Babylog VN500 Ventilator with Software 2.51.01 and Lower-… recallZ-0871-2022 · Draeger Medical, Inc. · initiated Feb 14, 2022
Terminated
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Draeger Medical, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Cybersecurity Improvement Action-not equipped against potential cyber security threats
FDA root cause category: Nonconforming Material/Component
What the company is doing
Draeger issued "Urgent Medical Device Correction Notices" on 2/14/22 to the Director, Clinical Engineering and Risk Manager via FedEx with tracking/confirmation of delivery. Letter states reason for recall, health risk and action to take: Continue to follow the recommendations in the instructions for use: o Limit or control physical access to the abovementioned ventilators. o Do not connect any non-approved devices to the USB, LAN and DVI interfaces. o Be attentive to notifications, alarms, and alerts Consider closing/covering all unused USB, LAN and DVI interfaces Should you decide to close/cover unused interfaces, Draeger offers to provide tools upon request to cover or close these data interfaces of the ventilators free of charge. For the intended and authorized use of the interfaces the USB locks and interface covers can be removed with the provided tool. To order this material please contact Dragger Customer Success via phone between the hours of 8:00 AM 8:00 PM EST at 1-800-437-2437 (press 1 at the prompt, the 2), via fax at 1-215-723-5935, or via email at [email protected]. You will need to know the material number to order, which is dependent on the model of the C500 Cockpit that is installed on your V500/VN500 Ventilator. If you have any questions regarding the operation of your Evita V500 or Babylog VN500, please contact Drager Service Technical Support between the hours of 8:00 AM 8:00 PM EST at 1-800-437-2437 (press 2 at the prompt, then 2, then 1).
Recall details
- Recall number
- Z-0871-2022
- Status
- Terminated
- Initiated by the company
- Feb 14, 2022
- Recalling firm
- Draeger Medical, Inc.
- Quantity
- 13,687 (2,987 units US)
- Product code
- CBK
- FDA submissions
- K093632
- Distribution
- US Nationwide
Product description
Babylog VN500 Ventilator with Software 2.51.01 and Lower- intended for the ventilation neonatal patients from 0.4 kg (0.88 lbs) up to 10 kg (22 lbs) and pediatric patients from 5 kg (11 lbs) up to 20 kg (44 lbs) bodyweight Catalog Number: 8417400
Code information (models, lots, serial numbers)
All Lots/Serial Numbers. UDI: 04048675250418
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- IACS Workstation Neonatal Care 8417400Drägerwerk AG & Co. KGaA · Ventilator, Continuous, Facility Use1 recall66 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 89702).
- Z-0870-2022 · TerminatedEvita V500 Ventilator with Software 2.51.01 and Lower - intended for the ventilation of adult, pediatric and neonatal patients Catalog Number: 8416400
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page