Revogene, For In Vitro Diagnostic use in performing nucleic… recallZ-0836-2022 · Meridian Bioscience Inc · initiated Feb 15, 2022
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Meridian Bioscience Inc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The photomultiplier tube, may have compromised functionality in some instruments. As a result, the user is likely to observe an increase in false-positive or indeterminate results.
FDA root cause category: Component design/selection
What the company is doing
A Letter dated 02/15/2022 sent via FedEx notified users of the risk of false-positives and provided instructions on how to recognize a compromised photomultiplier tube via error codes and based on various combinations of positive and indeterminate results. If a positive result with the negative control is obtained or false positive suspected, users are to contact Technical Service and cease using the instrument. The company is developing software solution aimed at identifying photomultiplier tubes that are at risk for failure and is seeking a hardware fix to this issue.
Recall details
- Recall number
- Z-0836-2022
- Status
- Open, Classified
- Initiated by the company
- Feb 15, 2022
- Recalling firm
- Meridian Bioscience Inc
- Quantity
- 452 units (398 US, 54 OUS)
- Product code
- OOI
- FDA submissions
- K170558
- Distribution
- Worldwide distribution - US Nationwide distribution in the states of AK, AL, AR, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, ME, MI, MN, MO, MS, MT, NC, NE, NJ, NV, NY, OH, OR, PA, PR, RI, SD, TN, TX, UT, VA, VT, WI, WV and the countries of Australia, Chile, Italy, Switzerland.
Product description
Revogene, For In Vitro Diagnostic use in performing nucleic acid testing in clinical laboratories. Meridian branded Revogene: Catalog Number 610210 GenePOC branded Revogene: Catalog Number 130840
Code information (models, lots, serial numbers)
UDI: 00840733102318 All units in the field, all serial numbers.
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Revogene 610210Meridian Bioscience, Inc. · Real Time Nucleic Acid Amplification System2 recalls10 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.