Skip to content

Comprehensive Shoulder System recallZ-0672-2022 · Biomet, Inc. · initiated Jan 7, 2022

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Biomet, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The products in scope underwent an incorrect rework operation for the porous plasma sprayed coating. Internal testing indicated that the parts may not have sufficient adhesion strength. The scope was identified to be all lots that were porous plasma coated or recoated at a vendor that were then reworked through the internal waterjet rework process. The waterjet rework process for porous plasma coating is not validated for products that were coated at a vendor.

FDA root cause category: Process control

What the company is doing

Zimmer Biomet issued Urgent Medical Device Recall Letter (ZFA 2021-00267) to Distributors, Risk Managers, and Surgeons on January 7, 2022 via FedEx. For US and OUS distributors, the notification was completed via email . Letter states reason for recall, health risk and action to take: Risk Manager Responsibilities: 1. Review this notification and ensure that affected personnel are aware of the contents. 2. If you have affected product at your facility, assist your Zimmer Biomet sales representative and quarantine all affected product. Your Zimmer Biomet sales representative will remove the affected product from your facility. 3. If the product has been further distributed, provide your customers with the field safety notice for hospitals and ensure documentation. 4. Complete Attachment 1 Certificate of Acknowledgement and send to [email protected]. This form shall be returned even if you do not have affected products at your facility. 5. Retain a copy of the acknowledgement form with your recall records in the event of a compliance audit of your facility s documentation. 6. If you have further questions or concerns after reviewing this notice, please call customer service at 574-371-3071 between 8:00 am and 5:00pm EST, Monday through Friday. Calls received outside of call center operating hours will receive a voicemail prompt or be transferred to an on-call representative in the event of an emergency. Alternatively, your questions may be emailed to [email protected]. Surgeon Responsibilities: 1. Review this notification for awareness of the contents. 2. If the product has been implanted, Zimmer Biomet recommends assessing the patient with imaging and monitoring the patient for third body debris or loosening that could potentially contribute to the potential health risks. 3. Complete Attachment 1 Certificate of Acknowledgement and send to [email protected]. 4. Retain a copy of the

Recall details

Recall number
Z-0672-2022
Status
Terminated
Initiated by the company
Jan 7, 2022
Recalling firm
Biomet, Inc.
Quantity
12 units
Product code
MBF
Distribution
Worldwide distribution - US Nationwide and the countries of Canada, Ecuador, Netherlands.

Product description

Comprehensive Shoulder System Primary Shoulder Stem, Micro Length, Porous Plasma, 14 MM, 55 MM Long Item Number: 113614

Code information (models, lots, serial numbers)

Lot Number: 439470 UDI: (01)00880304554153(17) 290803(10)439470

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other products in the same recall event

The FDA groups related products into one recall event (event 89544).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page