Atellica CH Calcium_2 (CA_2), Siemens Material Number… recallZ-0928-2022 · Siemens Healthcare Diagnostics, Inc. · initiated Feb 9, 2022
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Siemens Healthcare Diagnostics, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Siemens Healthcare Diagnostics Inc. has confirmed the potential for reagent carryover impacting quality control (QC), patient samples, and calibrator results.
FDA root cause category: Device Design
What the company is doing
Urgent Medical Device Correction notification letters dated 2/9/22 were sent to customers. Actions to be Taken by the Customer Please review this letter with your Medical Director. Perform the instructions provided in the Additional Information section below. Complete and return the Field Correction Effectiveness Check Form attached to this letter within 30 days. If you have received any complaints of illness or adverse events associated with the products listed in Table 1, immediately contact your local Siemens Healthineers Customer Care Center or your local Siemens Healthineers technical support representative. Please retain this letter with your laboratory records and forward this letter to those who may have received these products. Additional Information If your laboratory has multiple Atellica CH 930 Analyzers, separate the assays as follows: -LITH_2 from CA_2 and TBil_2 -A1c_E/A1c_H from CA_2 - CA_2 from UN_c (urine only); UN_c serum and plasma are not impacted and can be run on the same analyzer as CA_2 -Fruc from AAT, IgA_2, IgM_2 and Theo If you choose not to separate the assays as indicated above, batch testing of A1c_E, Fruc, LITH_2, and UN_c (urine only) may be considered. Note: An RPC2 wash mitigation must be initiated after completion of A1c_E, Fruc, LITH_2, and prior to UN_c (urine only). Any of the following will initiate the RPC2 wash: -After the Atellica CH 930 Analyzer has been in standby for 12 minutes. - After completion of any Open Channel assay. - Restarting the Atellica CH 930 Analyzer. Refer to the Atellica Solution Online Help for instructions on system restart. If you have any questions, contact your local Siemens Healthineers technical support representative.
Recall details
- Recall number
- Z-0928-2022
- Status
- Open, Classified
- Initiated by the company
- Feb 9, 2022
- Recalling firm
- Siemens Healthcare Diagnostics, Inc.
- Quantity
- 2,395
- Product code
- CJY
- FDA submissions
- K083386
- Distribution
- Global Distribution. US Nationwide.
Product description
Atellica CH Calcium_2 (CA_2), Siemens Material Number 11097644
Code information (models, lots, serial numbers)
UDI: 00630414220697; Lot Numbers: 211392 211438 211484 211522 211526
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Atellica CH CA_2 11097644Siemens Healthcare Diagnostics Inc. · AZO DYE, CALCIUM1 recall28 reports
Other devices under the same FDA submission
3 devices cleared under K083386 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- ADVIA® Chemistry CA_2c Calcium_2 Reagents, Concentrated - Arsenazo III 10283344Siemens Healthcare Diagnostics Inc. · AZO DYE, CALCIUM2 reports
- ADVIA® Chemistry CA_2 Calcium Reagents (Arsenazo III) 10316612Siemens Healthcare Diagnostics Inc. · AZO DYE, CALCIUM1 report
- ADVIA® Chemistry CA_2 Calcium_2 Reagents (Arsenazo III) 10341116Siemens Healthcare Diagnostics Inc. · AZO DYE, CALCIUM1 report
Other products in the same recall event
The FDA groups related products into one recall event (event 89740).
- Z-0929-2022 · Open, ClassifiedAtellica CH Enzymatic Hemoglobin A1c (A1c_E/A1c_H), Siemens Material Number 11097536
- Z-0930-2022 · Open, ClassifiedAtellica CH Fructosamine (Fruc), Siemens Material Number 11097637
- Z-0931-2022 · Open, ClassifiedAtellica CH Lithium_2 (LITH_2), Siemens Material Number 11097637
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page