ARCHITECT recallZ-0891-2022 · Abbott Laboratories · initiated Mar 8, 2022

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Abbott Laboratories or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

This calibrator lot did not meet acceptance criteria during ongoing stability testing. When a calibration is performed with this lot, controls may exceed the specified range in the Instructions for Use (IFU).

FDA root cause category: Under Investigation by firm

What the company is doing

The firm, Abbott, notified their consignees by "Urgent Medical Device Products Recall" letter on 03/08/2022. The letter explained the issue and requested the use of the affected item be discontinued and destroyed, and the Customer Reply Form be completed and returned via FAX #: 1-800-777-0051 or email [email protected]. The firm is replacing the product. If you or any of the health care providers you serve have questions regarding this information, U.S. Customers please contact Customer Service at 1-877-4ABBOTT (available 24 hours a day, 7 days a week).

Recall details

Recall number
Z-0891-2022
Status
Open, Classified
Initiated by the company
Mar 8, 2022
Recalling firm
Abbott Laboratories
Quantity
402 kits
Product code
JIS
FDA submissions
K042924
Distribution
Worldwide Distribution: US (nationwide) and OUS (foreign) including countries of: AUSTRIA, BELGIUM, BOSNIA AND HERZEGOVI, CROATIA, CYPRUS, CZECH REPUBLIC, DENMARK, FRANCE, GERMANY, GREECE, IRAQ, ITALY, KAZAKHSTAN, LITHUANIA, LUXEMBOURG, MACEDONIA, MOROCCO, PANAMA, PORTUGAL, QATAR, ROMANIA, SAUDI ARABIA, SOUTH AFRICA, SWEDEN, SWITZERLAND, and TURKEY

Product description

ARCHITECT STAT Myoglobin Calibrators, RES 2K43-01 The ARCHITECT STAT Myoglobin Calibrators are for the calibration of the ARCHITECT iSystem with STAT protocol capability when used for the quantitative determination of myoglobin in human serum and plasma.

Code information (models, lots, serial numbers)

UDI: (01)00380740003272(17)220531(10)166000

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page