Titanium MPJ (Toe Implant) TOE JOINT TITANIUM NPC MD, Part… recallZ-0683-2022 · BioPro, Inc. · initiated Feb 1, 2022
Terminated
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from BioPro, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
There is an incorrect statement regarding MR Compatibility in the BioPro Titanium MPJ implant IFU (IFU BP102303 rev 05) and patient implant cards (22266 rev 00) which indicates that the BioPro MPJ Implant has been tested for safety and compatibility in the MR Environment and is MR Conditional. MR Testing has not been conducted on the titanium version of the MPJ.
FDA root cause category: Labeling design
What the company is doing
An URGENT: MEDICAL DEVICE RECALL notification letter was distributed to customers on 2/1/22. Actions to be taken by the Customer BioPro has fully reviewed all records pertaining to the product and implant cards. To date, no implant cards have been issued for titanium MPJ implants. However, your facility has implanted one MPJ Implant whose IFU improperly contained the MR Conditional Statement. The information pertaining to that order is below. Please take the following actions 1. If the incorrect IFU was retained, replace it with the updated IFU, which has been included for your reference. Please dispose of the incorrect IFU. 2. Provide the enclosed corrected patient implant card to your patient who received the implant. 3. Please sign and return the following page acknowledging receipt of this notification. Actions Taken by the Recalling Firm All impacted labeling has been updated by BioPro. We apologize for any inconvenience or issue this may cause. Please feel free to contact BioPro directly if you have any questions. Hours of operation are Monday Friday from 8:00-5:00 PM est. If you are aware of any issues that may have occurred related to this implant, please notify BioPro, Inc immediately. You may additionally contact the FDA directly via the MedWatch Adverse Event Reporting program. Submit reports to the FDA through the MedWatch program in one of the following ways: " Complete the MedWatch Online Reporting Form (https://www.accessdata.fda.gov/scripts/medwatch/) " Download form or call 1-800-332-1088 to request a reporting form, then complete and return to the address on the pre-addressed form, or submit by fax to 1-800-FDA-0178
Recall details
- Recall number
- Z-0683-2022
- Status
- Terminated
- Initiated by the company
- Feb 1, 2022
- Recalling firm
- BioPro, Inc.
- Product code
- KWD
- FDA submissions
- K041595
- Distribution
- US Nationwide distribution in the states of CT, KY, NM, PA, TN, and SD.
Product description
Titanium MPJ (Toe Implant) TOE JOINT TITANIUM NPC MD, Part ID 16814
Code information (models, lots, serial numbers)
Lot Number: 125423, 128356; UDI: 00810012480093
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- TOE JOINT TITANIUM NPC MD 20MM 16814Biopro, Inc. · Prosthesis, Toe, Hemi-, Phalangeal1 recall0 reports
Other devices under the same FDA submission
6 devices cleared under K041595 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- TOE JOINT TITANIUM NPC SM 17MM 16813Biopro, Inc. · Prosthesis, Toe, Hemi-, Phalangeal1 recall0 reports
- TOE JOINT TITANIUM NPC M/L 21.5MM 16815Biopro, Inc. · Prosthesis, Toe, Hemi-, Phalangeal1 recall0 reports
- TOE JOINT TITANIUM NPC LG 23MM 16816Biopro, Inc. · Prosthesis, Toe, Hemi-, Phalangeal1 recall0 reports
- TOE JOINT TITANIUM PC SM 17MM 17035Biopro, Inc. · Prosthesis, Toe, Hemi-, Phalangeal1 recall0 reports
- TOE JOINT TITANIUM PC MD 20MM 17036Biopro, Inc. · Prosthesis, Toe, Hemi-, Phalangeal1 recall0 reports
- TOE JOINT TITANIUM PC M/S 18.5MM 17197Biopro, Inc. · Prosthesis, Toe, Hemi-, Phalangeal1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 89605).
- Z-0682-2022 · TerminatedTitanium MPJ (Toe Implant) TOE JOINT TITANIUM NPC SM, Part ID 16813
- Z-0684-2022 · TerminatedTitanium MPJ (Toe Implant) TOE JOINT TITANIUM NPC ML, Part ID 16815
- Z-0685-2022 · TerminatedTitanium MPJ (Toe Implant)TOE JOINT TITANIUM NPC LG, Part ID 16816
- Z-0686-2022 · TerminatedTitanium MPJ (Toe Implant)TOE JOINT TITANIUM PC SM, Part ID 17035
- Z-0687-2022 · TerminatedTitanium MPJ (Toe Implant) TOE JOINT TITANIUM PC MD, Part ID 17036
- Z-0688-2022 · TerminatedTitanium MPJ (Toe Implant) TOE JOINT TITANIUM PC M/L, Part ID 17037
- Z-0689-2022 · TerminatedTitanium MPJ (Toe Implant) TOE JOINT TITANIUM PC LG, Part ID 17038
- Z-0690-2022 · TerminatedTitanium MPJ (Toe Implant) TOE JOINT TITANIUM PC MS, Part ID 17197
- Z-0691-2022 · TerminatedTitanium MPJ (Toe Implant) TOE JOINT TITANIUM NPC MS, Part ID 17198
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page