EVEREST Spinal System recallZ-0706-2022 · K2M, Inc · initiated Jan 13, 2022
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from K2M, Inc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Stryker received two (2) complaints for units from specific lots of Inner and Outer Dilators not fitting together properly during a surgical procedure. Some dilators from these lots and from other lots were subsequently determined to be affected by a manufacturing nonconformance.
FDA root cause category: Process control
What the company is doing
URGENT: MEDICAL DEVICE RECALL notification letters dated 1/13/22 were sent to customers. Hospital Quality or Materials Managers/Inventory Contacts instructions: Actions needed 1. Immediately check your internal inventory to locate the product listed on the attached Business Reply Form and remove them from their point of use. 2. Use the Business Reply Form to reconcile any affected product. Complete the Business Reply Form even if there is no affected product identified. 3. Return the enclosed Business Reply Form by email to [email protected] or fax to 201- 575-4675 to confirm receipt of this notification/document product segregation. Your account s Stryker Spine Sales Representative will arrange for retrieval and removal of any instruments associated with this recall. 4. In the interim, until units can be removed from your facility by Stryker, there are no additional actions that users can and should take once the product has been segregated and removed from point of use. - If you desire additional training associated with these instructions, please contact our Regulatory Compliance team by calling 201.749.8090 or by email at [email protected], or contact your local Sales representative. 5. Maintain awareness of this communication internally until all required actions have been completed within your facility. 6. Inform Stryker if any of the subject devices have been distributed to other organizations. If so, provide contact details so Stryker can inform the recipients appropriately. 7. Under 21 CFR 803, manufacturers are required to report any serious injuries where a product has contributed or may have contributed to the event. Please keep Stryker informed of any adverse events associated with this product by using Stryker s online reporting site: https://www.stryker.com/productexperience. Additionally, adverse reactions or quality problems experienced with the use of this product may also be reported to the FDA's MedWatch Adver
Recall details
- Recall number
- Z-0706-2022
- Status
- Open, Classified
- Initiated by the company
- Jan 13, 2022
- Recalling firm
- K2M, Inc
- Quantity
- 686 Total Devices
- Product code
- KWP
- Distribution
- Worldwide distribution - US Nationwide.
Product description
Everest MI XT Outer Dilator, Catalog Number 5101-90168
Code information (models, lots, serial numbers)
GTIN: 10888857261211; Lot Numbers: HCBN HCBP JUJJ KFMX KUPH KYYV MDPF NAKJ NDGT PCCN
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- EVEREST Spinal System 5101-90168VB Spine LLC · Orthopedic manual surgical instrument1 recall0 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 89522).
- Z-0705-2022 · Open, ClassifiedEverest MI XT Inner Dilator, Catalog Number 5101-90167
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page