Temperature Probes recallZ-0926-2022 · Philips North America, LLC · initiated Feb 7, 2022

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Philips North America, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Update to instructions for use regarding the cleaning and disinfection process for the reusable probes.

FDA root cause category: Device Design

What the company is doing

The firm, Philips, issued "URGENT Medical Device Recall" Customer notification letters on 07 Feb 2022 via certified mail. Actions planned by the firm: An IFU addendum (part# 453564978601) was created that provides updated instructions for cleaning, disinfection, and sterilization. This addendum is to be downloaded www.philips.com/ifu and kept with the main IFU (part# 453564635891). Customer should take the following actions to prevent risk to patients or users: Prepare and soak in an enzymatic solution (detergent) to remove any debris prior to disinfection or sterilization. To reduce risk of cross-infection always clean all probes before applying to a different patient. Esophageal/Rectal Probes must be cleaned and sterilized between patient use. Skin probes can be cleaned and disinfected or cleaned and sterilized between patient use. Contact Philips Customer Services with questions. US customers: 1 (800) 722-9377, Worldwide customers: Contact local Philips representative.

Recall details

Recall number
Z-0926-2022
Status
Open, Classified
Initiated by the company
Feb 7, 2022
Recalling firm
Philips North America, LLC
Quantity
478,931 units
Product code
FFL
FDA submissions
K962070
Distribution
Worldwide distribution: US (nationwide) Including Puerto Rico and OUS (foreign) countries including: Afghanistan, Albania, Algeria, Angola, Argentina, Aruba, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Barbados, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Botswana, Brazil, Brunei Darussalam, Bulgaria, Burundi, Cambodia, Canada, Chile, China, Columbia, Congo (Democratic Republic of the), Costa Rica, Cote D'lvoire, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador, Estonia, Ethiopia, Finland, France, French Guiana, French Polynesia, Gabon, Georgia, Germany, Ghana, Greece, Guadeloupe, Guatemala, Honduras, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Ireland, Israel, Japan, Jordan, Kazakhstan, Korea Republic of, Kuwait, Lao People's Democratic Republic, Latvia, Lebanon, Liberia, Libya, Lithuania, Luxembourg, Macao, Macedonia, Malaysia, Maldives, Mali, Malta, Mauritius, Mayotte, Mexico, Moldova, Mongolia, Montenegro, Morocco, Mozambique, Myanmar, Namibia, Nepal, Netherlands, Netherlands Antilles, New Zealand, Nicaragua, Nigeria, Norway, Oman, Pakistan, Palestine, State of, Panama, Papua New Guinea, Paraguay, Peru, Philippines, Poland, Qatar, Reunion, Romania, Russian Federation, Rwanda, Saint Lucia, Saint Pierre and Miquelon, Saudi Arabia, Senegal, Serbia, Singapore, Slovakia, South Africa, Spain, Sri Lanka, Suriname, Swaziland, Sweden, Switzerland, Syrian and Republic, Taiwan, Tajikistan, Tanzania, Thailand, Trinidad and Tobago, Tunisia, Turkey, Turkmenistan, Uganda, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Uzbekistan, Venezuela, Viet Nam, Yemen and Zambia.

Product description

The IFU for the Autoclavable Temperature Probes: Part # 453564635891, providing in the packaging for the following product model #s: 21075A, 21076A, 21078A, 989803162621, 989803162631, 989803162641.

Code information (models, lots, serial numbers)

All lots manufactured IFUs from 2008 to present. UDI for probe model 21075A (01)00884838000995, model 21076A (01)00884838001008, model 21078A (01)00884838001015, model 989803162621 (01)00884838008045, model 989803162631 (01)00884838008052, and model 989803162641 (01) 00884838008069