Philips Respironics recallZ-0662-2022 · Respironics California, LLC · initiated Jan 24, 2022
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Respironics California, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
A limited number of ventilators were assembled with expired adhesive. If the adhesive fails, a bracket may become loose and potentially damage the capacitors, which could cause the ventilator to stop providing ventilation to the patient. This failure may either activate both visual and audible alarms, or it may not sound or display an alarm (a silent shutdown).
FDA root cause category: Process control
What the company is doing
On 01/24/2022, the firm sent an "URGENT MEDICAL DEVICE CORRECTION" Letter via priority mail to customer informing them that a limited number of ventilators have been identified to have an internal component assembled with expired adhesive. There is a possibility that two failures may occur simultaneously activated by adhesive failure and if the component that the adhesive is assembled to also fails. The results of the failures could lead to: 1) The ventilator ceases to operate, activating both visual and audible alarms, or 2) The ventilator ceases to operate and does not activate either visual or audible alarm causing a "silent shutdown". Customers are instructed to: 1) Compare the serial numbers of their device(s) to Attachment A: List of Impacted Serial Numbers to determine if the ventilator is impacted. Device serial number information can be located at the rear of the ventilator. Alternatively, the serial number of the ventilator may be viewed from the display while the ventilators is operation. Select the Menu tab at the bottom of the screen then select Vent Info. 2) Additional actions that should be taken by the customer: -It is not necessary to remove affected ventilators from service. -The ventilators have a remote alarm system capability that allows the ventilators to be connected to a remote alarm system. The Recall Firm is recommending to utilize a remote alarm. The remote alarm will provide a backup warning even if the ventilator's primary alarm system does not alarm. Directions for connecting a remote alarm system can be found in Section B-5: "Remote Alarm Port" of the Operation's Manual. -It is important to follow directions in the Operation's Manual and the Urgent Medical Device Correction Letter to further reduce any risk associated with this potential failure. -From the Operator's Manual: 1. Provide external oxygen monitoring to minimize patient risk in case of O2 supply loss or ventilator failure. 2. Promptly respond to all low
Recall details
- Recall number
- Z-0662-2022
- Status
- Open, Classified
- Initiated by the company
- Jan 24, 2022
- Posted by the FDA
- Feb 25, 2022
- Recalling firm
- Respironics California, LLC
- Quantity
- 16 systems
- Product code
- MNT
- FDA submissions
- K102985
- Distribution
- U.S.: AR, CA, FL, GA, IN, LA, MD, MO, NY, PA, SC, TN and TX. O.U.S. N/A
Product description
Philips Respironics V60 Ventilator Part Number 1053617
Code information (models, lots, serial numbers)
Show the full code information (1k characters)
The Component of the Ventilators that is impacted is: N 1034132 BRACKET, CAPACITOR,MOTOR DRIVER PCB Model V60 Part Number: 1053617 Serial Numbers/UDI Code: MX0016593 / (01)00884838020054(21)MX0016593 MX0016797 / (01)00884838020054(21)MX0016797 MX0016824 / (01)00884838020054(21)MX0016824 MX0016827 / (01)00884838020054(21)MX0016827 MX0016830 / (01)00884838020054(21)MX0016830 MX0016839 / (01)00884838020054(21)MX0016839 MX0016840 / (01)00884838020054(21)MX0016840 MX0016841 / (01)00884838020054(21)MX0016841 MX0016845 (01)00884838020054(21)MX0016845 MX0016853 (01)00884838020054(21)MX0016853 MX0016867 (01)00884838020054(21)MX0016867 MX0016870 (01)00884838020054(21)MX0016870 MX0016871 (01)00884838020054(21)MX0016871 MX0016872 (01)00884838020054(21)MX0016872 MX0016875 (01)00884838020054(21)MX0016875 MX0016878 (01)00884838020054(21)MX0016878
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Philips Respironics V60Respironics California, Inc. · Ventilator, Continuous, Minimal Ventilatory Support,Facility Use3 recalls30 deaths13,525 reports
Product lines
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page