cobas LiatSystem, respiratory virus panel nucleic acid… recallZ-0828-2022 · Roche Molecular Systems, Inc. · initiated Nov 16, 2021
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Roche Molecular Systems, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Roche has identified a small number of cobas Liat analyzer units that have consistently high noise levels (background signal) originating from the amber detector (used for Influenza B detection). This noise increases the probability of false positive Influenza B results.
FDA root cause category: Nonconforming Material/Component
What the company is doing
A Customer Bulletin dated 11/16/21 was sent to customers. Analyzer Return Procedure Our records indicate that you have one of the affected cobas Liat analyzers listed on the previous page. Because of the issue, Roche will send you a new cobas Liat analyzer. Please return your affected analyzer immediately to Roche for investigation by following the steps below: 1. A representative from the Roche Support Network Customer Support Center will contact you to initiate analyzer replacement and confirm shipping information. 2. When you receive the new cobas Liat analyzer from Roche, remove the new analyzer and FedEx shipping label from the box. 3. Place the affected analyzer in the box and seal it using packing tape. 4. Affix the FedEx postage paid shipping label to the outside of the box. 5. Take the box containing the affected analyzer to an appropriate FedEx location for return shipping at no charge to you. Actions Required " Follow the steps in the Analyzer Return Procedure section above. " File this Customer Bulletin for future reference. Please contact the Roche Support Network Customer Support Center at 1-800-800-5973 if you have questions about the information in this Customer Bulletin.
Recall details
- Recall number
- Z-0828-2022
- Status
- Open, Classified
- Initiated by the company
- Nov 16, 2021
- Recalling firm
- Roche Molecular Systems, Inc.
- Quantity
- 14 units
- Product code
- OOI
- FDA submissions
- K111387
- Distribution
- Worldwide distribution - US Nationwide distribution in the states of CO, IL, IN, KY, MI, NC, and TX. The countries of Cyprus, Spain, and Hong Kong.
Product description
cobas LiatSystem, respiratory virus panel nucleic acid assay system, Material Number 07341920190
Code information (models, lots, serial numbers)
UDI: 07613336100097; Serial Numbers: 16333, 16454, 16688, 16695, 16871, 16914, 17048, 17266, 19924, 18094, 18323, 18357, 17116, 17086
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- cobas Liat Analyzer 07341920190Roche Molecular Systems, Inc. · real time Nucleic acid amplification system1 recall4 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page