QC Sets and Panels: Helix Elite recallZ-0697-2022 · Microbiologics Inc · initiated Jun 6, 2017

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Microbiologics Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Real-time shelf life testing failed at 24 months

FDA root cause category: Process design

What the company is doing

Microbiologics notified customers on about 06/06/2017 via email containing a letter. The letters indicated the affected products and instructed customers to update their records with the corrected expiration date, which was provided for each affected lot.

Recall details

Recall number
Z-0697-2022
Status
Terminated
Initiated by the company
Jun 6, 2017
Recalling firm
Microbiologics Inc
Quantity
8 units
Product code
OHQ
Distribution
Worldwide distribution - US Nationwide distribution in the states of AK, CA, DC, FL, IA, ID, KS, LA, MD, MN, MO, NE, NH, NJ, NY, OH, OK, PA, SC, TX, VA, WA, WV and the countries of Australia, Belgium, Canada, India, United Kingdom.

Product description

Microbiologics QC Sets and Panels: Helix Elite. Extended Enteric Bacterial Verification Panel (Inactivated Pellet) includes 21 pellets each comprised of the following 8 organisms: 1. S. Typhimurium derived from NCTC 74, 2. S. sonnei derived from NCTC 12984, 3. E. coli derived from CDC 2010C-3114 (STEC), 4. C. jejuni derived from NCTC 11322, 5. P. shigelloides derived from NCTC 10363, 6. Y. enterocolitica derived from NCTC 11174, 7. E. coli derived from NCTC 11602 (ETEC), 8. V. parahaemolyticus derived from NCTC 13209

Code information (models, lots, serial numbers)

Catalog Number: 8191, Lot Number: 8191-01, UDI: 70845357041240

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other products in the same recall event

The FDA groups related products into one recall event (event 89446).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page