DVR CROSSLOCK recallZ-0810-2022 · Biomet, Inc. · initiated Feb 14, 2022
Terminated
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Biomet, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Packaging of various implants may not have sufficient adhesion on all sides, leading to loss of sterile barrier integrity. Use of product poses risk of infection, leading to surgical intervention, and/or extension of surgery to find another part.
FDA root cause category: Under Investigation by firm
What the company is doing
A recall notification was sent via courier 24 Feb 2022 to distributors, surgeons, and hospital risk managers. Customers advised to locate and quarantine product, return to Biomet and identify any further customers receiving product. Recalled product is identified using specific lot and item numbers. Additionally, surgeons and hospital risk managers are to follow any specified patient monitoring instructions and return the Certificate of Acknowledgement to Zimmer Biomet. If you have further questions or concerns, please call customer service at 574-371-3071 or send questions via email to [email protected]
Recall details
- Recall number
- Z-0810-2022
- Status
- Terminated
- Initiated by the company
- Feb 14, 2022
- Recalling firm
- Biomet, Inc.
- Quantity
- 38
- Product code
- JDR
- FDA submissions
- K133939
- Distribution
- Worldwide: US (28 States AL, AZ, CA, CO, CT, FL, GA, ID, IL, IN, KS, KY, LA, MD, MI, MS, NC, NE, NM, NY, OH, OK, OR, SC, TX, UT, WA, WY), Australia, Japan, and Netherlands
Product description
BIOMET Trauma, [Item # 13182-22-26, DVR CROSSLOCK EXTRA LONG PLATE LEFT], [Item # 1318-12-126, DVR CROSSLOCK EXTRA LONG PLATE RIGHT], [Item # 1318-12-176, DVR CROSSLOCK EXTRA EXTRA LONG PLATE RIGHT]. Devices are packaged in a vacuum sealed nylon pouch, which is placed in a carton with the appropriate IFU and patient record labels. Identifying labels on the pouch and the carton.
Code information (models, lots, serial numbers)
6 lots: Lot #, UDI - [101220, (01)00887868446191(17)310623(10)101220], [101340, (01)00887868446207(17)310625(10)101340], [101450, (01)00887868446313(17)310630(10)101450], [101560, (01)00887868446191(17)310624(10)101560], [119180, (01)00887868446313(17)310630(10)119180], [154060, (01)00887868446313(17)310630(10)154060]
Devices named in this recall
3 devices in the FDA device registry whose identifier appears in the recall's code information.
- DVR® CROSSLOCK 1318-12-126Biomet Orthopedics, LLC · Plate, fixation, bone1 recall0 reports
- DVR® CROSSLOCK 1318-12-176Biomet Orthopedics, LLC · Plate, fixation, bone1 recall0 reports
- DVR® CROSSLOCK 1318-22-126Biomet Orthopedics, LLC · Plate, fixation, bone1 recall0 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 89701).
- Z-0809-2022 · TerminatedBIOMET Sports Medicine JUGGERLOC SLOTTED REAMER: [Item # 110010371, 6MM], [Item #110010372, 7MM], [Item # 110010373, 8MM], [Item #110018275, 5MM]. Devices are packaged in a vacuum …
- Z-0811-2022 · TerminatedBIOMET TRAUMA A.L.P.S. FIBULA COMPOSITE LOCKING PLATE 8 HOLE, Item # 856204008. Devices are packaged in a vacuum sealed nylon pouch, which is placed in a carton with the appropriat…
- Z-0812-2022 · TerminatedBIOMET TRAUMA Drill Bit, Tap, Guide Pin: [DVR Drill Bit, Item # 212000022, DVR 2.2MM DRILL BIT], [ALPS Drill Bit, Item # 214227160, 2.7X160MM CALIBRATED DRILL], [ALPS Drill Bit, It…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page