Skip to content

Liko Universal Twinbar 670 QRH, an accessory that can be… recallZ-0829-2022 · Baxter Healthcare Corporation · initiated Feb 21, 2022

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Baxter Healthcare Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

An incorrect center bolt was used in the assembly of the product which could result in a patient fall during use.

FDA root cause category: Nonconforming Material/Component

What the company is doing

The recalling firm issued letters to U.S. consignees on 2/21/2022 via UPS ground. The letter provided a description of the problem with the background which included photographs and views of the issue. The potential risk was explained and actions to be taken by the customer were provided, which included to identify whether they have an impacted product and to return the enclosed response form within two weeks. Upon receipt of the response form, Hillrom will contact the consignee to arrange to inspect the Twinbar and if the short bolts have been used, the Twinbar will be corrected. The consignee was instructed not to use the Twinbar together with a Liko Scale until an inspection has been completed. Distributors were instructed to share the Medical Device Correction with their end users and complete the attached response form for return within two weeks. Hillrom will provide replacements for distributors to facilitate exchange among end users. A 3rd party recall service will be issuing notifications to the OUS consignees.

Recall details

Recall number
Z-0829-2022
Status
Open, Classified
Initiated by the company
Feb 21, 2022
Recalling firm
Baxter Healthcare Corporation
Quantity
1,399 twinbars
Product code
FSA
Distribution
Worldwide distribution - US Nationwide distribution in the states of AK, AZ, CA, CO, DC, FL, GA, HI, IA, ID, IN, KS, KY, LA, MA, MD, MI, MN, MO, ND, NE, NJ, NV, NY, OH, OK, OR, PA, TN, TX, UT, VA, WA, and WI. There was government distribution but no military distribution. The countries of Austria, Australia, Brazil, Canada, Chile, China, Colombia, Denmark, Finland, France, Germany, Hong Kong, Ireland, Japan, Jordan, Latvia, Netherlands, Portugal, South Africa, Spain, Sweden, Switzerland, Taiwan, Tunisia, United Kingdom, and Uruguay.

Product description

Liko Universal Twinbar 670 QRH, an accessory that can be used together with most Liko lifts, model numbers 3156087 and P3156087.

Code information (models, lots, serial numbers)

Manufactured before 28OCT2021, UDI 00887761982468

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page