Skip to content

Gunther Tulip recallZ-0875-2022 · Cook Medical Incorporated · initiated Feb 14, 2022

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Cook Medical Incorporated or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

IFU update includes an update to the Device description, Intended Use/Indications for Use, Contraindications, Warnings, Precautions, Potential adverse events

FDA root cause category: Labeling design

What the company is doing

Cook issued Urgent Medical Device Correction letter on February 14, 2022 via courier to US consignees. Letter states reason for recall, health risk and action to take: 1. Please read the new IFU in entirety. 2. The electronic versions of the IFUs can be found on the Cook Medical website by Catalogue Number (RPN): https://ifu.cookmedical.com/ifuPub/searchIfu.jsf 3. Your Cook Medical Sales Representative can facilitate delivery of new IFUs for your inventory upon request. 4. Please complete and return the Acknowledgement and Receipt Form within 5 business days of receiving this letter regardless of whether you have affected product(s) on hand. You may return the Form via fax (812.339.7316) or email ([email protected]). 5. Please continue to report any adverse events to Cook Medical Customer Relations by phone at 800.457.4500 or 812.339.2235, Monday through Friday between 7:30 am and 5:00 pm (Eastern Time) or by email to [email protected]. 6. Adverse events or quality problems experienced with the use of this product may also be reported to the FDA. Visit http://www.fda.gov/Safety/MedWatch/HowToReport/default.htm to obtain a form to fax or mail or call the FDA at 800.FDA.1088. This notice must be shared with appropriate personnel, including down to the user level, within your organization or with any organization where the potentially affected devices may have been transferred. If you have any questions or concerns, please contact Cook Medical Customer Relations at 800.457.4500 or 812.339.2235

Recall details

Recall number
Z-0875-2022
Status
Terminated
Initiated by the company
Feb 14, 2022
Recalling firm
Cook Medical Incorporated
Quantity
24,641 units
Product code
DTK
FDA submissions
K211874
Distribution
US Nationwide - Worldwide Distribution Foreign: Canada, Bahamas ,Chile Guam

Product description

Gunther Tulip Vena Cava Filter Set for Femoral Vein Approach Catalog Number/GPN: IGTCFS-65-1-FEM-TULIP G52917; IGTCFS-65-1-JUG-TULIP G52916; IGTCFS-65-1-UNI-TULIP G52918

Code information (models, lots, serial numbers)

Show the full code information (86k characters)

All Lots manufactured within 24 Jan 2019 through 24 Jan 2022. Expiration Dates from January 8, 2022 through January 18 2025. RPN GPN LOT UDI IGTCFS-65-1-UNI-TULIP G52918 E3814493 1082700252918920220114E3814493 IGTCFS-65-1-UNI-TULIP G52918 E3817291 1082700252918920220123E3817291 IGTCFS-65-1-UNI-TULIP G52918 E3817898 1082700252918920220125E3817898 IGTCFS-65-1-UNI-TULIP G52918 E3818494 1082700252918920220128E3818494 IGTCFS-65-1-FEM-TULIP G52917 E3813763 1082700252917220220110E3813763 IGTCFS-65-1-FEM-TULIP G52917 E3815476 1082700252917220220116E3815476 IGTCFS-65-1-FEM-TULIP G52917 E3815477 1082700252917220220116E3815477 IGTCFS-65-1-FEM-TULIP G52917 E3817125 1082700252917220220123E3817125 IGTCFS-65-1-FEM-TULIP G52917 E3817562 1082700252917220220124E3817562 IGTCFS-65-1-FEM-TULIP G52917 E3818360 1082700252917220220128E3818360 IGTCFS-65-1-FEM-TULIP G52917 E3818361 1082700252917220220128E3818361 IGTCFS-65-1-FEM-TULIP G52917 E3819474 1082700252917220220131E3819474 IGTCFS-65-1-JUG-TULIP G52916 E3814878 1082700252916520220115E3814878 IGTCFS-65-1-JUG-TULIP G52916 E3815833 1082700252916520220122E3815833 IGTCFS-65-1-JUG-TULIP G52916 E3815866 1082700252916520220122E3815866 IGTCFS-65-1-JUG-TULIP G52916 E3817486 1082700252916520220124E3817486 IGTCFS-65-1-JUG-TULIP G52916 E3817561 1082700252916520220124E3817561 IGTCFS-65-1-JUG-TULIP G52916 E3818790 1082700252916520220129E3818790 IGTCFS-65-1-UNI-TULIP G52918 E3820413 1082700252918920220205E3820413 IGTCFS-65-1-UNI-TULIP G52918 E3821142 1082700252918920220207E3821142 IGTCFS-65-1-UNI-TULIP G52918 E3821784 1082700252918920220211E3821784 IGTCFS-65-1-UNI-TULIP G52918 E3822904 1082700252918920220214E3822904 IGTCFS-65-1-UNI-TULIP G52918 E3825281 1082700252918920220226E3825281 IGTCFS-65-1-UNI-TULIP G52918 E3826948 1082700252918920220228E3826948 IGTCFS-65-1-FEM-TULIP G52917 E3820061 1082700252917220220204E3820061 IGTCFS-65-1-FEM-TULIP G52917 E3820252 1082700252917220220204E3820252 IGTCFS-65-1-FEM-TULIP G52917 E3821156 1082700252917220220207E3821156 IGTCFS-65-1-FEM-TULIP G52917 E3821463 1082700252917220220208E3821463 IGTCFS-65-1-FEM-TULIP G52917 E3821811 1082700252917220220211E3821811 IGTCFS-65-1-FEM-TULIP G52917 E3822125 1082700252917220220212E3822125 IGTCFS-65-1-FEM-TULIP G52917 E3822898 1082700252917220220214E3822898 IGTCFS-65-1-FEM-TULIP G52917 E3825267 1082700252917220220225E3825267 IGTCFS-65-1-FEM-TULIP G52917 E3826004 1082700252917220220226E3826004 IGTCFS-65-1-FEM-TULIP G52917 E3826512 1082700252917220220227E3826512 IGTCFS-65-1-FEM-TULIP G52917 E3826938 1082700252917220220228E3826938 IGTCFS-65-1-JUG-TULIP G52916 E3819879 1082700252916520220201E3819879 IGTCFS-65-1-JUG-TULIP G52916 E3819880 1082700252916520220201E3819880 IGTCFS-65-1-JUG-TULIP G52916 E3820105 1082700252916520220204E3820105 IGTCFS-65-1-JUG-TULIP G52916 E3820736 1082700252916520220206E3820736 IGTCFS-65-1-JUG-TULIP G52916 E3820824 1082700252916520220206E3820824 IGTCFS-65-1-JUG-TULIP G52916 E3821141 1082700252916520220207E3821141 IGTCFS-65-1-JUG-TULIP G52916 E3821812 1082700252916520220211E3821812 IGTCFS-65-1-JUG-TULIP G52916 E3822124 1082700252916520220212E3822124 IGTCFS-65-1-JUG-TULIP G52916 E3822544 1082700252916520220213E3822544 IGTCFS-65-1-JUG-TULIP G52916 E3823251 1082700252916520220215E3823251 IGTCFS-65-1-JUG-TULIP G52916 E3825280 1082700252916520220225E3825280 IGTCFS-65-1-JUG-TULIP G52916 E3825903 1082700252916520220226E3825903 IGTCFS-65-1-JUG-TULIP G52916 E3825939 1082700252916520220226E3825939 IGTCFS-65-1-JUG-TULIP G52916 E3826432 1082700252916520220227E3826432 IGTCFS-65-1-JUG-TULIP G52916 E3826939 1082700252916520220228E3826939 IGTCFS-65-1-UNI-TULIP G52918 E3830691 1082700252918920220312E3830691 IGTCFS-65-1-UNI-TULIP G52918 E3830692 1082700252918920220312E3830692 IGTCFS-65-1-UNI-TULIP G52918 E3831343 1082700252918920220313E3831343 IGTCFS-65-1-UNI-TULIP G52918 E3832271 1082700252918920220315E3832271 IGTCFS-65-1-UNI-TULIP G52918 E3832885 1082700252918920220318E3832885 IGTCFS-65-1-UNI-TULIP G52918 E3833908 1082700252918920220320E3833908 IGTCFS-65-1-UNI-TULIP G52918 E3836608 1082700252918920220327E3836608 IGTCFS-65-1-UNI-TULIP G52918 E3837201 1082700252918920220328E3837201 IGTCFS-65-1-FEM-TULIP G52917 E3827399 1082700252917220220301E3827399 IGTCFS-65-1-FEM-TULIP G52917 E3831235 1082700252917220220313E3831235 IGTCFS-65-1-FEM-TULIP G52917 E3831777 1082700252917220220314E3831777 IGTCFS-65-1-FEM-TULIP G52917 E3831845 1082700252917220220314E3831845 IGTCFS-65-1-FEM-TULIP G52917 E3832884 1082700252917220220318E3832884 IGTCFS-65-1-FEM-TULIP G52917 E3833491 1082700252917220220319E3833491 IGTCFS-65-1-FEM-TULIP G52917 E3833996 1082700252917220220320E3833996 IGTCFS-65-1-FEM-TULIP G52917 E3835423 1082700252917220220325E3835423 IGTCFS-65-1-FEM-TULIP G52917 E3835494 1082700252917220220325E3835494 IGTCFS-65-1-FEM-TULIP G52917 E3836013 1082700252917220220326E3836013 IGTCFS-65-1-FEM-TULIP G52917 E3836636 1082700252917220220327E3836636 IGTCFS-65-1-FEM-TULIP G52917 E3837153 1082700252917220220328E3837153 IGTCFS-65-1-FEM-TULIP G52917 E3838005 1082700252917220220329E3838005 IGTCFS-65-1-JUG-TULIP G52916 E3827747 1082700252916520220304E3827747 IGTCFS-65-1-JUG-TULIP G52916 E3831778 1082700252916520220314E3831778 IGTCFS-65-1-JUG-TULIP G52916 E3832304 1082700252916520220315E3832304 IGTCFS-65-1-JUG-TULIP G52916 E3833437 1082700252916520220319E3833437 IGTCFS-65-1-JUG-TULIP G52916 E3833505 1082700252916520220319E3833505 IGTCFS-65-1-JUG-TULIP G52916 E3833536 1082700252916520220319E3833536 IGTCFS-65-1-JUG-TULIP G52916 E3834015 1082700252916520220320E3834015 IGTCFS-65-1-JUG-TULIP G52916 E3835562 1082700252916520220325E3835562 IGTCFS-65-1-JUG-TULIP G52916 E3836157 1082700252916520220326E3836157 IGTCFS-65-1-JUG-TULIP G52916 E3836517 1082700252916520220327E3836517 IGTCFS-65-1-JUG-TULIP G52916 E3837955 1082700252916520220329E3837955 IGTCFS-65-1-JUG-TULIP G52916 E3838006 1082700252916520220329E3838006 IGTCFS-65-1-UNI-TULIP G52918 E3838873 1082700252918920220402E3838873 IGTCFS-65-1-UNI-TULIP G52918 E3839398 1082700252918920220403E3839398 IGTCFS-65-1-UNI-TULIP G52918 E3841740 1082700252918920220409E3841740 IGTCFS-65-1-UNI-TULIP G52918 E3843594 1082700252918920220412E3843594 IGTCFS-65-1-UNI-TULIP G52918 E3846865 1082700252918920220426E3846865 IGTCFS-65-1-UNI-TULIP G52918 E3847342 1082700252918920220429E3847342 IGTCFS-65-1-FEM-TULIP G52917 E3838416 1082700252917220220401E3838416 IGTCFS-65-1-FEM-TULIP G52917 E3838923 1082700252917220220402E3838923 IGTCFS-65-1-FEM-TULIP G52917 E3840225 1082700252917220220404E3840225 IGTCFS-65-1-FEM-TULIP G52917 E3840828 1082700252917220220405E3840828 IGTCFS-65-1-FEM-TULIP G52917 E3841731 1082700252917220220409E3841731 IGTCFS-65-1-FEM-TULIP G52917 E3842191 1082700252917220220410E3842191 IGTCFS-65-1-FEM-TULIP G52917 E3843633 1082700252917220220412E3843633 IGTCFS-65-1-FEM-TULIP G52917 E3844250 1082700252917220220416E3844250 IGTCFS-65-1-FEM-TULIP G52917 E3844561 1082700252917220220417E3844561 IGTCFS-65-1-FEM-TULIP G52917 E3845029 1082700252917220220423E3845029 IGTCFS-65-1-FEM-TULIP G52917 E3845680 1082700252917220220424E3845680 IGTCFS-65-1-FEM-TULIP G52917 E3846894 1082700252917220220426E3846894 IGTCFS-65-1-FEM-TULIP G52917 E3847303 1082700252917220220429E3847303 IGTCFS-65-1-FEM-TULIP G52917 E3847748 1082700252917220220430E3847748 IGTCFS-65-1-JUG-TULIP G52916 E3838383 1082700252916520220401E3838383 IGTCFS-65-1-JUG-TULIP G52916 E3838874 1082700252916520220402E3838874 IGTCFS-65-1-JUG-TULIP G52916 E3840814 1082700252916520220405E3840814 IGTCFS-65-1-JUG-TULIP G52916 E3841694 1082700252916520220409E3841694 IGTCFS-65-1-JUG-TULIP G52916 E3841695 1082700252916520220409E3841695 IGTCFS-65-1-JUG-TULIP G52916 E3842192 1082700252916520220410E3842192 IGTCFS-65-1-JUG-TULIP G52916 E3844512 1082700252916520220417E3844512 IGTCFS-65-1-JUG-TULIP G52916 E3844513 1082700252916520220417E3844513 IGTCFS-65-1-JUG-TULIP G52916 E3846247 1082700252916520220425E3846247 IGTCFS-65-1-JUG-TULIP G52916 E3847246 1082700252916520220429E3847246 IGTCFS-65-1-UNI-TULIP G52918 E3848786 1082700252918920220502E3848786 IGTCFS-65-1-UNI-TULIP G52918 E3850691 1082700252918920220507E3850691 IGTCFS-65-1-UNI-TULIP G52918 E3855390 1082700252918920220521E3855390 IGTCFS-65-1-UNI-TULIP G52918 E3857284 1082700252918920220524E3857284 IGTCFS-65-1-UNI-TULIP G52918 E3859338 1082700252918920220531E3859338 IGTCFS-65-1-FEM-TULIP G52917 E3848711 1082700252917220220502E3848711 IGTCFS-65-1-FEM-TULIP G52917 E3849811 1082700252917220220506E3849811 IGTCFS-65-1-FEM-TULIP G52917 E3850306 1082700252917220220507E3850306 IGTCFS-65-1-FEM-TULIP G52917 E3850307 1082700252917220220507E3850307 IGTCFS-65-1-FEM-TULIP G52917 E3851540 1082700252917220220509E3851540 IGTCFS-65-1-FEM-TULIP G52917 E3851541 1082700252917220220509E3851541 IGTCFS-65-1-FEM-TULIP G52917 E3855832 1082700252917220220522E3855832 IGTCFS-65-1-FEM-TULIP G52917 E3857907 1082700252917220220527E3857907 IGTCFS-65-1-FEM-TULIP G52917 E3858477 1082700252917220220528E3858477 IGTCFS-65-1-FEM-TULIP G52917 E3859131 1082700252917220220529E3859131 IGTCFS-65-1-FEM-TULIP G52917 E3859351 1082700252917220220531E3859351 IGTCFS-65-1-JUG-TULIP G52916 E3848066 1082700252916520220501E3848066 IGTCFS-65-1-JUG-TULIP G52916 E3848073 1082700252916520220501E3848073 IGTCFS-65-1-JUG-TULIP G52916 E3848819 1082700252916520220502E3848819 IGTCFS-65-1-JUG-TULIP G52916 E3849561 1082700252916520220506E3849561 IGTCFS-65-1-JUG-TULIP G52916 E3850454 1082700252916520220507E3850454 IGTCFS-65-1-JUG-TULIP G52916 E3850690 1082700252916520220508E3850690 IGTCFS-65-1-JUG-TULIP G52916 E3851347 1082700252916520220509E3851347 IGTCFS-65-1-JUG-TULIP G52916 E3851348 1082700252916520220509E3851348 IGTCFS-65-1-JUG-TULIP G52916 E3851893 1082700252916520220510E3851893 IGTCFS-65-1-JUG-TULIP G52916 E3856529 1082700252916520220523E3856529 IGTCFS-65-1-JUG-TULIP G52916 E3857936 1082700252916520220527E3857936 IGTCFS-65-1-JUG-TULIP G52916 E3858370 1082700252916520220528E3858370 IGTCFS-65-1-JUG-TULIP G52916 E3859132 1082700252916520220529E3859132 IGTCFS-65-1-UNI-TULIP G52918 E3860499 1082700252918920220604E3860499 IGTCFS-65-1-UNI-TULIP G52918 E3861511 1082700252918920220607E3861511 IGTCFS-65-1-UNI-TULIP G52918 E3862912 1082700252918920220612E3862912 IGTCFS-65-1-UNI-TULIP G52918 E3864792 1082700252918920220617E3864792 IGTCFS-65-1-UNI-TULIP G52918 E3866427 1082700252918920220619E3866427 IGTCFS-65-1-UNI-TULIP G52918 E3867167 1082700252918920220621E3867167 IGTCFS-65-1-UNI-TULIP G52918 E3869595 1082700252918920220627E3869595 IGTCFS-65-1-FEM-TULIP G52917 E3859845 (01)10827002529172(17)220306(10)E3859845 IGTCFS-65-1-FEM-TULIP G52917 E3860395 1082700252917220220604E3860395 IGTCFS-65-1-FEM-TULIP G52917 E3861510 1082700252917220220607E3861510 IGTCFS-65-1-FEM-TULIP G52917 E3862082 1082700252917220220611E3862082 IGTCFS-65-1-FEM-TULIP G52917 E3862911 1082700252917220220612E3862911 IGTCFS-65-1-FEM-TULIP G52917 E3863953 1082700252917220220614E3863953 IGTCFS-65-1-FEM-TULIP G52917 E3865557 1082700252917220220618E3865557 IGTCFS-65-1-FEM-TULIP G52917 E3866116 1082700252917220220619E3866116 IGTCFS-65-1-FEM-TULIP G52917 E3867856 1082700252917220220624E3867856 IGTCFS-65-1-FEM-TULIP G52917 E3868500 1082700252917220220625E3868500 IGTCFS-65-1-FEM-TULIP G52917 E3868997 1082700252917220220626E3868997 IGTCFS-65-1-FEM-TULIP G52917 E3869594 1082700252917220220627E3869594 IGTCFS-65-1-JUG-TULIP G52916 E3859697 1082700252916520220603E3859697 IGTCFS-65-1-JUG-TULIP G52916 E3860406 1082700252916520220604E3860406 IGTCFS-65-1-JUG-TULIP G52916 E3860442 1082700252916520220604E3860442 IGTCFS-65-1-JUG-TULIP G52916 E3860443 1082700252916520220604E3860443 IGTCFS-65-1-JUG-TULIP G52916 E3862110 1082700252916520220611E3862110 IGTCFS-65-1-JUG-TULIP G52916 E3863563 1082700252916520220613E3863563 IGTCFS-65-1-JUG-TULIP G52916 E3865448 1082700252916520220618E3865448 IGTCFS-65-1-JUG-TULIP G52916 E3865558 1082700252916520220618E3865558 IGTCFS-65-1-JUG-TULIP G52916 E3866779 1082700252916520220620E3866779 IGTCFS-65-1-JUG-TULIP G52916 E3867857 1082700252916520220624E3867857 IGTCFS-65-1-JUG-TULIP G52916 E3868422 1082700252916520220625E3868422 IGTCFS-65-1-JUG-TULIP G52916 E3868998 1082700252916520220626E3868998 IGTCFS-65-1-JUG-TULIP G52916 E3869494 1082700252916520220627E3869494 IGTCFS-65-1-UNI-TULIP G52918 E3870714 1082700252918920220701E3870714 IGTCFS-65-1-UNI-TULIP G52918 E3872515 1082700252918920220704E3872515 IGTCFS-65-1-UNI-TULIP G52918 E3873982 1082700252918920220710E3873982 IGTCFS-65-1-UNI-TULIP G52918 E3876732 1082700252918920220717E3876732 IGTCFS-65-1-UNI-TULIP G52918 E3877609 1082700252918920220719E3877609 IGTCFS-65-1-UNI-TULIP G52918 E3880488 1082700252918920220730E3880488 IGTCFS-65-1-FEM-TULIP G52917 E3871118 1082700252917220220702E3871118 IGTCFS-65-1-FEM-TULIP G52917 E3871215 1082700252917220220702E3871215 IGTCFS-65-1-FEM-TULIP G52917 E3871906 1082700252917220220703E3871906 IGTCFS-65-1-FEM-TULIP G52917 E3873314 1082700252917220220708E3873314 IGTCFS-65-1-FEM-TULIP G52917 E3875048 1082700252917220220711E3875048 IGTCFS-65-1-FEM-TULIP G52917 E3877303 1082700252917220220718E3877303 IGTCFS-65-1-FEM-TULIP G52917 E3877608 1082700252917220220719E3877608 IGTCFS-65-1-FEM-TULIP G52917 E3878399 1082700252917220220723E3878399 IGTCFS-65-1-FEM-TULIP G52917 E3878920 1082700252917220220724E3878920 IGTCFS-65-1-FEM-TULIP G52917 E3879513 1082700252917220220725E3879513 IGTCFS-65-1-FEM-TULIP G52917 E3880095 1082700252917220220729E3880095 IGTCFS-65-1-FEM-TULIP G52917 E3881044 1082700252917220220731E3881044 IGTCFS-65-1-JUG-TULIP G52916 E3871119 1082700252916520220702E3871119 IGTCFS-65-1-JUG-TULIP G52916 E3871216 1082700252916520220702E3871216 IGTCFS-65-1-JUG-TULIP G52916 E3871825 1082700252916520220703E3871825 IGTCFS-65-1-JUG-TULIP G52916 E3872793 1082700252916520220705E3872793 IGTCFS-65-1-JUG-TULIP G52916 E3874483 1082700252916520220710E3874483 IGTCFS-65-1-JUG-TULIP G52916 E3874559 1082700252916520220710E3874559 IGTCFS-65-1-JUG-TULIP G52916 E3876308 1082700252916520220716E3876308 IGTCFS-65-1-JUG-TULIP G52916 E3878019 1082700252916520220722E3878019 IGTCFS-65-1-JUG-TULIP G52916 E3878553 1082700252916520220723E3878553 IGTCFS-65-1-JUG-TULIP G52916 E3878943 1082700252916520220724E3878943 IGTCFS-65-1-JUG-TULIP G52916 E3879753 1082700252916520220726E3879753 IGTCFS-65-1-JUG-TULIP G52916 E3881016 1082700252916520220731E3881016 IGTCFS-65-1-UNI-TULIP G52918 E3882718 1082700252918920220805E3882718 IGTCFS-65-1-UNI-TULIP G52918 E3883877 1082700252918920220808E3883877 IGTCFS-65-1-UNI-TULIP G52918 E3886771 1082700252918920220816E3886771 IGTCFS-65-1-UNI-TULIP G52918 E3890987 1082700252918920220828E3890987 IGTCFS-65-1-FEM-TULIP G52917 E3883070 1082700252917220220806E3883070 IGTCFS-65-1-FEM-TULIP G52917 E3883896 1082700252917220220808E3883896 IGTCFS-65-1-FEM-TULIP G52917 E3886370 1082700252917220220815E3886370 IGTCFS-65-1-FEM-TULIP G52917 E3887001 1082700252917220220819E3887001 IGTCFS-65-1-FEM-TULIP G52917 E3887002 1082700252917220220819E3887002 IGTCFS-65-1-FEM-TULIP G52917 E3887679 1082700252917220220820E3887679 IGTCFS-65-1-FEM-TULIP G52917 E3888996 1082700252917220220822E3888996 IGTCFS-65-1-FEM-TULIP G52917 E3891512 1082700252917220220829E3891512 IGTCFS-65-1-FEM-TULIP G52917 E3891513 1082700252917220220829E3891513 IGTCFS-65-1-FEM-TULIP G52917 E3891897 1082700252917220220830E3891897 IGTCFS-65-1-FEM-TULIP G52917 E3891898 1082700252917220220830E3891898 IGTCFS-65-1-JUG-TULIP G52916 E3881620 1082700252916520220801E3881620 IGTCFS-65-1-JUG-TULIP G52916 E3882381 1082700252916520220802E3882381 IGTCFS-65-1-JUG-TULIP G52916 E3884199 1082700252916520220809E3884199 IGTCFS-65-1-JUG-TULIP G52916 E3886369 1082700252916520220815E3886369 IGTCFS-65-1-JUG-TULIP G52916 E3886755 1082700252916520220816E3886755 IGTCFS-65-1-JUG-TULIP G52916 E3887146 1082700252916520220819E3887146 IGTCFS-65-1-JUG-TULIP G52916 E3887694 1082700252916520220820E3887694 IGTCFS-65-1-JUG-TULIP G52916 E3889372 1082700252916520220823E3889372 IGTCFS-65-1-JUG-TULIP G52916 E3891012 1082700252916520220828E3891012 IGTCFS-65-1-UNI-TULIP G52918 E3899596 (01)10827002529189(17)221909(10)E3899596 IGTCFS-65-1-UNI-TULIP G52918 E3895481 1082700252918920220910E3895481 IGTCFS-65-1-UNI-TULIP G52918 E3897136 1082700252918920220913E3897136 IGTCFS-65-1-UNI-TULIP G52918 E3898888 1082700252918920220918E3898888 IGTCFS-65-1-UNI-TULIP G52918 E3902096 1082700252918920220925E3902096 IGTCFS-65-1-UNI-TULIP G52918 E3902675 1082700252918920220926E3902675 IGTCFS-65-1-FEM-TULIP G52917 E3892917 (01)10827002529172(17)220309(10)E3892917 IGTCFS-65-1-FEM-TULIP G52917 E3892916 1082700252917220220903E3892916 IGTCFS-65-1-FEM-TULIP G52917 E3893094 1082700252917220220903E3893094 IGTCFS-65-1-FEM-TULIP G52917 E3893707 1082700252917220220904E3893707 IGTCFS-65-1-FEM-TULIP G52917 E3893728 1082700252917220220905E3893728 IGTCFS-65-1-FEM-TULIP G52917 E3897782 1082700252917220220916E3897782 IGTCFS-65-1-FEM-TULIP G52917 E3899108 1082700252917220220918E3899108 IGTCFS-65-1-FEM-TULIP G52917 E3900108 1082700252917220220920E3900108 IGTCFS-65-1-FEM-TULIP G52917 E3902674 1082700252917220220926E3902674 IGTCFS-65-1-FEM-TULIP G52917 E3903705 1082700252917220220927E3903705 IGTCFS-65-1-FEM-TULIP G52917 E3904350 1082700252917220220930E3904350 IGTCFS-65-1-JUG-TULIP G52916 E3892185 1082700252916520220902E3892185 IGTCFS-65-1-JUG-TULIP G52916 E3892186 1082700252916520220902E3892186 IGTCFS-65-1-JUG-TULIP G52916 E3892331 1082700252916520220902E3892331 IGTCFS-65-1-JUG-TULIP G52916 E3894195 1082700252916520220905E3894195 IGTCFS-65-1-JUG-TULIP G52916 E3894473 1082700252916520220906E3894473 IGTCFS-65-1-JUG-TULIP G52916 E3896108 1082700252916520220911E3896108 IGTCFS-65-1-JUG-TULIP G52916 E3896109 1082700252916520220911E3896109 IGTCFS-65-1-JUG-TULIP G52916 E3897129 1082700252916520220913E3897129 IGTCFS-65-1-JUG-TULIP G52916 E3898163 1082700252916520220917E3898163 IGTCFS-65-1-JUG-TULIP G52916 E3898396 1082700252916520220917E3898396 IGTCFS-65-1-JUG-TULIP G52916 E3899562 1082700252916520220919E3899562 IGTCFS-65-1-JUG-TULIP G52916 E3899595 1082700252916520220919E3899595 IGTCFS-65-1-JUG-TULIP G52916 E3901522 1082700252916520220924E3901522 IGTCFS-65-1-JUG-TULIP G52916 E3903704 1082700252916520220927E3903704 IGTCFS-65-1-JUG-TULIP G52916 E3904339 1082700252916520220930E3904339 IGTCFS-65-1-UNI-TULIP G52918 E3905960 1082700252918920221004E3905960 IGTCFS-65-1-UNI-TULIP G52918 E3907187 1082700252918920221008E3907187 IGTCFS-65-1-UNI-TULIP G52918 E3909031 1082700252918920221014E3909031 IGTCFS-65-1-UNI-TULIP G52918 E3911157 1082700252918920221017E3911157 IGTCFS-65-1-UNI-TULIP G52918 E3912451 1082700252918920221022E3912451 IGTCFS-65-1-UNI-TULIP G52918 E3914289 1082700252918920221025E3914289 IGTCFS-65-1-UNI-TULIP G52918 E3916290 1082700252918920221031E3916290 IGTCFS-65-1-FEM-TULIP G52917 E3905890 1082700252917220221004E3905890 IGTCFS-65-1-FEM-TULIP G52917 E3906655 1082700252917220221007E3906655 IGTCFS-65-1-FEM-TULIP G52917 E3906726 1082700252917220221007E3906726 IGTCFS-65-1-FEM-TULIP G52917 E3907811 1082700252917220221009E3907811 IGTCFS-65-1-FEM-TULIP G52917 E3907892 1082700252917220221009E3907892 IGTCFS-65-1-FEM-TULIP G52917 E3909461 1082700252917220221014E3909461 IGTCFS-65-1-FEM-TULIP G52917 E3910045 1082700252917220221015E3910045 IGTCFS-65-1-FEM-TULIP G52917 E3913162 1082700252917220221023E3913162 IGTCFS-65-1-FEM-TULIP G52917 E3913528 1082700252917220221024E3913528 IGTCFS-65-1-FEM-TULIP G52917 E3913529 1082700252917220221024E3913529 IGTCFS-65-1-FEM-TULIP G52917 E3915973 1082700252917220221030E3915973 IGTCFS-65-1-FEM-TULIP G52917 E3916472 1082700252917220221031E3916472 IGTCFS-65-1-JUG-TULIP G52916 E3905891 1082700252916520221004E3905891 IGTCFS-65-1-JUG-TULIP G52916 E3906552 1082700252916520221007E3906552 IGTCFS-65-1-JUG-TULIP G52916 E3907272 1082700252916520221008E3907272 IGTCFS-65-1-JUG-TULIP G52916 E3907729 1082700252916520221009E3907729 IGTCFS-65-1-JUG-TULIP G52916 E3909030 1082700252916520221014E3909030 IGTCFS-65-1-JUG-TULIP G52916 E3909478 1082700252916520221014E3909478 IGTCFS-65-1-JUG-TULIP G52916 E3910046 1082700252916520221015E3910046 IGTCFS-65-1-JUG-TULIP G52916 E3911187 1082700252916520221017E3911187 IGTCFS-65-1-JUG-TULIP G52916 E3914350 1082700252916520221025E3914350 IGTCFS-65-1-JUG-TULIP G52916 E3914790 1082700252916520221028E3914790 IGTCFS-65-1-JUG-TULIP G52916 E3915868 1082700252916520221030E3915868 IGTCFS-65-1-UNI-TULIP G52918 E3927997 (01)10827002529189(17)222911(10)E3927997 IGTCFS-65-1-UNI-TULIP G52918 E3917337 1082700252918920221104E3917337 IGTCFS-65-1-UNI-TULIP G52918 E3918473 1082700252918920221106E3918473 IGTCFS-65-1-UNI-TULIP G52918 E3919513 1082700252918920221108E3919513 IGTCFS-65-1-UNI-TULIP G52918 E3920690 1082700252918920221112E3920690 IGTCFS-65-1-UNI-TULIP G52918 E3921280 1082700252918920221113E3921280 IGTCFS-65-1-UNI-TULIP G52918 E3923493 1082700252918920221119E3923493 IGTCFS-65-1-UNI-TULIP G52918 E3924627 1082700252918920221121E3924627 IGTCFS-65-1-UNI-TULIP G52918 E3926151 1082700252918920221126E3926151 IGTCFS-65-1-UNI-TULIP G52918 E3926394 1082700252918920221126E3926394 IGTCFS-65-1-FEM-TULIP G52917 E3916980 1082700252917220221101E3916980 IGTCFS-65-1-FEM-TULIP G52917 E3917757 1082700252917220221105E3917757 IGTCFS-65-1-FEM-TULIP G52917 E3918419 1082700252917220221106E3918419 IGTCFS-65-1-FEM-TULIP G52917 E3918420 1082700252917220221106E3918420 IGTCFS-65-1-FEM-TULIP G52917 E3919113 1082700252917220221107E3919113 IGTCFS-65-1-FEM-TULIP G52917 E3919512 1082700252917220221108E3919512 IGTCFS-65-1-FEM-TULIP G52917 E3920143 1082700252917220221111E3920143 IGTCFS-65-1-FEM-TULIP G52917 E3920689 1082700252917220221112E3920689 IGTCFS-65-1-FEM-TULIP G52917 E3921279 1082700252917220221113E3921279 IGTCFS-65-1-FEM-TULIP G52917 E3921977 1082700252917220221114E3921977 IGTCFS-65-1-FEM-TULIP G52917 E3922874 1082700252917220221118E3922874 IGTCFS-65-1-FEM-TULIP G52917 E3923470 1082700252917220221119E3923470 IGTCFS-65-1-FEM-TULIP G52917 E3924516 1082700252917220221121E3924516 IGTCFS-65-1-FEM-TULIP G52917 E3924595 1082700252917220221121E3924595 IGTCFS-65-1-FEM-TULIP G52917 E3925197 1082700252917220221122E3925197 IGTCFS-65-1-FEM-TULIP G52917 E3925566 1082700252917220221125E3925566 IGTCFS-65-1-FEM-TULIP G52917 E3926753 1082700252917220221127E3926753 IGTCFS-65-1-FEM-TULIP G52917 E3926754 1082700252917220221127E3926754 IGTCFS-65-1-FEM-TULIP G52917 E3927497 1082700252917220221128E3927497 IGTCFS-65-1-JUG-TULIP G52916 E3916979 1082700252916520221101E3916979 IGTCFS-65-1-JUG-TULIP G52916 E3917299 1082700252916520221104E3917299 IGTCFS-65-1-JUG-TULIP G52916 E3917331 1082700252916520221104E3917331 IGTCFS-65-1-JUG-TULIP G52916 E3917758 1082700252916520221105E3917758 IGTCFS-65-1-JUG-TULIP G52916 E3917915 1082700252916520221105E3917915 IGTCFS-65-1-JUG-TULIP G52916 E3919112 1082700252916520221107E3919112 IGTCFS-65-1-JUG-TULIP G52916 E3919149 1082700252916520221107E3919149 IGTCFS-65-1-JUG-TULIP G52916 E3919579 1082700252916520221108E3919579 IGTCFS-65-1-JUG-TULIP G52916 E3919991 1082700252916520221111E3919991 IGTCFS-65-1-JUG-TULIP G52916 E3920160 1082700252916520221111E3920160 IGTCFS-65-1-JUG-TULIP G52916 E3920760 1082700252916520221113E3920760 IGTCFS-65-1-JUG-TULIP G52916 E3921873 1082700252916520221114E3921873 IGTCFS-65-1-JUG-TULIP G52916 E3923469 1082700252916520221119E3923469 IGTCFS-65-1-JUG-TULIP G52916 E3924011 1082700252916520221120E3924011 IGTCFS-65-1-JUG-TULIP G52916 E3924012 1082700252916520221120E3924012 IGTCFS-65-1-JUG-TULIP G52916 E3925196 1082700252916520221122E3925196 IGTCFS-65-1-JUG-TULIP G52916 E3925585 1082700252916520221125E3925585 IGTCFS-65-1-JUG-TULIP G52916 E3925586 1082700252916520221125E3925586 IGTCFS-65-1-JUG-TULIP G52916 E3927430 1082700252916520221128E3927430 IGTCFS-65-1-JUG-TULIP G52916 E3927492 1082700252916520221128E3927492 IGTCFS-65-1-JUG-TULIP G52916 E3927508 1082700252916520221128E3927508 IGTCFS-65-1-UNI-TULIP G52918 E3928602 1082700252918920221202E3928602 IGTCFS-65-1-UNI-TULIP G52918 E3929671 1082700252918920221204E3929671 IGTCFS-65-1-UNI-TULIP G52918 E3932395 1082700252918920221211E3932395 IGTCFS-65-1-UNI-TULIP G52918 E3934462 1082700252918920221217E3934462 IGTCFS-65-1-UNI-TULIP G52918 E3935923 1082700252918920221223E3935923 IGTCFS-65-1-FEM-TULIP G52917 E3929151 1082700252917220221203E3929151 IGTCFS-65-1-FEM-TULIP G52917 E3929710 1082700252917220221204E3929710 IGTCFS-65-1-FEM-TULIP G52917 E3930479 1082700252917220221206E3930479 IGTCFS-65-1-FEM-TULIP G52917 E3931847 1082700252917220221210E3931847 IGTCFS-65-1-FEM-TULIP G52917 E3933647 1082700252917220221216E3933647 IGTCFS-65-1-FEM-TULIP G52917 E3934480 1082700252917220221217E3934480 IGTCFS-65-1-FEM-TULIP G52917 E3934942 1082700252917220221218E3934942 IGTCFS-65-1-FEM-TULIP G52917 E3935497 1082700252917220221219E3935497 IGTCFS-65-1-FEM-TULIP G52917 E3935922 1082700252917220221223E3935922 IGTCFS-65-1-JUG-TULIP G52916 E3928590 1082700252916520221202E3928590 IGTCFS-65-1-JUG-TULIP G52916 E3929169 1082700252916520221203E3929169 IGTCFS-65-1-JUG-TULIP G52916 E3930201 1082700252916520221205E3930201 IGTCFS-65-1-JUG-TULIP G52916 E3932797 1082700252916520221212E3932797 IGTCFS-65-1-JUG-TULIP G52916 E3933179 1082700252916520221213E3933179 IGTCFS-65-1-JUG-TULIP G52916 E3935448 1082700252916520221219E3935448 IGTCFS-65-1-JUG-TULIP G52916 E3935742 1082700252916520221220E3935742 IGTCFS-65-1-JUG-TULIP G52916 E3935805 1082700252916520221220E3935805 IGTCFS-65-1-UNI-TULIP G52918 E3942705 (01)10827002529189(17)232001(10)E3942705 IGTCFS-65-1-UNI-TULIP G52918 E3937484 1082700252918920230106E3937484 IGTCFS-65-1-UNI-TULIP G52918 E3943842 1082700252918920230122E3943842 IGTCFS-65-1-UNI-TULIP G52918 E3945857 1082700252918920230128E3945857 IGTCFS-65-1-FEM-TULIP G52917 E3937033 1082700252917220230103E3937033 IGTCFS-65-1-FEM-TULIP G52917 E3937804 1082700252917220230107E3937804 IGTCFS-65-1-FEM-TULIP G52917 E3938472 1082700252917220230108E3938472 IGTCFS-65-1-FEM-TULIP G52917 E3938906 1082700252917220230109E3938906 IGTCFS-65-1-FEM-TULIP G52917 E3942218 1082700252917220230117E3942218 IGTCFS-65-1-FEM-TULIP G52917 E3942704 1082700252917220230120E3942704 IGTCFS-65-1-FEM-TULIP G52917 E3943841 1082700252917220230122E3943841 IGTCFS-65-1-FEM-TULIP G52917 E3944826 1082700252917220230124E3944826 IGTCFS-65-1-FEM-TULIP G52917 E3945882 1082700252917220230128E3945882 IGTCFS-65-1-FEM-TULIP G52917 E3946313 1082700252917220230129E3946313 IGTCFS-65-1-FEM-TULIP G52917 E3946314 1082700252917220230129E3946314 IGTCFS-65-1-FEM-TULIP G52917 E3947037 1082700252917220230130E3947037 IGTCFS-65-1-FEM-TULIP G52917 E3947260 1082700252917220230131E3947260 IGTCFS-65-1-JUG-TULIP G52916 E3940661 (01)10827002529165(17)231401(10)E3940661 IGTCFS-65-1-JUG-TULIP G52916 E3936429 1082700252916520230102E3936429 IGTCFS-65-1-JUG-TULIP G52916 E3938019 1082700252916520230107E3938019 IGTCFS-65-1-JUG-TULIP G52916 E3938436 1082700252916520230108E3938436 IGTCFS-65-1-JUG-TULIP G52916 E3938838 1082700252916520230109E3938838 IGTCFS-65-1-JUG-TULIP G52916 E3939891 1082700252916520230113E3939891 IGTCFS-65-1-JUG-TULIP G52916 E3941657 1082700252916520230116E3941657 IGTCFS-65-1-JUG-TULIP G52916 E3942226 1082700252916520230117E3942226 IGTCFS-65-1-JUG-TULIP G52916 E3944341 1082700252916520230123E3944341 IGTCFS-65-1-JUG-TULIP G52916 E3945187 1082700252916520230127E3945187 IGTCFS-65-1-JUG-TULIP G52916 E3945188 1082700252916520230127E3945188 IGTCFS-65-1-JUG-TULIP G52916 E3945883 1082700252916520230128E3945883 IGTCFS-65-1-JUG-TULIP G52916 E3946533 1082700252916520230129E3946533 IGTCFS-65-1-JUG-TULIP G52916 E3947030 1082700252916520230130E3947030 IGTCFS-65-1-JUG-TULIP G52916 E3947239 1082700252916520230131E3947239 IGTCFS-65-1-UNI-TULIP G52918 E3947845 1082700252918920230203E3947845 IGTCFS-65-1-UNI-TULIP G52918 E3949597 1082700252918920230206E3949597 IGTCFS-65-1-UNI-TULIP G52918 E3952229 1082700252918920230213E3952229 IGTCFS-65-1-UNI-TULIP G52918 E3952759 1082700252918920230214E3952759 IGTCFS-65-1-UNI-TULIP G52918 E3953091 1082700252918920230217E3953091 IGTCFS-65-1-UNI-TULIP G52918 E3955465 1082700252918920230221E3955465 IGTCFS-65-1-UNI-TULIP G52918 E3956430 1082700252918920230225E3956430 IGTCFS-65-1-UNI-TULIP G52918 E3957209 1082700252918920230226E3957209 IGTCFS-65-1-FEM-TULIP G52917 E3947741 1082700252917220230203E3947741 IGTCFS-65-1-FEM-TULIP G52917 E3949086 1082700252917220230205E3949086 IGTCFS-65-1-FEM-TULIP G52917 E3949695 1082700252917220230206E3949695 IGTCFS-65-1-FEM-TULIP G52917 E3950523 1082700252917220230210E3950523 IGTCFS-65-1-FEM-TULIP G52917 E3951738 1082700252917220230212E3951738 IGTCFS-65-1-FEM-TULIP G52917 E3952758 1082700252917220230214E3952758 IGTCFS-65-1-FEM-TULIP G52917 E3953090 1082700252917220230217E3953090 IGTCFS-65-1-FEM-TULIP G52917 E3953926 1082700252917220230218E3953926 IGTCFS-65-1-FEM-TULIP G52917 E3954234 1082700252917220230219E3954234 IGTCFS-65-1-FEM-TULIP G52917 E3954731 1082700252917220230220E3954731 IGTCFS-65-1-FEM-TULIP G52917 E3956431 1082700252917220230225E3956431 IGTCFS-65-1-FEM-TULIP G52917 E3957822 1082700252917220230227E3957822 IGTCFS-65-1-FEM-TULIP G52917 E3958152 1082700252917220230228E3958152 IGTCFS-65-1-FEM-TULIP G52917 E3958153 1082700252917220230228E3958153 IGTCFS-65-1-JUG-TULIP G52916 E3947760 1082700252916520230203E3947760 IGTCFS-65-1-JUG-TULIP G52916 E3948419 1082700252916520230204E3948419 IGTCFS-65-1-JUG-TULIP G52916 E3949128 1082700252916520230205E3949128 IGTCFS-65-1-JUG-TULIP G52916 E3950020 1082700252916520230207E3950020 IGTCFS-65-1-JUG-TULIP G52916 E3950021 1082700252916520230207E3950021 IGTCFS-65-1-JUG-TULIP G52916 E3951131 1082700252916520230211E3951131 IGTCFS-65-1-JUG-TULIP G52916 E3951192 1082700252916520230211E3951192 IGTCFS-65-1-JUG-TULIP G52916 E3952122 1082700252916520230213E3952122 IGTCFS-65-1-JUG-TULIP G52916 E3953163 1082700252916520230217E3953163 IGTCFS-65-1-JUG-TULIP G52916 E3953808 1082700252916520230218E3953808 IGTCFS-65-1-JUG-TULIP G52916 E3953809 1082700252916520230218E3953809 IGTCFS-65-1-JUG-TULIP G52916 E3954893 1082700252916520230220E3954893 IGTCFS-65-1-JUG-TULIP G52916 E3954974 1082700252916520230220E3954974 IGTCFS-65-1-JUG-TULIP G52916 E3955466 1082700252916520230221E3955466 IGTCFS-65-1-JUG-TULIP G52916 E3955757 1082700252916520230224E3955757 IGTCFS-65-1-JUG-TULIP G52916 E3955853 1082700252916520230224E3955853 IGTCFS-65-1-JUG-TULIP G52916 E3956377 1082700252916520230225E3956377 IGTCFS-65-1-JUG-TULIP G52916 E3957057 1082700252916520230226E3957057 IGTCFS-65-1-JUG-TULIP G52916 E3957829 1082700252916520230227E3957829 IGTCFS-65-1-UNI-TULIP G52918 E3958365 1082700252918920230302E3958365 IGTCFS-65-1-UNI-TULIP G52918 E3958366 1082700252918920230302E3958366 IGTCFS-65-1-UNI-TULIP G52918 E3960328 1082700252918920230305E3960328 IGTCFS-65-1-UNI-TULIP G52918 E3961932 1082700252918920230310E3961932 IGTCFS-65-1-UNI-TULIP G52918 E3966333 1082700252918920230323E3966333 IGTCFS-65-1-UNI-TULIP G52918 E3969305 1082700252918920230330E3969305 IGTCFS-65-1-FEM-TULIP G52917 E3959150 1082700252917220230303E3959150 IGTCFS-65-1-FEM-TULIP G52917 E3959151 1082700252917220230303E3959151 IGTCFS-65-1-FEM-TULIP G52917 E3959723 1082700252917220230304E3959723 IGTCFS-65-1-FEM-TULIP G52917 E3960406 1082700252917220230305E3960406 IGTCFS-65-1-FEM-TULIP G52917 E3960955 1082700252917220230306E3960955 IGTCFS-65-1-FEM-TULIP G52917 E3961222 1082700252917220230309E3961222 IGTCFS-65-1-FEM-TULIP G52917 E3962454 1082700252917220230311E3962454 IGTCFS-65-1-FEM-TULIP G52917 E3963667 1082700252917220230316E3963667 IGTCFS-65-1-FEM-TULIP G52917 E3967079 1082700252917220230324E3967079 IGTCFS-65-1-FEM-TULIP G52917 E3968861 1082700252917220230327E3968861 IGTCFS-65-1-JUG-TULIP G52916 E3958603 1082700252916520230302E3958603 IGTCFS-65-1-JUG-TULIP G52916 E3959092 1082700252916520230303E3959092 IGTCFS-65-1-JUG-TULIP G52916 E3959722 1082700252916520230304E3959722 IGTCFS-65-1-JUG-TULIP G52916 E3959746 1082700252916520230304E3959746 IGTCFS-65-1-JUG-TULIP G52916 E3960982 1082700252916520230306E3960982 IGTCFS-65-1-JUG-TULIP G52916 E3961205 1082700252916520230309E3961205 IGTCFS-65-1-JUG-TULIP G52916 E3961223 1082700252916520230309E3961223 IGTCFS-65-1-JUG-TULIP G52916 E3961952 1082700252916520230310E3961952 IGTCFS-65-1-JUG-TULIP G52916 E3962604 1082700252916520230311E3962604 IGTCFS-65-1-JUG-TULIP G52916 E3963263 1082700252916520230312E3963263 IGTCFS-65-1-JUG-TULIP G52916 E3966957 1082700252916520230324E3966957 IGTCFS-65-1-JUG-TULIP G52916 E3967450 1082700252916520230325E3967450 IGTCFS-65-1-UNI-TULIP G52918 E3974538 1082700252918920230415E3974538 IGTCFS-65-1-UNI-TULIP G52918 E3975836 1082700252918920230420E3975836 IGTCFS-65-1-UNI-TULIP G52918 E3978194 1082700252918920230427E3978194 IGTCFS-65-1-UNI-TULIP G52918 E3978541 1082700252918920230428E3978541 IGTCFS-65-1-UNI-TULIP G52918 E3979351 1082700252918920230430E3979351 IGTCFS-65-1-FEM-TULIP G52917 E3971683 1082700252917220230403E3971683 IGTCFS-65-1-FEM-TULIP G52917 E3971684 1082700252917220230403E3971684 IGTCFS-65-1-FEM-TULIP G52917 E3972382 1082700252917220230406E3972382 IGTCFS-65-1-FEM-TULIP G52917 E3975084 1082700252917220230416E3975084 IGTCFS-65-1-FEM-TULIP G52917 E3975576 1082700252917220230417E3975576 IGTCFS-65-1-FEM-TULIP G52917 E3976456 1082700252917220230421E3976456 IGTCFS-65-1-FEM-TULIP G52917 E3976993 1082700252917220230422E3976993 IGTCFS-65-1-FEM-TULIP G52917 E3978193 1082700252917220230427E3978193 IGTCFS-65-1-FEM-TULIP G52917 E3978540 1082700252917220230428E3978540 IGTCFS-65-1-FEM-TULIP G52917 E3978895 1082700252917220230429E3978895 IGTCFS-65-1-FEM-TULIP G52917 E3979130 1082700252917220230430E3979130 IGTCFS-65-1-FEM-TULIP G52917 E3979131 1082700252917220230430E3979131 IGTCFS-65-1-JUG-TULIP G52916 E3970502 1082700252916520230401E3970502 IGTCFS-65-1-JUG-TULIP G52916 E3970640 1082700252916520230401E3970640 IGTCFS-65-1-JUG-TULIP G52916 E3974032 1082700252916520230414E3974032 IGTCFS-65-1-JUG-TULIP G52916 E3975577 1082700252916520230417E3975577 IGTCFS-65-1-JUG-TULIP G52916 E3975814 1082700252916520230420E3975814 IGTCFS-65-1-JUG-TULIP G52916 E3976774 1082700252916520230422E3976774 IGTCFS-65-1-JUG-TULIP G52916 E3978520 1082700252916520230428E3978520 IGTCFS-65-1-JUG-TULIP G52916 E3978903 1082700252916520230429E3978903 IGTCFS-65-1-JUG-TULIP G52916 E3979350 1082700252916520230430E3979350 IGTCFS-65-1-UNI-TULIP G52918 E3980490 1082700252918920230504E3980490 IGTCFS-65-1-UNI-TULIP G52918 E3981243 1082700252918920230505E3981243 IGTCFS-65-1-UNI-TULIP G52918 E3981967 1082700252918920230507E3981967 IGTCFS-65-1-UNI-TULIP G52918 E3984329 1082700252918920230513E3984329 IGTCFS-65-1-FEM-TULIP G52917 E3979589 1082700252917220230501E3979589 IGTCFS-65-1-FEM-TULIP G52917 E3980304 1082700252917220230504E3980304 IGTCFS-65-1-FEM-TULIP G52917 E3981607 1082700252917220230506E3981607 IGTCFS-65-1-FEM-TULIP G52917 E3981966 1082700252917220230507E3981966 IGTCFS-65-1-FEM-TULIP G52917 E3983309 1082700252917220230511E3983309 IGTCFS-65-1-FEM-TULIP G52917 E3983310 1082700252917220230511E3983310 IGTCFS-65-1-FEM-TULIP G52917 E3984694 1082700252917220230514E3984694 IGTCFS-65-1-FEM-TULIP G52917 E3986278 1082700252917220230525E3986278 IGTCFS-65-1-FEM-TULIP G52917 E3987013 1082700252917220230526E3987013 IGTCFS-65-1-FEM-TULIP G52917 E3987780 1082700252917220230528E3987780 IGTCFS-65-1-JUG-TULIP G52916 E3979588 1082700252916520230501E3979588 IGTCFS-65-1-JUG-TULIP G52916 E3979631 1082700252916520230501E3979631 IGTCFS-65-1-JUG-TULIP G52916 E3979632 1082700252916520230501E3979632 IGTCFS-65-1-JUG-TULIP G52916 E3980515 1082700252916520230504E3980515 IGTCFS-65-1-JUG-TULIP G52916 E3981244 1082700252916520230505E3981244 IGTCFS-65-1-JUG-TULIP G52916 E3983792 1082700252916520230512E3983792 IGTCFS-65-1-JUG-TULIP G52916 E3983793 1082700252916520230512E3983793 IGTCFS-65-1-JUG-TULIP G52916 E3984415 1082700252916520230513E3984415 IGTCFS-65-1-JUG-TULIP G52916 E3986279 1082700252916520230525E3986279 IGTCFS-65-1-JUG-TULIP G52916 E3986966 1082700252916520230526E3986966 IGTCFS-65-1-JUG-TULIP G52916 E3988671 1082700252916520230529E3988671 IGTCFS-65-1-UNI-TULIP G52918 E3989536 1082700252918920230609E3989536 IGTCFS-65-1-UNI-TULIP G52918 E3990559 1082700252918920230612E3990559 IGTCFS-65-1-UNI-TULIP G52918 E3993076 1082700252918920230626E3993076 IGTCFS-65-1-FEM-TULIP G52917 E3989262 1082700252917220230608E3989262 IGTCFS-65-1-FEM-TULIP G52917 E3989535 1082700252917220230609E3989535 IGTCFS-65-1-FEM-TULIP G52917 E3990329 1082700252917220230611E3990329 IGTCFS-65-1-FEM-TULIP G52917 E3991440 1082700252917220230622E3991440 IGTCFS-65-1-FEM-TULIP G52917 E3991855 1082700252917220230623E3991855 IGTCFS-65-1-FEM-TULIP G52917 E3992360 1082700252917220230624E3992360 IGTCFS-65-1-FEM-TULIP G52917 E3992811 1082700252917220230625E3992811 IGTCFS-65-1-FEM-TULIP G52917 E3993077 1082700252917220230626E3993077 IGTCFS-65-1-JUG-TULIP G52916 E3989229 1082700252916520230608E3989229 IGTCFS-65-1-JUG-TULIP G52916 E3989824 1082700252916520230610E3989824 IGTCFS-65-1-JUG-TULIP G52916 E3989825 1082700252916520230610E3989825 IGTCFS-65-1-JUG-TULIP G52916 E3990280 1082700252916520230611E3990280 IGTCFS-65-1-JUG-TULIP G52916 E3993335 1082700252916520230629E3993335 IGTCFS-65-1-JUG-TULIP G52916 E3993490 1082700252916520230629E3993490 IGTCFS-65-1-UNI-TULIP G52918 E3994995 1082700252918920230702E3994995 IGTCFS-65-1-UNI-TULIP G52918 E3996239 1082700252918920230707E3996239 IGTCFS-65-1-UNI-TULIP G52918 E3997449 1082700252918920230710E3997449 IGTCFS-65-1-UNI-TULIP G52918 E3999387 1082700252918920230717E3999387 IGTCFS-65-1-UNI-TULIP G52918 E4000424 1082700252918920230722E4000424 IGTCFS-65-1-UNI-TULIP G52918 E4001803 1082700252918920230728E4001803 IGTCFS-65-1-FEM-TULIP G52917 E3994068 1082700252917220230701E3994068 IGTCFS-65-1-FEM-TULIP G52917 E3994988 1082700252917220230702E3994988 IGTCFS-65-1-FEM-TULIP G52917 E3995874 1082700252917220230706E3995874 IGTCFS-65-1-FEM-TULIP G52917 E3996260 1082700252917220230707E3996260 IGTCFS-65-1-FEM-TULIP G52917 E3996919 1082700252917220230709E3996919 IGTCFS-65-1-FEM-TULIP G52917 E3996920 1082700252917220230709E3996920 IGTCFS-65-1-FEM-TULIP G52917 E3997795 1082700252917220230713E3997795 IGTCFS-65-1-FEM-TULIP G52917 E3998212 1082700252917220230714E3998212 IGTCFS-65-1-FEM-TULIP G52917 E3998792 1082700252917220230716E3998792 IGTCFS-65-1-FEM-TULIP G52917 E3999578 1082700252917220230720E3999578 IGTCFS-65-1-FEM-TULIP G52917 E4000415 1082700252917220230722E4000415 IGTCFS-65-1-FEM-TULIP G52917 E4000736 1082700252917220230723E4000736 IGTCFS-65-1-FEM-TULIP G52917 E4001578 1082700252917220230727E4001578 IGTCFS-65-1-FEM-TULIP G52917 E4001882 1082700252917220230728E4001882 IGTCFS-65-1-FEM-TULIP G52917 E4002426 1082700252917220230729E4002426 IGTCFS-65-1-JUG-TULIP G52916 E3994495 1082700252916520230701E3994495 IGTCFS-65-1-JUG-TULIP G52916 E3995237 1082700252916520230703E3995237 IGTCFS-65-1-JUG-TULIP G52916 E3996679 1082700252916520230708E3996679 IGTCFS-65-1-JUG-TULIP G52916 E3997448 1082700252916520230710E3997448 IGTCFS-65-1-JUG-TULIP G52916 E3997853 1082700252916520230713E3997853 IGTCFS-65-1-JUG-TULIP G52916 E3998567 1082700252916520230715E3998567 IGTCFS-65-1-JUG-TULIP G52916 E3999386 1082700252916520230717E3999386 IGTCFS-65-1-JUG-TULIP G52916 E3999698 1082700252916520230720E3999698 IGTCFS-65-1-JUG-TULIP G52916 E4000702 1082700252916520230723E4000702 IGTCFS-65-1-JUG-TULIP G52916 E4001807 1082700252916520230728E4001807 IGTCFS-65-1-JUG-TULIP G52916 E4002414 1082700252916520230729E4002414 IGTCFS-65-1-JUG-TULIP G52916 E4002823 1082700252916520230730E4002823 IGTCFS-65-1-JUG-TULIP G52916 E4002824 1082700252916520230730E4002824 IGTCFS-65-1-UNI-TULIP G52918 E4004146 1082700252918920230804E4004146 IGTCFS-65-1-UNI-TULIP G52918 E4004875 1082700252918920230806E4004875 IGTCFS-65-1-UNI-TULIP G52918 E4007029 1082700252918920230813E4007029 IGTCFS-65-1-UNI-TULIP G52918 E4009099 1082700252918920230819E4009099 IGTCFS-65-1-UNI-TULIP G52918 E4012699 1082700252918920230828E4012699 IGTCFS-65-1-FEM-TULIP G52917 E4003840 1082700252917220230803E4003840 IGTCFS-65-1-FEM-TULIP G52917 E4004145 1082700252917220230804E4004145 IGTCFS-65-1-FEM-TULIP G52917 E4004874 1082700252917220230806E4004874 IGTCFS-65-1-FEM-TULIP G52917 E4005695 1082700252917220230807E4005695 IGTCFS-65-1-FEM-TULIP G52917 E4006000 1082700252917220230810E4006000 IGTCFS-65-1-FEM-TULIP G52917 E4006295 1082700252917220230811E4006295 IGTCFS-65-1-FEM-TULIP G52917 E4007830 1082700252917220230814E4007830 IGTCFS-65-1-FEM-TULIP G52917 E4009090 1082700252917220230819E4009090 IGTCFS-65-1-FEM-TULIP G52917 E4010279 1082700252917220230821E4010279 IGTCFS-65-1-FEM-TULIP G52917 E4011893 1082700252917220230827E4011893 IGTCFS-65-1-FEM-TULIP G52917 E4012698 1082700252917220230828E4012698 IGTCFS-65-1-FEM-TULIP G52917 E4013088 1082700252917220230831E4013088 IGTCFS-65-1-JUG-TULIP G52916 E4004505 1082700252916520230805E4004505 IGTCFS-65-1-JUG-TULIP G52916 E4004506 1082700252916520230805E4004506 IGTCFS-65-1-JUG-TULIP G52916 E4005999 1082700252916520230810E4005999 IGTCFS-65-1-JUG-TULIP G52916 E4006342 1082700252916520230811E4006342 IGTCFS-65-1-JUG-TULIP G52916 E4006640 1082700252916520230812E4006640 IGTCFS-65-1-JUG-TULIP G52916 E4007030 1082700252916520230813E4007030 IGTCFS-65-1-JUG-TULIP G52916 E4008595 1082700252916520230818E4008595 IGTCFS-65-1-JUG-TULIP G52916 E4009463 1082700252916520230820E4009463 IGTCFS-65-1-JUG-TULIP G52916 E4010324 1082700252916520230821E4010324 IGTCFS-65-1-JUG-TULIP G52916 E4011894 1082700252916520230827E4011894 IGTCFS-65-1-JUG-TULIP G52916 E4013141 1082700252916520230831E4013141 IGTCFS-65-1-UNI-TULIP G52918 E4016464 (01)10827002529189(17)230909(10)E4016464 IGTCFS-65-1-UNI-TULIP G52918 E4013922 1082700252918920230902E4013922 IGTCFS-65-1-UNI-TULIP G52918 E4015626 1082700252918920230907E4015626 IGTCFS-65-1-UNI-TULIP G52918 E4018577 1082700252918920230915E4018577 IGTCFS-65-1-UNI-TULIP G52918 E4021570 1082700252918920230923E4021570 IGTCFS-65-1-UNI-TULIP G52918 E4023944 1082700252918920230928E4023944 IGTCFS-65-1-UNI-TULIP G52918 E4024432 1082700252918920230929E4024432 IGTCFS-65-1-FEM-TULIP G52917 E4013550 1082700252917220230901E4013550 IGTCFS-65-1-FEM-TULIP G52917 E4013907 1082700252917220230902E4013907 IGTCFS-65-1-FEM-TULIP G52917 E4014268 1082700252917220230903E4014268 IGTCFS-65-1-FEM-TULIP G52917 E4015136 1082700252917220230904E4015136 IGTCFS-65-1-FEM-TULIP G52917 E4015566 1082700252917220230907E4015566 IGTCFS-65-1-FEM-TULIP G52917 E4016090 1082700252917220230908E4016090 IGTCFS-65-1-FEM-TULIP G52917 E4016699 1082700252917220230910E4016699 IGTCFS-65-1-FEM-TULIP G52917 E4017577 1082700252917220230911E4017577 IGTCFS-65-1-FEM-TULIP G52917 E4018246 1082700252917220230914E4018246 IGTCFS-65-1-FEM-TULIP G52917 E4018576 1082700252917220230915E4018576 IGTCFS-65-1-FEM-TULIP G52917 E4020757 1082700252917220230921E4020757 IGTCFS-65-1-FEM-TULIP G52917 E4020783 1082700252917220230921E4020783 IGTCFS-65-1-FEM-TULIP G52917 E4021131 1082700252917220230922E4021131 IGTCFS-65-1-FEM-TULIP G52917 E4021552 1082700252917220230923E4021552 IGTCFS-65-1-FEM-TULIP G52917 E4022087 1082700252917220230924E4022087 IGTCFS-65-1-FEM-TULIP G52917 E4024377 1082700252917220230929E4024377 IGTCFS-65-1-FEM-TULIP G52917 E4024904 1082700252917220230930E4024904 IGTCFS-65-1-JUG-TULIP G52916 E4013551 1082700252916520230901E4013551 IGTCFS-65-1-JUG-TULIP G52916 E4015137 1082700252916520230904E4015137 IGTCFS-65-1-JUG-TULIP G52916 E4016121 1082700252916520230908E4016121 IGTCFS-65-1-JUG-TULIP G52916 E4016379 1082700252916520230909E4016379 IGTCFS-65-1-JUG-TULIP G52916 E4016743 1082700252916520230910E4016743 IGTCFS-65-1-JUG-TULIP G52916 E4017578 1082700252916520230911E4017578 IGTCFS-65-1-JUG-TULIP G52916 E4019275 1082700252916520230917E4019275 IGTCFS-65-1-JUG-TULIP G52916 E4019276 1082700252916520230917E4019276 IGTCFS-65-1-JUG-TULIP G52916 E4021132 1082700252916520230922E4021132 IGTCFS-65-1-JUG-TULIP G52916 E4021553 1082700252916520230923E4021553 IGTCFS-65-1-JUG-TULIP G52916 E4021995 1082700252916520230924E4021995 IGTCFS-65-1-JUG-TULIP G52916 E4022855 1082700252916520230925E4022855 IGTCFS-65-1-JUG-TULIP G52916 E4024447 1082700252916520230929E4024447 IGTCFS-65-1-JUG-TULIP G52916 E4024903 1082700252916520230930E4024903 IGTCFS-65-1-UNI-TULIP G52918 E4026670 1082700252918920231005E4026670 IGTCFS-65-1-UNI-TULIP G52918 E4027587 1082700252918920231007E4027587 IGTCFS-65-1-UNI-TULIP G52918 E4027968 1082700252918920231008E4027968 IGTCFS-65-1-UNI-TULIP G52918 E4029584 1082700252918920231013E4029584 IGTCFS-65-1-UNI-TULIP G52918 E4031558 1082700252918920231019E4031558 IGTCFS-65-1-UNI-TULIP G52918 E4032664 1082700252918920231022E4032664 IGTCFS-65-1-UNI-TULIP G52918 E4034123 1082700252918920231026E4034123 IGTCFS-65-1-UNI-TULIP G52918 E4035366 1082700252918920231029E4035366 IGTCFS-65-1-FEM-TULIP G52917 E4025396 1082700252917220231001E4025396 IGTCFS-65-1-FEM-TULIP G52917 E4025788 1082700252917220231002E4025788 IGTCFS-65-1-FEM-TULIP G52917 E4027254 1082700252917220231006E4027254 IGTCFS-65-1-FEM-TULIP G52917 E4027937 1082700252917220231008E4027937 IGTCFS-65-1-FEM-TULIP G52917 E4028740 1082700252917220231009E4028740 IGTCFS-65-1-FEM-TULIP G52917 E4030319 1082700252917220231015E4030319 IGTCFS-65-1-FEM-TULIP G52917 E4030635 1082700252917220231016E4030635 IGTCFS-65-1-FEM-TULIP G52917 E4031513 1082700252917220231019E4031513 IGTCFS-65-1-FEM-TULIP G52917 E4032321 1082700252917220231021E4032321 IGTCFS-65-1-FEM-TULIP G52917 E4034124 1082700252917220231027E4034124 IGTCFS-65-1-FEM-TULIP G52917 E4035396 1082700252917220231029E4035396 IGTCFS-65-1-FEM-TULIP G52917 E4035978 1082700252917220231030E4035978 IGTCFS-65-1-JUG-TULIP G52916 E4025323 1082700252916520231001E4025323 IGTCFS-65-1-JUG-TULIP G52916 E4025335 1082700252916520231001E4025335 IGTCFS-65-1-JUG-TULIP G52916 E4027253 1082700252916520231006E4027253 IGTCFS-65-1-JUG-TULIP G52916 E4027652 1082700252916520231007E4027652 IGTCFS-65-1-JUG-TULIP G52916 E4029237 1082700252916520231012E4029237 IGTCFS-65-1-JUG-TULIP G52916 E4029550 1082700252916520231013E4029550 IGTCFS-65-1-JUG-TULIP G52916 E4030257 1082700252916520231015E4030257 IGTCFS-65-1-JUG-TULIP G52916 E4030634 1082700252916520231016E4030634 IGTCFS-65-1-JUG-TULIP G52916 E4031565 1082700252916520231019E4031565 IGTCFS-65-1-JUG-TULIP G52916 E4031927 1082700252916520231020E4031927 IGTCFS-65-1-JUG-TULIP G52916 E4032320 1082700252916520231021E4032320 IGTCFS-65-1-JUG-TULIP G52916 E4032663 1082700252916520231022E4032663 IGTCFS-65-1-JUG-TULIP G52916 E4034601 1082700252916520231027E4034601 IGTCFS-65-1-JUG-TULIP G52916 E4035061 1082700252916520231028E4035061 IGTCFS-65-1-UNI-TULIP G52918 E4044548 (01)10827002529189(17)232311(10)E4044548 IGTCFS-65-1-UNI-TULIP G52918 E4036937 1082700252918920231103E4036937 IGTCFS-65-1-UNI-TULIP G52918 E4037748 1082700252918920231105E4037748 IGTCFS-65-1-UNI-TULIP G52918 E4040438 1082700252918920231112E4040438 IGTCFS-65-1-UNI-TULIP G52918 E4042212 1082700252918920231117E4042212 IGTCFS-65-1-UNI-TULIP G52918 E4045488 1082700252918920231125E4045488 IGTCFS-65-1-UNI-TULIP G52918 E4047358 1082700252918920231130E4047358 IGTCFS-65-1-FEM-TULIP G52917 E4036592 1082700252917220231102E4036592 IGTCFS-65-1-FEM-TULIP G52917 E4037002 1082700252917220231103E4037002 IGTCFS-65-1-FEM-TULIP G52917 E4037775 1082700252917220231105E4037775 IGTCFS-65-1-FEM-TULIP G52917 E4038172 1082700252917220231106E4038172 IGTCFS-65-1-FEM-TULIP G52917 E4039931 1082700252917220231111E4039931 IGTCFS-65-1-FEM-TULIP G52917 E4040752 1082700252917220231113E4040752 IGTCFS-65-1-FEM-TULIP G52917 E4041736 1082700252917220231116E4041736 IGTCFS-65-1-FEM-TULIP G52917 E4043093 1082700252917220231119E4043093 IGTCFS-65-1-FEM-TULIP G52917 E4043588 1082700252917220231120E4043588 IGTCFS-65-1-FEM-TULIP G52917 E4045489 1082700252917220231125E4045489 IGTCFS-65-1-FEM-TULIP G52917 E4046012 1082700252917220231126E4046012 IGTCFS-65-1-FEM-TULIP G52917 E4046276 1082700252917220231127E4046276 IGTCFS-65-1-JUG-TULIP G52916 E4036544 1082700252916520231102E4036544 IGTCFS-65-1-JUG-TULIP G52916 E4036593 1082700252916520231102E4036593 IGTCFS-65-1-JUG-TULIP G52916 E4037323 1082700252916520231104E4037323 IGTCFS-65-1-JUG-TULIP G52916 E4037324 1082700252916520231104E4037324 IGTCFS-65-1-JUG-TULIP G52916 E4039128 1082700252916520231109E4039128 IGTCFS-65-1-JUG-TULIP G52916 E4039520 1082700252916520231110E4039520 IGTCFS-65-1-JUG-TULIP G52916 E4039621 1082700252916520231110E4039621 IGTCFS-65-1-JUG-TULIP G52916 E4041661 1082700252916520231116E4041661 IGTCFS-65-1-JUG-TULIP G52916 E4042197 1082700252916520231117E4042197 IGTCFS-65-1-JUG-TULIP G52916 E4042622 1082700252916520231118E4042622 IGTCFS-65-1-JUG-TULIP G52916 E4044600 1082700252916520231123E4044600 IGTCFS-65-1-JUG-TULIP G52916 E4045041 1082700252916520231124E4045041 IGTCFS-65-1-JUG-TULIP G52916 E4045042 1082700252916520231124E4045042 IGTCFS-65-1-JUG-TULIP G52916 E4045779 1082700252916520231126E4045779 IGTCFS-65-1-JUG-TULIP G52916 E4047409 1082700252916520231130E4047409 IGTCFS-65-1-UNI-TULIP G52918 E4049889 1082700252918920231207E4049889 IGTCFS-65-1-UNI-TULIP G52918 E4051485 1082700252918920231211E4051485 IGTCFS-65-1-FEM-TULIP G52917 E4047823 1082700252917220231201E4047823 IGTCFS-65-1-FEM-TULIP G52917 E4048702 1082700252917220231203E4048702 IGTCFS-65-1-FEM-TULIP G52917 E4048969 1082700252917220231204E4048969 IGTCFS-65-1-FEM-TULIP G52917 E4049876 1082700252917220231207E4049876 IGTCFS-65-1-FEM-TULIP G52917 E4050199 1082700252917220231208E4050199 IGTCFS-65-1-FEM-TULIP G52917 E4050789 1082700252917220231209E4050789 IGTCFS-65-1-FEM-TULIP G52917 E4051207 1082700252917220231210E4051207 IGTCFS-65-1-FEM-TULIP G52917 E4053356 1082700252917220231216E4053356 IGTCFS-65-1-FEM-TULIP G52917 E4053368 1082700252917220231216E4053368 IGTCFS-65-1-JUG-TULIP G52916 E4047676 1082700252916520231201E4047676 IGTCFS-65-1-JUG-TULIP G52916 E4048191 1082700252916520231202E4048191 IGTCFS-65-1-JUG-TULIP G52916 E4048761 1082700252916520231203E4048761 IGTCFS-65-1-JUG-TULIP G52916 E4050224 1082700252916520231208E4050224 IGTCFS-65-1-JUG-TULIP G52916 E4050880 1082700252916520231209E4050880 IGTCFS-65-1-JUG-TULIP G52916 E4051206 1082700252916520231210E4051206 IGTCFS-65-1-JUG-TULIP G52916 E4052379 1082700252916520231214E4052379 IGTCFS-65-1-JUG-TULIP G52916 E4052915 1082700252916520231215E4052915 IGTCFS-65-1-JUG-TULIP G52916 E4053751 1082700252916520231217E4053751 IGTCFS-65-1-UNI-TULIP G52918 E4060787 1082700252918920240113E4060787 IGTCFS-65-1-UNI-TULIP G52918 E4062614 1082700252918920240118E4062614 IGTCFS-65-1-UNI-TULIP G52918 E4063585 1082700252918920240120E4063585 IGTCFS-65-1-UNI-TULIP G52918 E4066849 1082700252918920240128E4066849 IGTCFS-65-1-FEM-TULIP G52917 E4060196 (01)10827002529172(17)241201(10)E4060196 IGTCFS-65-1-FEM-TULIP G52917 E4056849 1082700252917220240104E4056849 IGTCFS-65-1-FEM-TULIP G52917 E4057306 1082700252917220240105E4057306 IGTCFS-65-1-FEM-TULIP G52917 E4058325 1082700252917220240107E4058325 IGTCFS-65-1-FEM-TULIP G52917 E4058617 1082700252917220240108E4058617 IGTCFS-65-1-FEM-TULIP G52917 E4062652 1082700252917220240118E4062652 IGTCFS-65-1-FEM-TULIP G52917 E4063097 1082700252917220240119E4063097 IGTCFS-65-1-FEM-TULIP G52917 E4064043 1082700252917220240121E4064043 IGTCFS-65-1-FEM-TULIP G52917 E4064576 1082700252917220240122E4064576 IGTCFS-65-1-FEM-TULIP G52917 E4065975 1082700252917220240126E4065975 IGTCFS-65-1-FEM-TULIP G52917 E4066396 1082700252917220240127E4066396 IGTCFS-65-1-FEM-TULIP G52917 E4066817 1082700252917220240128E4066817 IGTCFS-65-1-JUG-TULIP G52916 E4057807 1082700252916520240106E4057807 IGTCFS-65-1-JUG-TULIP G52916 E4057808 1082700252916520240106E4057808 IGTCFS-65-1-JUG-TULIP G52916 E4059720 1082700252916520240111E4059720 IGTCFS-65-1-JUG-TULIP G52916 E4060803 1082700252916520240113E4060803 IGTCFS-65-1-JUG-TULIP G52916 E4062653 1082700252916520240118E4062653 IGTCFS-65-1-JUG-TULIP G52916 E4063081 1082700252916520240119E4063081 IGTCFS-65-1-JUG-TULIP G52916 E4063539 1082700252916520240120E4063539 IGTCFS-65-1-JUG-TULIP G52916 E4064044 1082700252916520240121E4064044 IGTCFS-65-1-JUG-TULIP G52916 E4065292 1082700252916520240125E4065292 IGTCFS-65-1-JUG-TULIP G52916 E4065909 1082700252916520240126E4065909 IGTCFS-65-1-JUG-TULIP G52916 E4066850 1082700252916520240128E4066850 IGTCFS-65-1-UNI-TULIP G52918 E4068586 (01)10827002529189(17)240202(10)E4068586 IGTCFS-65-1-UNI-TULIP G52918 E4068147 1082700252918920240201E4068147 IGTCFS-65-1-UNI-TULIP G52918 E4070944 1082700252918920240208E4070944 IGTCFS-65-1-UNI-TULIP G52918 E4073741 1082700252918920240215E4073741 IGTCFS-65-1-UNI-TULIP G52918 E4075994 1082700252918920240219E4075994 IGTCFS-65-1-UNI-TULIP G52918 E4076680 1082700252918920240222E4076680 IGTCFS-65-1-UNI-TULIP G52918 E4078705 1082700252918920240226E4078705 IGTCFS-65-1-FEM-TULIP G52917 E4073904 (01)10827002529172(17)241502(10)E4073904 IGTCFS-65-1-FEM-TULIP G52917 E4069103 1082700252917220240203E4069103 IGTCFS-65-1-FEM-TULIP G52917 E4069603 1082700252917220240204E4069603 IGTCFS-65-1-FEM-TULIP G52917 E4070993 1082700252917220240208E4070993 IGTCFS-65-1-FEM-TULIP G52917 E4071558 1082700252917220240209E4071558 IGTCFS-65-1-FEM-TULIP G52917 E4072067 1082700252917220240210E4072067 IGTCFS-65-1-FEM-TULIP G52917 E4072545 1082700252917220240211E4072545 IGTCFS-65-1-FEM-TULIP G52917 E4072881 1082700252917220240212E4072881 IGTCFS-65-1-FEM-TULIP G52917 E4074369 1082700252917220240216E4074369 IGTCFS-65-1-FEM-TULIP G52917 E4075175 1082700252917220240218E4075175 IGTCFS-65-1-FEM-TULIP G52917 E4075992 1082700252917220240219E4075992 IGTCFS-65-1-FEM-TULIP G52917 E4076679 1082700252917220240222E4076679 IGTCFS-65-1-FEM-TULIP G52917 E4077070 1082700252917220240223E4077070 IGTCFS-65-1-FEM-TULIP G52917 E4077444 1082700252917220240224E4077444 IGTCFS-65-1-JUG-TULIP G52916 E4068092 1082700252916520240201E4068092 IGTCFS-65-1-JUG-TULIP G52916 E4068608 1082700252916520240202E4068608 IGTCFS-65-1-JUG-TULIP G52916 E4069637 1082700252916520240205E4069637 IGTCFS-65-1-JUG-TULIP G52916 E4071470 1082700252916520240209E4071470 IGTCFS-65-1-JUG-TULIP G52916 E4072592 1082700252916520240211E4072592 IGTCFS-65-1-JUG-TULIP G52916 E4074187 1082700252916520240216E4074187 IGTCFS-65-1-JUG-TULIP G52916 E4074380 1082700252916520240216E4074380 IGTCFS-65-1-JUG-TULIP G52916 E4074819 1082700252916520240217E4074819 IGTCFS-65-1-JUG-TULIP G52916 E4075993 1082700252916520240219E4075993 IGTCFS-65-1-JUG-TULIP G52916 E4076702 1082700252916520240222E4076702 IGTCFS-65-1-JUG-TULIP G52916 E4077012 1082700252916520240223E4077012 IGTCFS-65-1-JUG-TULIP G52916 E4077013 1082700252916520240223E4077013 IGTCFS-65-1-UNI-TULIP G52918 E4080170 1082700252918920240302E4080170 IGTCFS-65-1-UNI-TULIP G52918 E4080796 1082700252918920240303E4080796 IGTCFS-65-1-UNI-TULIP G52918 E4083345 1082700252918920240309E4083345 IGTCFS-65-1-UNI-TULIP G52918 E4084357 1082700252918920240311E4084357 IGTCFS-65-1-UNI-TULIP G52918 E4084358 1082700252918920240311E4084358 IGTCFS-65-1-FEM-TULIP G52917 E4083162 10827002529172 IGTCFS-65-1-FEM-TULIP G52917 E4080706 1082700252917220240303E4080706 IGTCFS-65-1-FEM-TULIP G52917 E4080707 1082700252917220240303E4080707 IGTCFS-65-1-FEM-TULIP G52917 E4080795 1082700252917220240303E4080795 IGTCFS-65-1-FEM-TULIP G52917 E4082825 1082700252917220240308E4082825 IGTCFS-65-1-FEM-TULIP G52917 E4083161 1082700252917220240309E4083161 IGTCFS-65-1-FEM-TULIP G52917 E4083344 1082700252917220240309E4083344 IGTCFS-65-1-FEM-TULIP G52917 E4083692 1082700252917220240310E4083692 IGTCFS-65-1-FEM-TULIP G52917 E4083899 1082700252917220240310E4083899 IGTCFS-65-1-FEM-TULIP G52917 E4084497 1082700252917220240311E4084497 IGTCFS-65-1-FEM-TULIP G52917 E4084498 1082700252917220240311E4084498 IGTCFS-65-1-JUG-TULIP G52916 E4080809 (01)10827002529165(17)240303(10)E4080809 IGTCFS-65-1-JUG-TULIP G52916 E4080130 1082700252916520240302E4080130 IGTCFS-65-1-JUG-TULIP G52916 E4080131 1082700252916520240302E4080131 IGTCFS-65-1-JUG-TULIP G52916 E4082907 1082700252916520240308E4082907 IGTCFS-65-1-JUG-TULIP G52916 E4083313 1082700252916520240309E4083313 IGTCFS-65-1-JUG-TULIP G52916 E4083314 1082700252916520240309E4083314 IGTCFS-65-1-JUG-TULIP G52916 E4083356 1082700252916520240309E4083356 IGTCFS-65-1-JUG-TULIP G52916 E4083909 1082700252916520240310E4083909 IGTCFS-65-1-JUG-TULIP G52916 E4083965 1082700252916520240310E4083965 IGTCFS-65-1-JUG-TULIP G52916 E4084484 1082700252916520240311E4084484 IGTCFS-65-1-JUG-TULIP G52916 E4084485 1082700252916520240311E4084485 IGTCFS-65-1-UNI-TULIP G52918 E4098450 1082700252918920240415E4098450 IGTCFS-65-1-UNI-TULIP G52918 E4098986 1082700252918920240416E4098986 IGTCFS-65-1-UNI-TULIP G52918 E4100348 1082700252918920240420E4100348 IGTCFS-65-1-UNI-TULIP G52918 E4101335 1082700252918920240422E4101335 IGTCFS-65-1-UNI-TULIP G52918 E4103711 1082700252918920240428E4103711 IGTCFS-65-1-FEM-TULIP G52917 E4099374 1082700252917220240416E4099374 IGTCFS-65-1-FEM-TULIP G52917 E4101114 1082700252917220240422E4101114 IGTCFS-65-1-FEM-TULIP G52917 E4101115 1082700252917220240422E4101115 IGTCFS-65-1-FEM-TULIP G52917 E4103335 1082700252917220240427E4103335 IGTCFS-65-1-FEM-TULIP G52917 E4105216 1082700252917220240429E4105216 IGTCFS-65-1-JUG-TULIP G52916 E4098842 1082700252916520240416E4098842 IGTCFS-65-1-JUG-TULIP G52916 E4099952 1082700252916520240419E4099952 IGTCFS-65-1-JUG-TULIP G52916 E4100705 1082700252916520240421E4100705 IGTCFS-65-1-JUG-TULIP G52916 E4100706 1082700252916520240421E4100706 IGTCFS-65-1-JUG-TULIP G52916 E4100807 1082700252916520240421E4100807 IGTCFS-65-1-JUG-TULIP G52916 E4102264 1082700252916520240423E4102264 IGTCFS-65-1-JUG-TULIP G52916 E4102795 1082700252916520240426E4102795 IGTCFS-65-1-JUG-TULIP G52916 E4105215 1082700252916520240429E4105215 IGTCFS-65-1-UNI-TULIP G52918 E4110991 (01)10827002529189(17)241705(10)E4110991 IGTCFS-65-1-UNI-TULIP G52918 E4105628 1082700252918920240503E4105628 IGTCFS-65-1-UNI-TULIP G52918 E4110572 1082700252918920240517E4110572 IGTCFS-65-1-UNI-TULIP G52918 E4114326 1082700252918920240526E4114326 IGTCFS-65-1-UNI-TULIP G52918 E4116802 1082700252918920240531E4116802 IGTCFS-65-1-UNI-TULIP G52918 E4113371 (01)10827002529189(17)240521(10)E4113371 IGTCFS-65-1-UNI-TULIP G52918 E4115049 (01)10827002529189(17)240527(10)E4115049 IGTCFS-65-1-FEM-TULIP G52917 E4106971 (01)10827002529172(17)240605(10)E4106971 IGTCFS-65-1-FEM-TULIP G52917 E4115050 (01)10827002529172(17)242705(10)E4115050 IGTCFS-65-1-FEM-TULIP G52917 E4105600 1082700252917220240503E4105600 IGTCFS-65-1-FEM-TULIP G52917 E4106085 1082700252917220240504E4106085 IGTCFS-65-1-FEM-TULIP G52917 E4106546 1082700252917220240505E4106546 IGTCFS-65-1-FEM-TULIP G52917 E4106547 1082700252917220240505E4106547 IGTCFS-65-1-FEM-TULIP G52917 E4106970 1082700252917220240506E4106970 IGTCFS-65-1-FEM-TULIP G52917 E4111437 1082700252917220240518E4111437 IGTCFS-65-1-FEM-TULIP G52917 E4112004 1082700252917220240519E4112004 IGTCFS-65-1-FEM-TULIP G52917 E4113889 1082700252917220240525E4113889 IGTCFS-65-1-FEM-TULIP G52917 E4114327 1082700252917220240526E4114327 IGTCFS-65-1-FEM-TULIP G52917 E4116856 1082700252917220240531E4116856 IGTCFS-65-1-FEM-TULIP G52917 E4113888 (01)10827002529172(17)240525(10)E4113888 IGTCFS-65-1-JUG-TULIP G52916 E4105627 1082700252916520240503E4105627 IGTCFS-65-1-JUG-TULIP G52916 E4106086 1082700252916520240504E4106086 IGTCFS-65-1-JUG-TULIP G52916 E4106561 1082700252916520240505E4106561 IGTCFS-65-1-JUG-TULIP G52916 E4106995 1082700252916520240506E4106995 IGTCFS-65-1-JUG-TULIP G52916 E4107640 1082700252916520240507E4107640 IGTCFS-65-1-JUG-TULIP G52916 E4107878 1082700252916520240507E4107878 IGTCFS-65-1-JUG-TULIP G52916 E4113372 1082700252916520240521E4113372 IGTCFS-65-1-JUG-TULIP G52916 E4114369 1082700252916520240526E4114369 IGTCFS-65-1-JUG-TULIP G52916 E4115048 (01)10827002529165(17)240527(10)E4115048 IGTCFS-65-1-JUG-TULIP G52916 E4115390 (01)10827002529165(17)240528(10)E4115390 IGTCFS-65-1-UNI-TULIP G52918 E4120863 (01)10827002529189(17)240806(10)E4120863 IGTCFS-65-1-UNI-TULIP G52918 E4117852 1082700252918920240602E4117852 IGTCFS-65-1-UNI-TULIP G52918 E4119454 1082700252918920240604E4119454 IGTCFS-65-1-UNI-TULIP G52918 E4121700 1082700252918920240609E4121700 IGTCFS-65-1-UNI-TULIP G52918 E4124411 1082700252918920240615E4124411 IGTCFS-65-1-UNI-TULIP G52918 E4127619 1082700252918920240622E4127619 IGTCFS-65-1-UNI-TULIP G52918 E4128555 1082700252918920240624E4128555 IGTCFS-65-1-UNI-TULIP G52918 E4130503 1082700252918920240629E4130503 IGTCFS-65-1-UNI-TULIP G52918 E4126071 (01)10827002529189(17)240618(10)E4126071 IGTCFS-65-1-FEM-TULIP G52917 E4117813 1082700252917220240602E4117813 IGTCFS-65-1-FEM-TULIP G52917 E4117814 1082700252917220240602E4117814 IGTCFS-65-1-FEM-TULIP G52917 E4118701 1082700252917220240603E4118701 IGTCFS-65-1-FEM-TULIP G52917 E4118702 1082700252917220240603E4118702 IGTCFS-65-1-FEM-TULIP G52917 E4118717 1082700252917220240603E4118717 IGTCFS-65-1-FEM-TULIP G52917 E4120267 1082700252917220240607E4120267 IGTCFS-65-1-FEM-TULIP G52917 E4120268 1082700252917220240607E4120268 IGTCFS-65-1-FEM-TULIP G52917 E4121578 1082700252917220240609E4121578 IGTCFS-65-1-FEM-TULIP G52917 E4121677 1082700252917220240609E4121677 IGTCFS-65-1-FEM-TULIP G52917 E4122428 1082700252917220240610E4122428 IGTCFS-65-1-FEM-TULIP G52917 E4124523 1082700252917220240615E4124523 IGTCFS-65-1-FEM-TULIP G52917 E4125233 1082700252917220240616E4125233 IGTCFS-65-1-FEM-TULIP G52917 E4125758 1082700252917220240617E4125758 IGTCFS-65-1-FEM-TULIP G52917 E4126721 1082700252917220240621E4126721 IGTCFS-65-1-FEM-TULIP G52917 E4127508 1082700252917220240622E4127508 IGTCFS-65-1-FEM-TULIP G52917 E4127509 1082700252917220240622E4127509 IGTCFS-65-1-FEM-TULIP G52917 E4128001 1082700252917220240623E4128001 IGTCFS-65-1-FEM-TULIP G52917 E4128002 1082700252917220240623E4128002 IGTCFS-65-1-FEM-TULIP G52917 E4128739 1082700252917220240624E4128739 IGTCFS-65-1-FEM-TULIP G52917 E4130315 1082700252917220240628E4130315 IGTCFS-65-1-FEM-TULIP G52917 E4131243 1082700252917220240630E4131243 IGTCFS-65-1-FEM-TULIP G52917 E4131244 1082700252917220240630E4131244 IGTCFS-65-1-FEM-TULIP G52917 E4122430 (01)10827002529172(17)240610(10)E4122430 IGTCFS-65-1-FEM-TULIP G52917 E4126722 (01)10827002529172(17)240621(10)E4126722 IGTCFS-65-1-JUG-TULIP G52916 E4117086 1082700252916520240601E4117086 IGTCFS-65-1-JUG-TULIP G52916 E4117851 1082700252916520240602E4117851 IGTCFS-65-1-JUG-TULIP G52916 E4118718 1082700252916520240603E4118718 IGTCFS-65-1-JUG-TULIP G52916 E4119040 1082700252916520240604E4119040 IGTCFS-65-1-JUG-TULIP G52916 E4120781 1082700252916520240608E4120781 IGTCFS-65-1-JUG-TULIP G52916 E4120782 1082700252916520240608E4120782 IGTCFS-65-1-JUG-TULIP G52916 E4121300 1082700252916520240609E4121300 IGTCFS-65-1-JUG-TULIP G52916 E4121678 1082700252916520240609E4121678 IGTCFS-65-1-JUG-TULIP G52916 E4122429 1082700252916520240610E4122429 IGTCFS-65-1-JUG-TULIP G52916 E4122656 1082700252916520240611E4122656 IGTCFS-65-1-JUG-TULIP G52916 E4124968 1082700252916520240616E4124968 IGTCFS-65-1-JUG-TULIP G52916 E4125785 1082700252916520240617E4125785 IGTCFS-65-1-JUG-TULIP G52916 E4127086 1082700252916520240621E4127086 IGTCFS-65-1-JUG-TULIP G52916 E4127527 1082700252916520240622E4127527 IGTCFS-65-1-JUG-TULIP G52916 E4128069 1082700252916520240623E4128069 IGTCFS-65-1-JUG-TULIP G52916 E4128196 1082700252916520240623E4128196 IGTCFS-65-1-JUG-TULIP G52916 E4128745 1082700252916520240624E4128745 IGTCFS-65-1-JUG-TULIP G52916 E4130547 1082700252916520240629E4130547 IGTCFS-65-1-JUG-TULIP G52916 E4131303 1082700252916520240630E4131303 IGTCFS-65-1-JUG-TULIP G52916 E4117196 (01)10827002529165(17)240601(10)E4117196 IGTCFS-65-1-JUG-TULIP G52916 E4117296 (01)10827002529165(17)240601(10)E4117296 IGTCFS-65-1-JUG-TULIP G52916 E4122628 (01)10827002529165(17)240611(10)E4122628 IGTCFS-65-1-JUG-TULIP G52916 E4125949 (01)10827002529165(17)240618(10)E4125949 IGTCFS-65-1-JUG-TULIP G52916 E4128731 (01)10827002529165(17)240625(10)E4128731 IGTCFS-65-1-UNI-TULIP G52918 E4136221 1082700252918920240712E4136221 IGTCFS-65-1-UNI-TULIP G52918 E4139253 1082700252918920240721E4139253 IGTCFS-65-1-UNI-TULIP G52918 E4140666 1082700252918920240726E4140666 IGTCFS-65-1-UNI-TULIP G52918 E4133809 (01)10827002529189(17)240706(10)E4133809 IGTCFS-65-1-FEM-TULIP G52917 E4131937 1082700252917220240701E4131937 IGTCFS-65-1-FEM-TULIP G52917 E4133153 1082700252917220240705E4133153 IGTCFS-65-1-FEM-TULIP G52917 E4134294 1082700252917220240707E4134294 IGTCFS-65-1-FEM-TULIP G52917 E4134700 1082700252917220240708E4134700 IGTCFS-65-1-FEM-TULIP G52917 E4136989 1082700252917220240714E4136989 IGTCFS-65-1-FEM-TULIP G52917 E4137368 1082700252917220240715E4137368 IGTCFS-65-1-FEM-TULIP G52917 E4138541 1082700252917220240719E4138541 IGTCFS-65-1-FEM-TULIP G52917 E4138922 1082700252917220240720E4138922 IGTCFS-65-1-FEM-TULIP G52917 E4139594 1082700252917220240722E4139594 IGTCFS-65-1-FEM-TULIP G52917 E4140913 1082700252917220240727E4140913 IGTCFS-65-1-FEM-TULIP G52917 E4141310 1082700252917220240728E4141310 IGTCFS-65-1-JUG-TULIP G52916 E4131748 1082700252916520240701E4131748 IGTCFS-65-1-JUG-TULIP G52916 E4132998 1082700252916520240705E4132998 IGTCFS-65-1-JUG-TULIP G52916 E4133830 1082700252916520240706E4133830 IGTCFS-65-1-JUG-TULIP G52916 E4134369 1082700252916520240707E4134369 IGTCFS-65-1-JUG-TULIP G52916 E4136523 1082700252916520240713E4136523 IGTCFS-65-1-JUG-TULIP G52916 E4136958 1082700252916520240714E4136958 IGTCFS-65-1-JUG-TULIP G52916 E4137325 1082700252916520240715E4137325 IGTCFS-65-1-JUG-TULIP G52916 E4139226 1082700252916520240721E4139226 IGTCFS-65-1-JUG-TULIP G52916 E4139516 1082700252916520240722E4139516 IGTCFS-65-1-JUG-TULIP G52916 E4141293 1082700252916520240728E4141293 IGTCFS-65-1-JUG-TULIP G52916 E4141672 1082700252916520240729E4141672 IGTCFS-65-1-UNI-TULIP G52918 E4144870 1082700252918920240809E4144870 IGTCFS-65-1-UNI-TULIP G52918 E4145680 1082700252918920240811E4145680 IGTCFS-65-1-UNI-TULIP G52918 E4147396 1082700252918920240816E4147396 IGTCFS-65-1-UNI-TULIP G52918 E4148743 1082700252918920240819E4148743 IGTCFS-65-1-UNI-TULIP G52918 E4150610 1082700252918920240824E4150610 IGTCFS-65-1-UNI-TULIP G52918 E4153466 1082700252918920240831E4153466 IGTCFS-65-1-FEM-TULIP G52917 E4144871 1082700252917220240809E4144871 IGTCFS-65-1-FEM-TULIP G52917 E4145008 1082700252917220240810E4145008 IGTCFS-65-1-FEM-TULIP G52917 E4145662 1082700252917220240811E4145662 IGTCFS-65-1-FEM-TULIP G52917 E4146150 1082700252917220240812E4146150 IGTCFS-65-1-FEM-TULIP G52917 E4147395 1082700252917220240816E4147395 IGTCFS-65-1-FEM-TULIP G52917 E4147678 1082700252917220240817E4147678 IGTCFS-65-1-FEM-TULIP G52917 E4148152 1082700252917220240818E4148152 IGTCFS-65-1-FEM-TULIP G52917 E4149255 1082700252917220240820E4149255 IGTCFS-65-1-FEM-TULIP G52917 E4149256 1082700252917220240820E4149256 IGTCFS-65-1-FEM-TULIP G52917 E4150423 1082700252917220240824E4150423 IGTCFS-65-1-FEM-TULIP G52917 E4150942 1082700252917220240825E4150942 IGTCFS-65-1-FEM-TULIP G52917 E4150943 1082700252917220240825E4150943 IGTCFS-65-1-FEM-TULIP G52917 E4151902 1082700252917220240827E4151902 IGTCFS-65-1-JUG-TULIP G52916 E4145007 1082700252916520240809E4145007 IGTCFS-65-1-JUG-TULIP G52916 E4145320 1082700252916520240810E4145320 IGTCFS-65-1-JUG-TULIP G52916 E4145679 1082700252916520240811E4145679 IGTCFS-65-1-JUG-TULIP G52916 E4146152 1082700252916520240812E4146152 IGTCFS-65-1-JUG-TULIP G52916 E4147356 1082700252916520240816E4147356 IGTCFS-65-1-JUG-TULIP G52916 E4147787 1082700252916520240817E4147787 IGTCFS-65-1-JUG-TULIP G52916 E4148255 1082700252916520240818E4148255 IGTCFS-65-1-JUG-TULIP G52916 E4149677 1082700252916520240820E4149677 IGTCFS-65-1-JUG-TULIP G52916 E4150100 1082700252916520240823E4150100 IGTCFS-65-1-JUG-TULIP G52916 E4150522 1082700252916520240824E4150522 IGTCFS-65-1-JUG-TULIP G52916 E4151901 1082700252916520240827E4151901 IGTCFS-65-1-UNI-TULIP G52918 E4153944 1082700252918920240901E4153944 IGTCFS-65-1-UNI-TULIP G52918 E4154288 1082700252918920240902E4154288 IGTCFS-65-1-UNI-TULIP G52918 E4157108 1082700252918920240909E4157108 IGTCFS-65-1-UNI-TULIP G52918 E4158084 1082700252918920240910E4158084 IGTCFS-65-1-UNI-TULIP G52918 E4162290 1082700252918920240921E4162290 IGTCFS-65-1-UNI-TULIP G52918 E4164234 1082700252918920240925E4164234 IGTCFS-65-1-UNI-TULIP G52918 E4165095 1082700252918920240928E4165095 IGTCFS-65-1-UNI-TULIP G52918 E4155779 (01)10827002529189(17)240906(10)E4155779 IGTCFS-65-1-FEM-TULIP G52917 E4153943 1082700252917220240901E4153943 IGTCFS-65-1-FEM-TULIP G52917 E4154227 1082700252917220240902E4154227 IGTCFS-65-1-FEM-TULIP G52917 E4154228 1082700252917220240902E4154228 IGTCFS-65-1-FEM-TULIP G52917 E4155350 1082700252917220240903E4155350 IGTCFS-65-1-FEM-TULIP G52917 E4156182 1082700252917220240907E4156182 IGTCFS-65-1-FEM-TULIP G52917 E4156546 1082700252917220240908E4156546 IGTCFS-65-1-FEM-TULIP G52917 E4156547 1082700252917220240908E4156547 IGTCFS-65-1-FEM-TULIP G52917 E4158200 1082700252917220240910E4158200 IGTCFS-65-1-FEM-TULIP G52917 E4159377 1082700252917220240915E4159377 IGTCFS-65-1-FEM-TULIP G52917 E4159448 1082700252917220240915E4159448 IGTCFS-65-1-FEM-TULIP G52917 E4159667 1082700252917220240915E4159667 IGTCFS-65-1-FEM-TULIP G52917 E4162203 1082700252917220240921E4162203 IGTCFS-65-1-FEM-TULIP G52917 E4162789 1082700252917220240922E4162789 IGTCFS-65-1-FEM-TULIP G52917 E4163142 1082700252917220240923E4163142 IGTCFS-65-1-FEM-TULIP G52917 E4164295 1082700252917220240925E4164295 IGTCFS-65-1-FEM-TULIP G52917 E4164586 1082700252917220240927E4164586 IGTCFS-65-1-FEM-TULIP G52917 E4164587 1082700252917220240927E4164587 IGTCFS-65-1-FEM-TULIP G52917 E4165579 1082700252917220240929E4165579 IGTCFS-65-1-FEM-TULIP G52917 E4165875 1082700252917220240930E4165875 IGTCFS-65-1-FEM-TULIP G52917 E4165909 1082700252917220240930E4165909 IGTCFS-65-1-FEM-TULIP G52917 E4165910 1082700252917220240930E4165910 IGTCFS-65-1-JUG-TULIP G52916 E4153887 1082700252916520240901E4153887 IGTCFS-65-1-JUG-TULIP G52916 E4153888 1082700252916520240901E4153888 IGTCFS-65-1-JUG-TULIP G52916 E4153927 1082700252916520240901E4153927 IGTCFS-65-1-JUG-TULIP G52916 E4154289 1082700252916520240902E4154289 IGTCFS-65-1-JUG-TULIP G52916 E4155351 1082700252916520240903E4155351 IGTCFS-65-1-JUG-TULIP G52916 E4156153 1082700252916520240907E4156153 IGTCFS-65-1-JUG-TULIP G52916 E4156553 1082700252916520240909E4156553 IGTCFS-65-1-JUG-TULIP G52916 E4157107 1082700252916520240909E4157107 IGTCFS-65-1-JUG-TULIP G52916 E4157109 1082700252916520240909E4157109 IGTCFS-65-1-JUG-TULIP G52916 E4158700 1082700252916520240913E4158700 IGTCFS-65-1-JUG-TULIP G52916 E4158925 1082700252916520240914E4158925 IGTCFS-65-1-JUG-TULIP G52916 E4159684 1082700252916520240915E4159684 IGTCFS-65-1-JUG-TULIP G52916 E4161457 1082700252916520240920E4161457 IGTCFS-65-1-JUG-TULIP G52916 E4161724 1082700252916520240920E4161724 IGTCFS-65-1-JUG-TULIP G52916 E4162710 1082700252916520240922E4162710 IGTCFS-65-1-JUG-TULIP G52916 E4163804 1082700252916520240924E4163804 IGTCFS-65-1-JUG-TULIP G52916 E4163924 1082700252916520240924E4163924 IGTCFS-65-1-JUG-TULIP G52916 E4165889 1082700252916520240930E4165889 IGTCFS-65-1-UNI-TULIP G52918 E4166147 1082700252918920241001E4166147 IGTCFS-65-1-UNI-TULIP G52918 E4167221 1082700252918920241004E4167221 IGTCFS-65-1-UNI-TULIP G52918 E4167480 1082700252918920241005E4167480 IGTCFS-65-1-FEM-TULIP G52917 E4166427 1082700252917220241001E4166427 IGTCFS-65-1-FEM-TULIP G52917 E4169769 1082700252917220241011E4169769 IGTCFS-65-1-FEM-TULIP G52917 E4169770 1082700252917220241011E4169770 IGTCFS-65-1-FEM-TULIP G52917 E4169875 1082700252917220241011E4169875 IGTCFS-65-1-FEM-TULIP G52917 E4169876 1082700252917220241011E4169876 IGTCFS-65-1-FEM-TULIP G52917 E4170079 1082700252917220241012E4170079 IGTCFS-65-1-FEM-TULIP G52917 E4170080 1082700252917220241012E4170080 IGTCFS-65-1-FEM-TULIP G52917 E4170255 1082700252917220241012E4170255 IGTCFS-65-1-FEM-TULIP G52917 E4170256 1082700252917220241012E4170256 IGTCFS-65-1-FEM-TULIP G52917 E4170738 1082700252917220241013E4170738 IGTCFS-65-1-FEM-TULIP G52917 E4170739 1082700252917220241013E4170739 IGTCFS-65-1-FEM-TULIP G52917 E4170789 1082700252917220241013E4170789 IGTCFS-65-1-FEM-TULIP G52917 E4170790 1082700252917220241013E4170790 IGTCFS-65-1-FEM-TULIP G52917 E4172851 1082700252917220241018E4172851 IGTCFS-65-1-FEM-TULIP G52917 E4172852 1082700252917220241018E4172852 IGTCFS-65-1-FEM-TULIP G52917 E4172901 1082700252917220241018E4172901 IGTCFS-65-1-FEM-TULIP G52917 E4172924 1082700252917220241018E4172924 IGTCFS-65-1-FEM-TULIP G52917 E4172925 1082700252917220241018E4172925 IGTCFS-65-1-FEM-TULIP G52917 E4172940 1082700252917220241018E4172940 IGTCFS-65-1-FEM-TULIP G52917 E4173345 1082700252917220241019E4173345 IGTCFS-65-1-FEM-TULIP G52917 E4173346 1082700252917220241019E4173346 IGTCFS-65-1-FEM-TULIP G52917 E4173357 1082700252917220241019E4173357 IGTCFS-65-1-FEM-TULIP G52917 E4173421 1082700252917220241019E4173421 IGTCFS-65-1-FEM-TULIP G52917 E4173423 1082700252917220241019E4173423 IGTCFS-65-1-FEM-TULIP G52917 E4174004 1082700252917220241020E4174004 IGTCFS-65-1-FEM-TULIP G52917 E4174005 1082700252917220241020E4174005 IGTCFS-65-1-FEM-TULIP G52917 E4174006 1082700252917220241020E4174006 IGTCFS-65-1-FEM-TULIP G52917 E4174056 1082700252917220241020E4174056 IGTCFS-65-1-FEM-TULIP G52917 E4174057 1082700252917220241020E4174057 IGTCFS-65-1-FEM-TULIP G52917 E4174058 1082700252917220241020E4174058 IGTCFS-65-1-FEM-TULIP G52917 E4174531 1082700252917220241021E4174531 IGTCFS-65-1-FEM-TULIP G52917 E4174532 1082700252917220241021E4174532 IGTCFS-65-1-FEM-TULIP G52917 E4174544 1082700252917220241021E4174544 IGTCFS-65-1-FEM-TULIP G52917 E4174572 1082700252917220241021E4174572 IGTCFS-65-1-FEM-TULIP G52917 E4174573 1082700252917220241021E4174573 IGTCFS-65-1-FEM-TULIP G52917 E4174574 1082700252917220241021E4174574 IGTCFS-65-1-FEM-TULIP G52917 E4174862 1082700252917220241022E4174862 IGTCFS-65-1-FEM-TULIP G52917 E4174867 1082700252917220241022E4174867 IGTCFS-65-1-FEM-TULIP G52917 E4174868 1082700252917220241022E4174868 IGTCFS-65-1-FEM-TULIP G52917 E4176173 1082700252917220241025E4176173 IGTCFS-65-1-FEM-TULIP G52917 E4176174 1082700252917220241025E4176174 IGTCFS-65-1-FEM-TULIP G52917 E4176274 1082700252917220241025E4176274 IGTCFS-65-1-FEM-TULIP G52917 E4176306 1082700252917220241025E4176306 IGTCFS-65-1-FEM-TULIP G52917 E4176307 1082700252917220241025E4176307 IGTCFS-65-1-FEM-TULIP G52917 E4176619 1082700252917220241026E4176619 IGTCFS-65-1-FEM-TULIP G52917 E4176620 1082700252917220241026E4176620 IGTCFS-65-1-FEM-TULIP G52917 E4176691 1082700252917220241026E4176691 IGTCFS-65-1-FEM-TULIP G52917 E4176692 1082700252917220241026E4176692 IGTCFS-65-1-FEM-TULIP G52917 E4176772 1082700252917220241026E4176772 IGTCFS-65-1-FEM-TULIP G52917 E4176773 1082700252917220241026E4176773 IGTCFS-65-1-FEM-TULIP G52917 E4177087 1082700252917220241027E4177087 IGTCFS-65-1-FEM-TULIP G52917 E4177088 1082700252917220241027E4177088 IGTCFS-65-1-FEM-TULIP G52917 E4177155 1082700252917220241027E4177155 IGTCFS-65-1-FEM-TULIP G52917 E4177176 1082700252917220241027E4177176 IGTCFS-65-1-FEM-TULIP G52917 E4177177 1082700252917220241027E4177177 IGTCFS-65-1-FEM-TULIP G52917 E4177206 1082700252917220241027E4177206 IGTCFS-65-1-FEM-TULIP G52917 E4177595 1082700252917220241028E4177595 IGTCFS-65-1-FEM-TULIP G52917 E4177596 1082700252917220241028E4177596 IGTCFS-65-1-FEM-TULIP G52917 E4177614 1082700252917220241028E4177614 IGTCFS-65-1-FEM-TULIP G52917 E4177615 1082700252917220241028E4177615 IGTCFS-65-1-FEM-TULIP G52917 E4177640 1082700252917220241028E4177640 IGTCFS-65-1-FEM-TULIP G52917 E4177649 1082700252917220241028E4177649 IGTCFS-65-1-FEM-TULIP G52917 E4177926 1082700252917220241029E4177926 IGTCFS-65-1-FEM-TULIP G52917 E4177927 1082700252917220241029E4177927 IGTCFS-65-1-FEM-TULIP G52917 E4178268 1082700252917220241029E4178268 IGTCFS-65-1-FEM-TULIP G52917 E4178749 1082700252917220241029E4178749 IGTCFS-65-1-FEM-TULIP G52917 E4173422 (01)10827002529172(17)241019(10)E4173422 IGTCFS-65-1-FEM-TULIP G52917 E4176305 (01)10827002529172(17)241025(10)E4176305 IGTCFS-65-1-JUG-TULIP G52916 E4166428 1082700252916520241001E4166428 IGTCFS-65-1-JUG-TULIP G52916 E4167115 1082700252916520241004E4167115 IGTCFS-65-1-JUG-TULIP G52916 E4167205 1082700252916520241004E4167205 IGTCFS-65-1-JUG-TULIP G52916 E4167222 1082700252916520241004E4167222 IGTCFS-65-1-JUG-TULIP G52916 E4167469 1082700252916520241005E4167469 IGTCFS-65-1-JUG-TULIP G52916 E4167470 1082700252916520241005E4167470 IGTCFS-65-1-JUG-TULIP G52916 E4167821 1082700252916520241006E4167821 IGTCFS-65-1-JUG-TULIP G52916 E4167822 1082700252916520241006E4167822 IGTCFS-65-1-UNI-TULIP G52918 E4182876 1082700252918920241109E4182876 IGTCFS-65-1-UNI-TULIP G52918 E4182877 1082700252918920241109E4182877 IGTCFS-65-1-UNI-TULIP G52918 E4184110 1082700252918920241111E4184110 IGTCFS-65-1-UNI-TULIP G52918 E4184894 1082700252918920241112E4184894 IGTCFS-65-1-UNI-TULIP G52918 E4185896 1082700252918920241115E4185896 IGTCFS-65-1-UNI-TULIP G52918 E4186441 1082700252918920241116E4186441 IGTCFS-65-1-UNI-TULIP G52918 E4187014 1082700252918920241117E4187014 IGTCFS-65-1-UNI-TULIP G52918 E4187491 1082700252918920241118E4187491 IGTCFS-65-1-UNI-TULIP G52918 E4187593 1082700252918920241118E4187593 IGTCFS-65-1-UNI-TULIP G52918 E4189981 1082700252918920241123E4189981 IGTCFS-65-1-UNI-TULIP G52918 E4190283 1082700252918920241124E4190283 IGTCFS-65-1-FEM-TULIP G52917 E4188517 1082700252917220241119E4188517 IGTCFS-65-1-FEM-TULIP G52917 E4188518 1082700252917220241119E4188518 IGTCFS-65-1-FEM-TULIP G52917 E4188714 1082700252917220241119E4188714 IGTCFS-65-1-FEM-TULIP G52917 E4188715 1082700252917220241119E4188715 IGTCFS-65-1-FEM-TULIP G52917 E4188732 1082700252917220241119E4188732 IGTCFS-65-1-FEM-TULIP G52917 E4189659 1082700252917220241122E4189659 IGTCFS-65-1-FEM-TULIP G52917 E4190441 1082700252917220241124E4190441 IGTCFS-65-1-FEM-TULIP G52917 E4190949 1082700252917220241125E4190949 IGTCFS-65-1-FEM-TULIP G52917 E4190957 1082700252917220241125E4190957 IGTCFS-65-1-FEM-TULIP G52917 E4191246 1082700252917220241125E4191246 IGTCFS-65-1-FEM-TULIP G52917 E4191127 1082700252917220241126E4191127 IGTCFS-65-1-FEM-TULIP G52917 E4191174 1082700252917220241126E4191174 IGTCFS-65-1-FEM-TULIP G52917 E4191394 1082700252917220241126E4191394 IGTCFS-65-1-FEM-TULIP G52917 E4192609 1082700252917220241129E4192609 IGTCFS-65-1-FEM-TULIP G52917 E4192720 1082700252917220241129E4192720 IGTCFS-65-1-FEM-TULIP G52917 E4193194 1082700252917220241130E4193194 IGTCFS-65-1-FEM-TULIP G52917 E4193195 1082700252917220241130E4193195 IGTCFS-65-1-FEM-TULIP G52917 E4193252 1082700252917220241130E4193252 IGTCFS-65-1-FEM-TULIP G52917 E4193273 1082700252917220241130E4193273 IGTCFS-65-1-FEM-TULIP G52917 E4193342 1082700252917220241130E4193342 IGTCFS-65-1-FEM-TULIP G52917 E4193396 1082700252917220241130E4193396 IGTCFS-65-1-FEM-TULIP G52917 E4192608 (01)10827002529172(17)241129(10)E4192608 IGTCFS-65-1-JUG-TULIP G52916 E4183629 1082700252916520241110E4183629 IGTCFS-65-1-JUG-TULIP G52916 E4184123 1082700252916520241111E4184123 IGTCFS-65-1-JUG-TULIP G52916 E4184893 1082700252916520241112E4184893 IGTCFS-65-1-JUG-TULIP G52916 E4185783 1082700252916520241115E4185783 IGTCFS-65-1-JUG-TULIP G52916 E4185895 1082700252916520241115E4185895 IGTCFS-65-1-JUG-TULIP G52916 E4185917 1082700252916520241115E4185917 IGTCFS-65-1-JUG-TULIP G52916 E4186176 1082700252916520241116E4186176 IGTCFS-65-1-JUG-TULIP G52916 E4186346 1082700252916520241116E4186346 IGTCFS-65-1-JUG-TULIP G52916 E4186475 1082700252916520241116E4186475 IGTCFS-65-1-JUG-TULIP G52916 E4186939 1082700252916520241117E4186939 IGTCFS-65-1-JUG-TULIP G52916 E4186992 1082700252916520241117E4186992 IGTCFS-65-1-JUG-TULIP G52916 E4187070 1082700252916520241117E4187070 IGTCFS-65-1-JUG-TULIP G52916 E4187584 1082700252916520241118E4187584 IGTCFS-65-1-JUG-TULIP G52916 E4187591 1082700252916520241118E4187591 IGTCFS-65-1-JUG-TULIP G52916 E4189839 1082700252916520241123E4189839 IGTCFS-65-1-JUG-TULIP G52916 E4189932 1082700252916520241123E4189932 IGTCFS-65-1-JUG-TULIP G52916 E4189948 1082700252916520241123E4189948 IGTCFS-65-1-JUG-TULIP G52916 E4190029 1082700252916520241123E4190029 IGTCFS-65-1-JUG-TULIP G52916 E4190388 1082700252916520241124E4190388 IGTCFS-65-1-JUG-TULIP G52916 E4187592 (01)10827002529165(17)241118(10)E4187592 IGTCFS-65-1-JUG-TULIP G52916 E4190480 (01)10827002529165(17)241124(10)E4190480 IGTCFS-65-1-UNI-TULIP G52918 E4195304 1082700252918920241203E4195304 IGTCFS-65-1-UNI-TULIP G52918 E4195305 1082700252918920241203E4195305 IGTCFS-65-1-UNI-TULIP G52918 E4197239 1082700252918920241209E4197239 IGTCFS-65-1-UNI-TULIP G52918 E4197346 1082700252918920241209E4197346 IGTCFS-65-1-FEM-TULIP G52917 E4194443 1082700252917220241202E4194443 IGTCFS-65-1-FEM-TULIP G52917 E4194859 1082700252917220241203E4194859 IGTCFS-65-1-FEM-TULIP G52917 E4195515 1082700252917220241206E4195515 IGTCFS-65-1-FEM-TULIP G52917 E4195658 1082700252917220241206E4195658 IGTCFS-65-1-FEM-TULIP G52917 E4195719 1082700252917220241206E4195719 IGTCFS-65-1-FEM-TULIP G52917 E4195766 1082700252917220241206E4195766 IGTCFS-65-1-FEM-TULIP G52917 E4195827 1082700252917220241206E4195827 IGTCFS-65-1-FEM-TULIP G52917 E4195999 1082700252917220241207E4195999 IGTCFS-65-1-FEM-TULIP G52917 E4196151 1082700252917220241207E4196151 IGTCFS-65-1-FEM-TULIP G52917 E4196152 1082700252917220241207E4196152 IGTCFS-65-1-FEM-TULIP G52917 E4196157 1082700252917220241207E4196157 IGTCFS-65-1-FEM-TULIP G52917 E4196231 1082700252917220241207E4196231 IGTCFS-65-1-FEM-TULIP G52917 E4196948 1082700252917220241208E4196948 IGTCFS-65-1-FEM-TULIP G52917 E4197158 1082700252917220241209E4197158 IGTCFS-65-1-FEM-TULIP G52917 E4197159 1082700252917220241209E4197159 IGTCFS-65-1-FEM-TULIP G52917 E4197356 1082700252917220241209E4197356 IGTCFS-65-1-FEM-TULIP G52917 E4198256 1082700252917220241210E4198256 IGTCFS-65-1-FEM-TULIP G52917 E4198257 1082700252917220241210E4198257 IGTCFS-65-1-FEM-TULIP G52917 E4198360 1082700252917220241210E4198360 IGTCFS-65-1-FEM-TULIP G52917 E4198507 1082700252917220241210E4198507 IGTCFS-65-1-FEM-TULIP G52917 E4198688 1082700252917220241213E4198688 IGTCFS-65-1-FEM-TULIP G52917 E4198689 1082700252917220241213E4198689 IGTCFS-65-1-FEM-TULIP G52917 E4198797 1082700252917220241213E4198797 IGTCFS-65-1-FEM-TULIP G52917 E4198925 1082700252917220241213E4198925 IGTCFS-65-1-FEM-TULIP G52917 E4199319 1082700252917220241214E4199319 IGTCFS-65-1-FEM-TULIP G52917 E4199335 1082700252917220241214E4199335 IGTCFS-65-1-FEM-TULIP G52917 E4199336 1082700252917220241214E4199336 IGTCFS-65-1-FEM-TULIP G52917 E4199365 1082700252917220241214E4199365 IGTCFS-65-1-FEM-TULIP G52917 E4199730 1082700252917220241215E4199730 IGTCFS-65-1-FEM-TULIP G52917 E4199844 1082700252917220241215E4199844 IGTCFS-65-1-FEM-TULIP G52917 E4199947 1082700252917220241215E4199947 IGTCFS-65-1-FEM-TULIP G52917 E4199948 1082700252917220241216E4199948 IGTCFS-65-1-FEM-TULIP G52917 E4200239 1082700252917220241216E4200239 IGTCFS-65-1-FEM-TULIP G52917 E4200240 1082700252917220241216E4200240 IGTCFS-65-1-FEM-TULIP G52917 E4200399 1082700252917220241216E4200399 IGTCFS-65-1-FEM-TULIP G52917 E4200400 1082700252917220241216E4200400 IGTCFS-65-1-FEM-TULIP G52917 E4200727 1082700252917220241217E4200727 IGTCFS-65-1-FEM-TULIP G52917 E4200728 1082700252917220241217E4200728 IGTCFS-65-1-FEM-TULIP G52917 E4200411 (01)10827002529172(17)241216(10)E4200411 IGTCFS-65-1-JUG-TULIP G52916 E4193593 1082700252916520241201E4193593 IGTCFS-65-1-JUG-TULIP G52916 E4194367 1082700252916520241202E4194367 IGTCFS-65-1-JUG-TULIP G52916 E4194464 1082700252916520241202E4194464 IGTCFS-65-1-JUG-TULIP G52916 E4196676 1082700252916520241208E4196676 IGTCFS-65-1-JUG-TULIP G52916 E4196692 1082700252916520241208E4196692 IGTCFS-65-1-JUG-TULIP G52916 E4196843 1082700252916520241208E4196843 IGTCFS-65-1-JUG-TULIP G52916 E4196958 1082700252916520241208E4196958 IGTCFS-65-1-JUG-TULIP G52916 E4193592 (01)10827002529165(17)241201(10)E4193592 IGTCFS-65-1-FEM-TULIP G52917 E4194331 1082700252917219000101E4194331 IGTCFS-65-1-UNI-TULIP G52918 E4207739 1082700252918920250110E4207739 IGTCFS-65-1-UNI-TULIP G52918 E4208225 1082700252918920250111E4208225 IGTCFS-65-1-UNI-TULIP G52918 E4209452 (01)10827002529189(17)250113(10)E4209452 IGTCFS-65-1-UNI-TULIP G52918 E4209453 (01)10827002529189(17)250113(10)E4209453 IGTCFS-65-1-FEM-TULIP G52917 E4208292 1082700252917220250111E4208292 IGTCFS-65-1-FEM-TULIP G52917 E4208293 1082700252917220250112E4208293 IGTCFS-65-1-FEM-TULIP G52917 E4208637 1082700252917220250112E4208637 IGTCFS-65-1-FEM-TULIP G52917 E4208638 1082700252917220250112E4208638 IGTCFS-65-1-FEM-TULIP G52917 E4208656 1082700252917220250112E4208656 IGTCFS-65-1-FEM-TULIP G52917 E4208761 (01)10827002529172(17)250112(10)E4208761 IGTCFS-65-1-FEM-TULIP G52917 E4208762 (01)10827002529172(17)250112(10)E4208762 IGTCFS-65-1-FEM-TULIP G52917 E4208791 (01)10827002529172(17)250112(10)E4208791 IGTCFS-65-1-FEM-TULIP G52917 E4209462 (01)10827002529172(17)250113(10)E4209462 IGTCFS-65-1-FEM-TULIP G52917 E4209481 (01)10827002529172(17)250113(10)E4209481 IGTCFS-65-1-FEM-TULIP G52917 E4209482 (01)10827002529172(17)250113(10)E4209482 IGTCFS-65-1-FEM-TULIP G52917 E4209499 (01)10827002529172(17)250113(10)E4209499 IGTCFS-65-1-FEM-TULIP G52917 E4210488 (01)10827002529172(17)250114(10)E4210488 IGTCFS-65-1-FEM-TULIP G52917 E4211273 (01)10827002529172(17)250117(10)E4211273 IGTCFS-65-1-FEM-TULIP G52917 E4211274 (01)10827002529172(17)250117(10)E4211274 IGTCFS-65-1-FEM-TULIP G52917 E4211682 (01)10827002529172(17)250118(10)E4211682 IGTCFS-65-1-FEM-TULIP G52917 E4211687 (01)10827002529172(17)250118(10)E4211687 IGTCFS-65-1-FEM-TULIP G52917 E4211688 (01)10827002529172(17)250118(10)E4211688 IGTCFS-65-1-FEM-TULIP G52917 E4211689 (01)10827002529172(17)250118(10)E4211689 IGTCFS-65-1-JUG-TULIP G52916 E4207722 1082700252916520250110E4207722 IGTCFS-65-1-JUG-TULIP G52916 E4207723 1082700252916520250110E4207723 IGTCFS-65-1-JUG-TULIP G52916 E4207740 1082700252916520250110E4207740 IGTCFS-65-1-JUG-TULIP G52916 E4207781 1082700252916520250110E4207781 IGTCFS-65-1-JUG-TULIP G52916 E4207810 1082700252916520250110E4207810 IGTCFS-65-1-JUG-TULIP G52916 E4208226 1082700252916520250111E4208226 IGTCFS-65-1-JUG-TULIP G52916 E4208266 1082700252916520250111E4208266 IGTCFS-65-1-JUG-TULIP G52916 E4208316 1082700252916520250111E4208316 IGTCFS-65-1-JUG-TULIP G52916 E4211177 (01)10827002529165(17)250117(10)E4211177 IGTCFS-65-1-JUG-TULIP G52916 E4211220 (01)10827002529165(17)250117(10)E4211220 IGTCFS-65-1-JUG-TULIP G52916 E4211254 (01)10827002529165(17)250117(10)E4211254

Devices named in this recall

3 devices in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other products in the same recall event

The FDA groups related products into one recall event (event 89689).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page