.

Product description

KWIK-STIK(TM), Quality control kit for culture media, sold as: KWIK-STIK(TM) 2-pack Enterococcus faecalis derived from ATCC(R) 51299(TM)

Code information (models, lots, serial numbers)

Catalog Number 01023P, UDI: 20845357029502, Lot Number (Expiration Date): 1023-05-7 (04/30/2014)