Lumbar Puncture Tray Adult 20g X 3.5 Dp recallZ-0792-2022 · Bard Peripheral Vascular Inc · initiated Jan 20, 2022

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Bard Peripheral Vascular Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Due to lumbar puncture tray manometer failure mode of leaking and mating issues.

FDA root cause category: Process control

What the company is doing

Between 01/28 and 02/01/2022, the firm sent an "URGENT MEDICAL DEVICE CORRECTION" Letter via email and FedEx 2-day mailings to customers informing them that between 07/02 to 12/15/2021 the firm received 59 customer complaints for lumbar puncture tray manometers with reported failure modes of leaking and mating issues. Customers are asked to take the following actions: 1. Check all inventory locations within their institution for the product listed in Table A and immediately destroy all affected product remaining in their possession. The product should have the sterile barrier broken and destroyed per their local facility s destruction process. 2. Share this notification with all users within their facilities to ensure they are also aware of this Urgent Medical Device Correction. 3. If Customers purchased this product from a distributor, contact their distributor for further instructions and credit resolution. 4. Complete and return the attached Customer Response Form to the Recalling Firm contact noted on the form confirming acknowledgement of the correction notification, whether or not they have any affected product, so that the Recalling Firm may acknowledge their receipt of this notification and process their credit. 5. Indicate on the response form the quantity of affected product identified at their facility and confirm that this inventory was destroyed. 6. Contact their Recalling Firm's representative to assist in this process or they may also contact the listed in the notification for further assistance. For further assistance contact: North American Regional Complaint Center - 1-844-8BD- LIFE (1-844-823-5433), Say Recall when prompted M-F 8am - 5pm CT

Recall details

Recall number
Z-0792-2022
Status
Open, Classified
Initiated by the company
Jan 20, 2022
Recalling firm
Bard Peripheral Vascular Inc
Quantity
116,770 devices
Product code
PWY
Distribution
U.S. Nationwide Distribution: AK, AL, AR, AZ, CA, CO, CT, DC, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, Puerto Rico, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY. O.U.S. (foreign) including states: Costa Rica and Hong Kong.

Product description

LUMBAR PUNCTURE TRAY ADULT 20G X 3.5 DP

Code information (models, lots, serial numbers)

Show the full code information (2k characters)

Catalog Number: 4301C Lot Numbers/UDI (GTIN, DI + PI): 0001410711 (01)10885403057038(17)220331(10)0001410711 0001411299 (01)10885403057038(17)220331(10)0001411299 0001411867 (01)10885403057038(17)220331(10)0001411867 0001414258 (01)10885403057038(17)220331(10)0001414258 0001414358 (01)10885403057038(17)220331(10)0001414358 0001414359 (01)10885403057038(17)220331(10)0001414359 0001415031 (01)10885403057038(17)220430(10)0001415031 0001415038 (01)10885403057038(17)220430(10)0001415038 0001415850 (01)10885403057038(17)220430(10)0001415850 0001415809 (01)10885403057038(17)220430(10)0001415809 0001417243 (01)10885403057038(17)220430(10)0001417243 0001417244 (01)10885403057038(17)220430(10)0001417244 0001417734 (01)10885403057038(17)220430(10)0001417734 0001418034 (01)10885403057038(17)220430(10)0001418034 0001418308 (01)10885403057038(17)220430(10)0001418308 0001418410 (01)10885403057038(17)220430(10)0001418410 0001421652 (01)10885403057038(17)220630(10)0001421652 0001421921 (01)10885403057038(17)220630(10)0001421921 0001422197 (01)10885403057038(17)220630(10)0001422197 0001422150 (01)10885403057038(17)220630(10)0001422150 0001422604 (01)10885403057038(17)220630(10)0001422604 0001425527 (01)10885403057038(17)220630(10)0001425527 0001425874 (01)10885403057038(17)220630(10)0001425874 0001426189 (01)10885403057038(17)220630(10)0001426189 0001426502 (01)10885403057038(17)220630(10)0001426502 0001427225 (01)10885403057038(17)220630(10)0001427225 0001427380 (01)10885403057038(17)220630(10)0001427380 0001429130 (01)10885403057038(17)220630(10)0001429130 0001429519 (01)10885403057038(17)220630(10)0001429519 0001430121 (01)10885403057038(17)220630(10)0001430121 0001430608 (01)10885403057038(17)220630(10)0001430608 0001431407 (01)10885403057038(17)220630(10)0001431407 0001431472 (01)10885403057038(17)220630(10)0001431472 0001432176 (01)10885403057038(17)220630(10)0001432176 0001432655 (01)10885403057038(17)220831(10)0001432655 0001433601 (01)10885403057038(17)220930(10)0001433601 0001433023 (01)10885403057038(17)220930(10)0001433023 0001434251 (01)10885403057038(17)220930(10)0001434251 0001434730 (01)10885403057038(17)220930(10)0001434730 0001436141 (01)10885403057038(17)220930(10)0001436141 0001437469 (01)10885403057038(17)221031(10)0001437469 0001437660 (01)10885403057038(17)221031(10)0001437660 0001437658 (01)10885403057038(17)221031(10)0001437658 0001437768 (01)10885403057038(17)221031(10)0001437768

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Other products in the same recall event

The FDA groups related products into one recall event (event 89553).