Skip to content

Del(5q) Deletion FISH Probe Kit: In-Vitro Diagnostic to… recallZ-0741-2022 · Cytocell Ltd. · initiated Jan 18, 2022

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Cytocell Ltd. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

individual components have been labelled with incorrect colours. The red and green colours are opposite to those specified in the product labelling. The EGR1 probe is labelled in green rather than red and the control probe labelled in red rather than green. The expected positive signal pattern in the case of 1x deleted 5q will be 1G2R instead of 1R2G, an incorrect result

FDA root cause category: Software Manufacturing/Software Deployment

What the company is doing

Cytocell issued Urgent Medical Recall Reference: VC/2022/001 letter on 1/18/22 to Distributors and End Users via email. Letter states reason for recall, health risk and action to take: Immediately examine your inventory and quarantine all product subject to recall. Cytocell requests that you destroy the remaining inventory. We also suggest that laboratories undertake a review of any results obtained with the affected devices and check that signal patterns were not misinterpreted as a result of the probe being manufactured with the fluorophores the incorrect way around. Please complete and return the enclosed response form as soon as possible. If you have any questions, contact Cytocell on Tel: +44 (0)1223 294048, Fax +44(0)1223 294986 or [email protected].

Recall details

Recall number
Z-0741-2022
Status
Terminated
Initiated by the company
Jan 18, 2022
Recalling firm
Cytocell Ltd.
Quantity
19 kits
Product code
QDI
Distribution
US Nationwide distribution in the states of Hawaii, Missouri, Pennsylvania, Tennessee.

Product description

Del(5q) Deletion FISH Probe Kit: In-Vitro Diagnostic to detect deletions in chromosome location from patients with AML or MDS Catalog Number: USA-LPH024

Code information (models, lots, serial numbers)

Lot Number: 075711 Exp. Date: 2023-06-22 UDI: (01)05055844901551(17)230622(10)075 711

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page