Skip to content

NavitrackER Kit A: Knee, Model Number 20-8000-000-07 recallZ-0721-2022 · Orthosoft, Inc. dba Zimmer CAS · initiated Jan 7, 2022

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Orthosoft, Inc. dba Zimmer CAS or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The product was released for distribution without passing sterilization results, which could potentially lead to insufficient sterility of the product. The highest severity risk is infection leading to surgical intervention.

FDA root cause category: Under Investigation by firm

What the company is doing

On January 7, 2022, the firm issued Urgent Medical Device Recall letters to all affected consignees. Customers were instructed to assist Zimmer Biomet sales representatives and quarantine all affected product. Product will be removed from customer facilities. Customers are responsible for forwarding the recall notice to any accounts who received further distribution. If you have further questions or concerns after reviewing this notice, please call customer service at 574-371-3071 between 8:00 am and 5:00pm EST, Monday through Friday. Calls received outside of call center operating hours will receive a voicemail prompt or be transferred to an on-call representative in the event of an emergency. Alternatively, your questions may be emailed to [email protected].

Recall details

Recall number
Z-0721-2022
Status
Open, Classified
Initiated by the company
Jan 7, 2022
Recalling firm
Orthosoft, Inc. dba Zimmer CAS
Quantity
14 units US; 132 units OUS
Product code
OLO
FDA submissions
K110054
Distribution
Worldwide distribution - US Nationwide distribution in the states of AR, NY, OH, TN, TX and the countries of Canada and the Netherlands.

Product description

NavitrackER Kit A: Knee, Model Number 20-8000-000-07,

Code information (models, lots, serial numbers)

Lot number 110221A1 UDI Number (01)00889024304222 (17)231103(10)110221A1

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page